A validated, safer and more efficient closed-system method for carboxyhemoglobin saturation in postmortem blood
Forensic Science International: Reports · 2026-06 · 13 · 100484
DOI: 10.1016/j.fsir.2026.100484 ↗Kurz gefasst
Fachartikel in der Zeitschrift Forensic Science International: Reports, 2026-06. Verfasst von Songhie An und 2 weiteren Autor:innen. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Ministry of the Interior and Safety, National Forensic Service.
Abstract (Originalsprache)
Accurate COHb quantification is critical for confirming carbon monoxide related deaths. We developed and validated a closed-system, rotation-based syringe method to saturate postmortem blood with CO, replacing open bubbling. In ten confirmed cases, syringe rotation reached > 90% COHb in about half the time, supported simultaneous batch processing, reduced CO consumption, and eliminated analyst exposure. CO-oximetry and headspace GC-TCD showed excellent linearity, low detection and quantification limits, and high precision and accuracy. This method provides rapid, safe, and reliable preparation and measurement of COHb standards, markedly improving throughput, efficiency, and occupational safety for forensic toxicology laboratories.
Förderung
- Ministry of the Interior and Safety
- National Forensic Service
Zitieren
APA
An, S., Kim, Y., & Park, W. (2026). A validated, safer and more efficient closed-system method for carboxyhemoglobin saturation in postmortem blood. Forensic Science International: Reports, 13, 100484. https://doi.org/10.1016/j.fsir.2026.100484
Vancouver
An S, Kim Y, Park W. A validated, safer and more efficient closed-system method for carboxyhemoglobin saturation in postmortem blood. Forensic Science International: Reports. 2026;13:100484. doi:10.1016/j.fsir.2026.100484
BibTeX
@article{An2026,
author = {An, Songhie and Kim, Yujeong and Park, Wooyong},
title = {A validated, safer and more efficient closed-system method for carboxyhemoglobin saturation in postmortem blood},
journal = {Forensic Science International: Reports},
year = {2026},
volume = {13},
pages = {100484},
doi = {10.1016/j.fsir.2026.100484}
}
EUVIMED