Dose Reduction in 4-Dimensional Computed Tomography Imaging: Breathing Signal-Guided Deep Learning-Driven Data Acquisition

Lukas Wimmert, Tobias Gauer, Jannis Dickmann, Christian Hofmann, Thilo Sentker, Rene Werner

International Journal of Radiation Oncology*Biology*Physics · 2026

DOI: 10.1016/j.ijrobp.2025.08.047 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Radiation Oncology*Biology*Physics, 2026. Verfasst von Lukas Wimmert und 5 weiteren Autor:innen. Beteiligte Einrichtung: Institute for Applied Medical Informatics, University Medical Center Hamburg-Eppendorf, Hamburg, Germany; Institute of Computational Neuroscience, Un…. Hauptthemen laut MEDLINE: Deep Learning, Four-Dimensional Computed Tomography, Respiration, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Siemens Healthineers, German Research Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: Four-dimensional computed tomography (4D CT) imaging is essential for radiation therapy planning in thoracic tumors. However, current protocols tend to acquire more projection data than is strictly necessary for reconstructing the 4D CT, potentially leading to unnecessary radiation exposure and a misalignment with the ALARA (As Low As Reasonably Achievable) principle. We propose a deep learning (DL)-driven approach that uses the patient's breathing signal to guide data acquisition, aiming to acquire only necessary projection data. METHODS AND MATERIALS: This retrospective study analyzed 1415 breathing signals from 294 patients, with a 75/25 training/validation split at the patient level. On the basis of the signals, a DL model was trained to predict optimal beam-on events for projection data acquisition. Model testing was performed on 104 independent clinical 4D CT scans. The performance of the model was assessed by measuring temporal alignment between predicted and optimal beam-on events. To assess the impact on the reconstructed images, each 4D CT data set was reconstructed twice: using all clinically acquired projections (reference) and using only the model-selected projections (dose-reduced). Reference and dose-reduced images were compared using Dice coefficients for organ segmentations, deformable image registration-based displacement fields, artifact frequency, and tumor segmentation agreement, the latter evaluated in terms of Hausdorff distance and tumor motion ranges. RESULTS: The proposed approach reduced beam-on time and imaging dose by a median of 29% (IQR, 24%-35%), corresponding to 11.6 mGy dose reduction for a standard 4D CT CTDIvol of 40 mGy. Temporal alignment between predicted and optimal beam-on events showed marginal differences. Similarly, reconstructed dose-reduced images showed only minimal differences to the reference images, demonstrated by high lung and liver segmentation Dice values, small-magnitude (deformable image registration) displacement fields, and unchanged artifact frequency. Minor deviations of tumor segmentation and motion ranges compared with the reference suggest only minimal impact of the proposed approach on treatment planning. CONCLUSIONS: The proposed DL-driven data acquisition approach has the ability to reduce radiation exposure during 4D CT imaging while preserving diagnostic quality, offering a clinically viable, ALARA-adhering solution for 4D CT imaging.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Deep LearningFour-Dimensional Computed TomographyRespirationRadiotherapy Planning, Computer-AssistedRadiation DosageHumansRetrospective StudiesLung NeoplasmsFemaleMale

Autor:innen und Institutionen

  • Institute for Applied Medical Informatics, University Medical Center Hamburg-Eppendorf, Hamburg, Germany; Institute of Computational Neuroscience, University Medical Center Hamburg-Eppendorf, Hamburg, Germany; Center fo…
  • Department of Radiotherapy and Radiation Oncology, University Medical Center Hamburg-Eppendorf, Hamburg, Germany.
  • Siemens Healthineers AG, Forchheim, Germany.

Förderung

  • Siemens Healthineers
  • German Research Foundation

Zitieren

APA

Wimmert, L., Gauer, T., Dickmann, J., Hofmann, C., Sentker, T., & Werner, R. (2026). Dose Reduction in 4-Dimensional Computed Tomography Imaging: Breathing Signal-Guided Deep Learning-Driven Data Acquisition. International Journal of Radiation Oncology*Biology*Physics. https://doi.org/10.1016/j.ijrobp.2025.08.047

Vancouver

Wimmert L, Gauer T, Dickmann J, Hofmann C, Sentker T, Werner R. Dose Reduction in 4-Dimensional Computed Tomography Imaging: Breathing Signal-Guided Deep Learning-Driven Data Acquisition. International Journal of Radiation Oncology*Biology*Physics. 2026. doi:10.1016/j.ijrobp.2025.08.047

BibTeX
@article{Wimmert2026,
  author = {Wimmert, Lukas and Gauer, Tobias and Dickmann, Jannis and Hofmann, Christian and Sentker, Thilo and Werner, Rene},
  title = {Dose Reduction in 4-Dimensional Computed Tomography Imaging: Breathing Signal-Guided Deep Learning-Driven Data Acquisition},
  journal = {International Journal of Radiation Oncology*Biology*Physics},
  year = {2026},
  doi = {10.1016/j.ijrobp.2025.08.047}
}
Kostenlos starten →