Reliability testing of an IMU-based 2-segment foot model for clinical gait analysis

Leandra Bauer, Maximilian Anselm Hamberger, Wolfgang Böcker, Hans Polzer, Sebastian Felix Baumbach

Gait & Posture · 2024-10 · 114 · 112-118

DOI: 10.1016/j.gaitpost.2024.09.010 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Gait & Posture, 2024-10. Verfasst von Leandra Bauer und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics and Trauma Surgery, Musculoskeletal University Center Munich (MUM), University Hospital, LMU Munich, Germany. Electronic a…. Hauptthemen laut MEDLINE: Gait Analysis, Foot. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: The one of the most commonly used reference system for clinical gait analysis is an optical motion capture system (OMC) using a multi-segment foot model. A time- and cost-efficient alternative could be an inertial measurement unit (IMU)-based systems. However, these are limited to a single segment approach for the foot and ankle. Therefore, the current setup was modified to be based on a 2-segment foot model, allowing for a separate analysis of the hind- and midfoot. The study aimed to evaluate the reliability (inter-rater, intra-rater, and test-retest reliability) of an IMU-based 2-segment foot model. MATERIAL AND METHODS: Twelve healthy subjects were recruited to test the inter-rater, intra-rater, and test-retest reliability of the new IMU based 2-segment foot model. Gait analysis was performed on a treadmill at a constant speed of 4 km/h. Kinematic data of the tibia/hindfoot, tibia/forefoot and hindfoot/forefoot over 100 % gait cycle were analyzed. The reliability was tested by using statistical parametric mapping (SPM) and the intraclass correlation coefficient (ICC). RESULTS: The SPM showed no significant difference for inter-, intra-rater, and test-retest reliability, but for a small segment of tibia/forefoot dorsiflexion test-retest reliability (2.1° difference). The single standard deviation measurement error for the sagittal and transverse plane was

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Gait AnalysisFootHumansReproducibility of ResultsMaleAdultFemaleBiomechanical PhenomenaYoung AdultGaitHealthy Volunteers

Stichwörter der Autor:innen

Gait analysisMulti-segment Foot ModelReliability

Autor:innen und Institutionen

  • Department of Orthopaedics and Trauma Surgery, Musculoskeletal University Center Munich (MUM), University Hospital, LMU Munich, Germany. Electronic address: Leandra.bauer@med.uni-muenchen.de.
  • Department of Orthopaedics and Trauma Surgery, Musculoskeletal University Center Munich (MUM), University Hospital, LMU Munich, Germany.

Zitieren

APA

Bauer, L., Hamberger, M. A., Böcker, W., Polzer, H., & Baumbach, S. F. (2024). Reliability testing of an IMU-based 2-segment foot model for clinical gait analysis. Gait & Posture, 114, 112-118. https://doi.org/10.1016/j.gaitpost.2024.09.010

Vancouver

Bauer L, Hamberger MA, Böcker W, Polzer H, Baumbach SF. Reliability testing of an IMU-based 2-segment foot model for clinical gait analysis. Gait & Posture. 2024;114:112-118. doi:10.1016/j.gaitpost.2024.09.010

BibTeX
@article{Bauer2024,
  author = {Bauer, Leandra and Hamberger, Maximilian Anselm and Böcker, Wolfgang and Polzer, Hans and Baumbach, Sebastian Felix},
  title = {Reliability testing of an IMU-based 2-segment foot model for clinical gait analysis},
  journal = {Gait & Posture},
  year = {2024},
  volume = {114},
  pages = {112--118},
  doi = {10.1016/j.gaitpost.2024.09.010}
}
Kostenlos starten →