Smartwatch-Based Heart Rate Detection in Atrial, Ventricular and Pulseless Arrhythmias
Heart, Lung and Circulation · 2026-09 · 35 (9) · 1298-1305
DOI: 10.1016/j.hlc.2026.02.016 ↗Kurz gefasst
Fachartikel in der Zeitschrift Heart, Lung and Circulation, 2026-09. Verfasst von Samual Turnbull und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Westmead Hospital, Westmead Applied Research Centre, University of Sydney, Westmead, NSW, Australia. Hauptthemen laut MEDLINE: Heart Rate, Wearable Electronic Devices, Electrocardiography, Atrial Fibrillation. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: National Heart Foundation of Australia.
Abstract (Originalsprache)
BACKGROUND & AIM: Smartwatches can provide a non-invasive heart rate (HR) monitoring to aid screening for atrial fibrillation, however, the accuracy of HR detections in the context of other arrhythmias requires further investigation. We aimed to assess the accuracy of HR measurements of two commonly used wearables at onset, during and termination of various induced and paced arrhythmias. METHOD: Patients undergoing an electrophysiology study were fitted intraprocedurally with both the Apple Watch Series 4 (AW) and Fitbit Versa (FB). Detected HRs on the devices were recorded at initiation, during, and at termination of various paced and induced arrhythmias. True HR was confirmed via intracardiac electrograms and 12-lead electrocardiogram. Detected HRs on the AW and FB were assessed for accuracy, defined as HR with in ±10% of the true rate. RESULTS: A total 322 cardiac rhythms from 32 patients were analysed. Both AW and FB demonstrated reasonable HR detection during sinus rhythm (100% vs 87%; p=0.04) and supraventricular tachycardia (SVT) (89% vs 86%; p=1). Both devices performed poorly at detecting SVT termination (62% vs 41%; p=0.04) and ventricular tachycardia (41% vs 16%; p=0.007). Neither device identified changes in HR during ventricular fibrillation or asystole. CONCLUSIONS: Both the Apple Watch and Fitbit demonstrate reasonable accuracy in HR detection during regular rhythms of supraventricular origin, however, struggle with arrhythmia terminations and ventricular arrhythmias. Care should be taken when assessing HR trends for presence of cardiac arrhythmias.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Westmead Hospital, Westmead Applied Research Centre, University of Sydney, Westmead, NSW, Australia.
- Department of Cardiology, Westmead Hospital, Westmead Applied Research Centre, University of Sydney, Westmead, NSW, Australia. Electronic address: saurabh.kumar@sydney.edu.au.
Förderung
- National Heart Foundation of Australia
Zitieren
APA
Turnbull, S., Gupta, A., Garikapati, K., Bhaskaran, A., De Silva, K., Kotake, Y., Bennett, R. G., Campbell, T., Thomas, S. P., Chow, C. K., & Kumar, S. (2026). Smartwatch-Based Heart Rate Detection in Atrial, Ventricular and Pulseless Arrhythmias. Heart, Lung and Circulation, 35(9), 1298-1305. https://doi.org/10.1016/j.hlc.2026.02.016
Vancouver
Turnbull S, Gupta A, Garikapati K, Bhaskaran A, De Silva K, Kotake Y, et al. Smartwatch-Based Heart Rate Detection in Atrial, Ventricular and Pulseless Arrhythmias. Heart, Lung and Circulation. 2026;35(9):1298-1305. doi:10.1016/j.hlc.2026.02.016
BibTeX
@article{Turnbull2026,
author = {Turnbull, Samual and Gupta, Anunay and Garikapati, Kartheek and Bhaskaran, Ashwin and De Silva, Kasun and Kotake, Yasuhito and Bennett, Richard G and Campbell, Timothy and Thomas, Stuart P and Chow, Clara K and Kumar, Saurabh},
title = {Smartwatch-Based Heart Rate Detection in Atrial, Ventricular and Pulseless Arrhythmias},
journal = {Heart, Lung and Circulation},
year = {2026},
volume = {35},
number = {9},
pages = {1298--1305},
doi = {10.1016/j.hlc.2026.02.016}
}
EUVIMED