EUVIMEDEUVIMED entdecken
Crossref

A Pilot Study of Simulation-Free Hippocampal-Avoidance Whole Brain Radiation Therapy Using Diagnostic MRI-Based and Online Adaptive Planning

Kylie H Kang, Alex T Price, Francisco J Reynoso, Eric Laugeman, Eric D Morris, Pamela P Samson, Jiayi Huang, Shahed N Badiyan, Hyun Kim, Randall J Brenneman, Christopher D Abraham, Nels C Knutson und 1 weitere

International Journal of Radiation Oncology*Biology*Physics · 2024-08 · 119 (5) · 1422-1428

DOI: 10.1016/j.ijrobp.2024.03.039 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Radiation Oncology*Biology*Physics, 2024-08. Verfasst von Kylie H Kang und 12 weiteren Autor:innen. Beteiligte Einrichtung: Washington University School of Medicine in St Louis, Department of Radiation Oncology, St Louis, Missouri. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Magnetic Resonance Imaging, Hippocampus, Feasibility Studies. Die Arbeit wurde bisher 20-mal zitiert (Crossref) und nennt 18 Literaturangaben. Gefördert von: Varian Medical Systems.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: We aimed to demonstrate the clinical feasibility and safety of simulation-free hippocampal avoidance whole brain radiation therapy (HA-WBRT) in a pilot study (National Clinical Trial 05096286). METHODS AND MATERIALS: Ten HA-WBRT candidates were enrolled for treatment on a commercially available computed tomography (CT)-guided linear accelerator with online adaptive capabilities. Planning structures were contoured on patient-specific diagnostic magnetic resonance imaging (MRI), which were registered to a CT of similar head shape, obtained from an atlas-based database (AB-CT). These patient-specific diagnostic MRI and AB-CT data sets were used for preplan calculation, using NRG-CC001 constraints. At first fraction, AB-CTs were used as primary data sets and deformed to patient-specific cone beam CTs (CBCT) to give patient-matched density information. Brain, ventricle, and brain stem contours were matched through rigid translation and rotation to the corresponding anatomy on CBCT. Lens, optic nerve, and brain contours were manually edited based on CBCT visualization. Preplans were then reoptimized through online adaptation to create final, simulation-free plans, which were used if they met all objectives. Workflow tasks were timed. In addition, patients underwent CT-simulation to create immobilization devices and for prospective dosimetric comparison of simulation-free and simulation-based plans. RESULTS: Median time from MRI importation to completion of "preplan" was 1 weekday (range, 1-4). Median on-table workflow duration was 41 minutes (range, 34-70). NRG-CC001 constraints were achieved by 90% of the simulation-free plans. One patient's simulation-free plan failed a planning target volume coverage objective (89% instead of 90% coverage); this was deemed acceptable for first-fraction delivery, with an offline replan used for subsequent fractions. Both simulation-free and simulation CT-based plans otherwise met constraints, without clinically meaningful differences. CONCLUSIONS: Simulation-free HA-WBRT using online adaptive radiation therapy is feasible, safe, and results in dosimetrically comparable treatment plans to simulation CT-based workflows while providing convenience and time savings for patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedMagnetic Resonance ImagingHippocampusFeasibility StudiesBrain NeoplasmsCranial IrradiationCone-Beam Computed TomographyOrgans at RiskRadiotherapy, Image-GuidedHumansPilot ProjectsOrgan Sparing TreatmentsMale

Autor:innen und Institutionen

  • Washington University School of Medicine in St Louis, Department of Radiation Oncology, St Louis, Missouri.
  • University Hospitals, Department of Radiation Oncology, Case Western Reserve University, Cleveland, Ohio.
  • University of Texas Southwestern Medical Center, Department of Radiation Oncology, Dallas, Texas.
  • Banner MD Anderson Cancer Center at Banner North Colorado Medical Center, Greeley, Colorado.
  • University Hospitals, Department of Radiation Oncology, Case Western Reserve University, Cleveland, Ohio. Electronic address: lauren.henke@uhhospitals.org.

Förderung

  • Varian Medical Systems

Zitieren

APA

Kang, K. H., Price, A. T., Reynoso, F. J., Laugeman, E., Morris, E. D., Samson, P. P., Huang, J., Badiyan, S. N., Kim, H., Brenneman, R. J., Abraham, C. D., Knutson, N. C., & Henke, L. E. (2024). A Pilot Study of Simulation-Free Hippocampal-Avoidance Whole Brain Radiation Therapy Using Diagnostic MRI-Based and Online Adaptive Planning. International Journal of Radiation Oncology*Biology*Physics, 119(5), 1422-1428. https://doi.org/10.1016/j.ijrobp.2024.03.039

Vancouver

Kang KH, Price AT, Reynoso FJ, Laugeman E, Morris ED, Samson PP, et al. A Pilot Study of Simulation-Free Hippocampal-Avoidance Whole Brain Radiation Therapy Using Diagnostic MRI-Based and Online Adaptive Planning. International Journal of Radiation Oncology*Biology*Physics. 2024;119(5):1422-1428. doi:10.1016/j.ijrobp.2024.03.039

BibTeX
@article{Kang2024,
  author = {Kang, Kylie H and Price, Alex T and Reynoso, Francisco J and Laugeman, Eric and Morris, Eric D and Samson, Pamela P and Huang, Jiayi and Badiyan, Shahed N and Kim, Hyun and Brenneman, Randall J and Abraham, Christopher D and Knutson, Nels C and Henke, Lauren E},
  title = {A Pilot Study of Simulation-Free Hippocampal-Avoidance Whole Brain Radiation Therapy Using Diagnostic MRI-Based and Online Adaptive Planning},
  journal = {International Journal of Radiation Oncology*Biology*Physics},
  year = {2024},
  volume = {119},
  number = {5},
  pages = {1422--1428},
  doi = {10.1016/j.ijrobp.2024.03.039}
}