Real-world performance and diagnostic yield gains of a next-generation patch-based 14-day uninterrupted ambulatory electrocardiogram monitor

Bradley P. Knight, Relana Pinkerton, Jeffrey Ellis, Jeffrey M. Ashburner, Oussama M. Wazni, Rod S. Passman, Mintu P. Turakhia, Suneet Mittal

Heart Rhythm O2 · 2026-10

DOI: 10.1016/j.hroo.2026.10.005 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Heart Rhythm O2, 2026-10. Verfasst von Bradley P. Knight und 7 weiteren Autor:innen. Die Arbeit nennt 12 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: iRhythm Technologies Inc.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The impact of device-specific differences in wear performance and analyzable ECG duration on arrhythmia yield has not been well characterized at a population level. We evaluated device performance and arrhythmia yield in a large post-market comparison of two long-term continuous monitors.

Methods

We performed a retrospective study comparing the post-market performance of an updated patch-based long-term continuous monitor (LTCM; iRhythm Technologies, San Francisco, CA) with its predecessor among patients prescribed 14-day monitoring. Both devices used the same FDA-cleared algorithm and technician review workflow. Outcomes included wear performance, analyzable ECG duration, arrhythmia yield (atrial fibrillation, supraventricular tachycardia, ventricular tachycardia, pauses, and atrioventricular block), skin irritation, and patient satisfaction.

Results

In 911,819 patients (age 61.8 ± 17.7 years, 53.7% female), the new LTCM (n=446,344) had longer wear time (mean 12.8 ± 2.6 vs. 12.1 ± 3.3 days, p

Förderung

  • iRhythm Technologies Inc

Zitieren

APA

Knight, B. P., Pinkerton, R., Ellis, J., Ashburner, J. M., Wazni, O. M., Passman, R. S., Turakhia, M. P., & Mittal, S. (2026). Real-world performance and diagnostic yield gains of a next-generation patch-based 14-day uninterrupted ambulatory electrocardiogram monitor. Heart Rhythm O2. https://doi.org/10.1016/j.hroo.2026.10.005

Vancouver

Knight BP, Pinkerton R, Ellis J, Ashburner JM, Wazni OM, Passman RS, et al. Real-world performance and diagnostic yield gains of a next-generation patch-based 14-day uninterrupted ambulatory electrocardiogram monitor. Heart Rhythm O2. 2026. doi:10.1016/j.hroo.2026.10.005

BibTeX
@article{Knight2026,
  author = {Knight, Bradley P. and Pinkerton, Relana and Ellis, Jeffrey and Ashburner, Jeffrey M. and Wazni, Oussama M. and Passman, Rod S. and Turakhia, Mintu P. and Mittal, Suneet},
  title = {Real-world performance and diagnostic yield gains of a next-generation patch-based 14-day uninterrupted ambulatory electrocardiogram monitor},
  journal = {Heart Rhythm O2},
  year = {2026},
  doi = {10.1016/j.hroo.2026.10.005}
}
Kostenlos starten →