1-Year Results From a Multicenter Trial of a Polymer Surgical Mitral Valve
JACC · 2025-08 · 86 (7) · 515-526
DOI: 10.1016/j.jacc.2025.06.017 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift JACC, 2025-08. Verfasst von Isaac George und 15 weiteren Autor:innen. Beteiligte Einrichtung: Structural Heart and Valve Center, New York-Presbyterian Hospital/Columbia University Medical Center, New York, New York, USA. Electronic address: ig…. Hauptthemen laut MEDLINE: Heart Valve Prosthesis, Mitral Valve, Polymers, Heart Valve Prosthesis Implantation. Die Arbeit wurde bisher 15-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Foldax Inc..
Abstract (Originalsprache)
BACKGROUND: Polymer leaflet material may extend the durability of surgical mitral valve replacement (SMVR) to provide stable long-term hemodynamics. The India Mitral Surgical Trial sought to evaluate the safety and performance of a novel polymer leaflet material as part of a surgical mitral valve (MV) prosthesis. OBJECTIVES: In this study, the authors sought to report 1-year outcomes in patients undergoing SMVR for MV disease using the Tria Mitral Valve (Foldax). METHODS: Adult patients requiring MV replacement were enrolled in a prospective single-arm multicenter trial at 8 clinical sites in India from April to November 2023. An independent physician screening committee reviewed each patient for study eligibility before enrollment. Safety events were adjudicated per standard Valve Academic Research Consortium 3 criteria guidelines, and valve performance was assessed by means of echocardiographic and computed tomographic imaging at 30 days and 1 year. Patients were maintained on a vitamin K antagonist (target international normalized ratio: 2.5). RESULTS: Sixty-seven patients, of whom 64% were female (48% of childbearing age), with a mean age of 42 years (range: 19-67 years), mean body mass index of 22.7 kg/m2, and body surface area of 1.6 cm2 were treated with SMVR with 100% technical success. Most patients (54%) were NYHA functional class III or IV at baseline. The mean Society of Thoracic Surgeons score was 1.4%. The etiology of MV disease was stenosis in 27%, regurgitation in 30%, and mixed in 43% of patients, primarily secondary to rheumatic heart disease. The 1-year rates for all-cause mortality, thromboembolic events, stroke, structural valve deterioration, and valve reintervention were 9.1%, 7.5%, 4.9%, 0%, and 0%, respectively. No death was valve related. One-year effective orifice area and mean inflow gradient were 1.4 ± 0.4 cm2 and 4.6 ± 1.7 mm Hg, respectively. There were 2 thrombotic events and 3 ischemic strokes, all in patients with subtherapeutic international normalized ratio. CONCLUSIONS: The polymer surgical MV demonstrated an acceptable safety profile and maintained stable hemodynamic performance through 1 year in patients undergoing MV replacement. Further study of this promising polymer leaflet technology is ongoing. (Clinical Investigation for the Foldax Tria Mitral Valve-India; NCT06191718).
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Structural Heart and Valve Center, New York-Presbyterian Hospital/Columbia University Medical Center, New York, New York, USA. Electronic address: ig2006@cumc.columbia.edu.
- Indo-US Superspeciality Hospital, Hyderabad, Telangana, India.
- EPIC Hospital, Ahmedabad, Gujarat, India.
- Cardiovascular Research Foundation, New York, New York, USA.
- SMS Medical College and Attached Hospitals, Jaipur, Rajasthan, India.
- Fortis Escorts Heart Institute, New Delhi, Delhi, India.
Förderung
- Foldax Inc.
Zitieren
APA
George, I., Rao, D. P., Jain, A., Ascione, G., Sharma, M., Meharwal, Z. S., Sarkar, B., Kochar, N., Gan, M. D., Shastri, N., Runt, J., Whisenant, B., Wilson, B., Kiser, A., Leon, M. B., & Pandey, K. (2025). 1-Year Results From a Multicenter Trial of a Polymer Surgical Mitral Valve. JACC, 86(7), 515-526. https://doi.org/10.1016/j.jacc.2025.06.017
Vancouver
George I, Rao DP, Jain A, Ascione G, Sharma M, Meharwal ZS, et al. 1-Year Results From a Multicenter Trial of a Polymer Surgical Mitral Valve. JACC. 2025;86(7):515-526. doi:10.1016/j.jacc.2025.06.017
BibTeX
@article{George2025,
author = {George, Isaac and Rao, Dasari Prasada and Jain, Anil and Ascione, Guido and Sharma, Mohit and Meharwal, Z S and Sarkar, Binay and Kochar, Nitin and Gan, Mohan D and Shastri, Naman and Runt, James and Whisenant, Brian and Wilson, Brent and Kiser, Andy and Leon, Martin B and Pandey, Kaushal},
title = {1-Year Results From a Multicenter Trial of a Polymer Surgical Mitral Valve},
journal = {JACC},
year = {2025},
volume = {86},
number = {7},
pages = {515--526},
doi = {10.1016/j.jacc.2025.06.017}
}