Clinical Pilot Study of Ionizing Radiation Acoustic Imaging for Real-Time Visualization of Radiation Therapy Dose Delivery in Cancer Patients

Kyle C Cuneo, Wei Zhang, Dale Litzenberg, Yaocai Huang, Scott Hadley, Ibrahim Oraiqat, Kai-Wei Chang, Glebys Gonzalez, Sarah Dykstra, Man Zhang, Eduardo G Moros, Paul L Carson und 2 weitere

International Journal of Radiation Oncology*Biology*Physics · 2026

DOI: 10.1016/j.ijrobp.2025.12.006 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Radiation Oncology*Biology*Physics, 2026. Verfasst von Kyle C Cuneo und 13 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan. Hauptthemen laut MEDLINE: Radiosurgery, Acoustics, Abdominal Neoplasms, Radiotherapy, Image-Guided. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NCI NIH HHS, National Institutes of Health, University of Michigan.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: This study aimed to evaluate the feasibility of using ionizing radiation acoustic imaging (iRAI) to map the delivered dose in patients receiving radiation therapy (RT) with various treatment techniques, including 3-dimensional conformal RT, intensity modulated RT, and volumetric modulated arc therapy. METHODS AND MATERIALS: Patients with intra-abdominal cancer were enrolled in a prospective clinical trial after providing informed consent. Patients were treated with stereotactic body RT using standard clinical techniques with the addition of volumetric iRAI for real-time mapping of 3-dimensional radiation dose deposition. RESULTS: The minimal detectable dose was approximately 10 cGy. The overall shape of the dose distribution and the location of dose deposition, as determined by iRAI, matched the corresponding treatment plan. A gamma passing rate of 75.97% ± 11.12% and 90.99% ± 6.61% with a 10 mm/10% distance to agreement/dose difference suggests good agreement between the iRAI measurement and the treatment plan in both the liver volume and the planning target volume under the current system resolution. A structural similarity value of 0.6284 ± 0.1678 demonstrates that the iRAI measurements have shown structural pattern agreement with the treatment plan within the planning target volume. CONCLUSIONS: This study demonstrates the clinical feasibility of iRAI to monitor radiation dose delivery to deep targets in real time during treatment. This information can be integrated into treatment delivery systems to improve safety and facilitate the adaptation of RT. Despite the promising results achieved with the current form of technology, limitations in detection sensitivity, reconstruction accuracy, and acoustic coupling still need to be addressed.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

RadiosurgeryAcousticsAbdominal NeoplasmsRadiotherapy, Image-GuidedHumansPilot ProjectsFeasibility StudiesRadiotherapy DosageRadiotherapy, ConformalProspective StudiesRadiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-AssistedLiverLiver Neoplasms

Autor:innen und Institutionen

  • Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan.
  • Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan. Electronic address: zhawei@umich.edu.
  • Department of Biomedical Engineering, University of Michigan, Ann Arbor, Michigan.
  • Department of Machine Learning, Moffitt Cancer Center, Tampa, Florida; Department of Radiation Oncology, Moffitt Cancer Center, Tampa, Florida.
  • Department of Machine Learning, Moffitt Cancer Center, Tampa, Florida.
  • Department of Radiology, University of Michigan, Ann Arbor, Michigan.

Förderung

  • NCI NIH HHS
  • National Institutes of Health
  • University of Michigan
  • NIH National Cancer Institute

Zitieren

APA

Cuneo, K. C., Zhang, W., Litzenberg, D., Huang, Y., Hadley, S., Oraiqat, I., Chang, K. W., Gonzalez, G., Dykstra, S., Zhang, M., Moros, E. G., Carson, P. L., El Naqa, I., & Wang, X. (2026). Clinical Pilot Study of Ionizing Radiation Acoustic Imaging for Real-Time Visualization of Radiation Therapy Dose Delivery in Cancer Patients. International Journal of Radiation Oncology*Biology*Physics. https://doi.org/10.1016/j.ijrobp.2025.12.006

Vancouver

Cuneo KC, Zhang W, Litzenberg D, Huang Y, Hadley S, Oraiqat I, et al. Clinical Pilot Study of Ionizing Radiation Acoustic Imaging for Real-Time Visualization of Radiation Therapy Dose Delivery in Cancer Patients. International Journal of Radiation Oncology*Biology*Physics. 2026. doi:10.1016/j.ijrobp.2025.12.006

BibTeX
@article{Cuneo2026,
  author = {Cuneo, Kyle C and Zhang, Wei and Litzenberg, Dale and Huang, Yaocai and Hadley, Scott and Oraiqat, Ibrahim and Chang, Kai-Wei and Gonzalez, Glebys and Dykstra, Sarah and Zhang, Man and Moros, Eduardo G and Carson, Paul L and El Naqa, Issam and Wang, Xueding},
  title = {Clinical Pilot Study of Ionizing Radiation Acoustic Imaging for Real-Time Visualization of Radiation Therapy Dose Delivery in Cancer Patients},
  journal = {International Journal of Radiation Oncology*Biology*Physics},
  year = {2026},
  doi = {10.1016/j.ijrobp.2025.12.006}
}
Kostenlos starten →