Digital forensic readiness of regulated medical devices: An empirical study of the GE venue ultrasound system
Forensic Science International: Digital Investigation · 2026-06 · 57 · 302121
DOI: 10.1016/j.fsidi.2026.302121 ↗Kurz gefasst
Fachartikel in der Zeitschrift Forensic Science International: Digital Investigation, 2026-06. Verfasst von Muhammad Shaharyar Yaqub und 4 weiteren Autor:innen. Die Arbeit nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: National Science Foundation.
Abstract (Originalsprache)
Medical devices increasingly operate as networked computing platforms, yet incident response remains difficult because many regulated systems restrict logging, execution, and evidence collection. Point-of-care ultrasound (POCUS) devices are especially challenging: they are workflow-driven, safety-adjacent, and often run embedded, vendor-managed Windows stacks that handle patient-associated data while limiting investigator access. This paper, to our knowledge, presents the first end-to-end forensic readiness and physical-memory forensics study of a POCUS ultrasound platform, using the GE Venue Ultrasound system as a case study. We introduce URSA, a volatile-first readiness workflow for constrained ultrasound devices, and operationalize it with a UI-state-driven acquisition design that ties memory collection to observable clinical modes. Across seven full physical memory dumps captured before and after controlled scan-and-review interactions, we demonstrate repeatable acquisition under segmented storage and restrictive execution boundaries. Memory analysis reconstructs the clinical launch chain and yields both security-relevant artifacts and clinical exposure artifacts, including structured patient-associated logs and renderable diagnostic images recovered from RAM, with persistence characterized across states and time offsets. We complement volatile results with targeted disk-side context to quantify permission boundaries and corroborate selected findings. These findings provide practical guidance for improving forensic readiness and privacy-aware response on regulated ultrasound platforms.
Förderung
- National Science Foundation
Zitieren
APA
Yaqub, M. S., Jo, W., Nadeem, T., Topsakal, E., & Ahmed, I. (2026). Digital forensic readiness of regulated medical devices: An empirical study of the GE venue ultrasound system. Forensic Science International: Digital Investigation, 57, 302121. https://doi.org/10.1016/j.fsidi.2026.302121
Vancouver
Yaqub MS, Jo W, Nadeem T, Topsakal E, Ahmed I. Digital forensic readiness of regulated medical devices: An empirical study of the GE venue ultrasound system. Forensic Science International: Digital Investigation. 2026;57:302121. doi:10.1016/j.fsidi.2026.302121
BibTeX
@article{Yaqub2026,
author = {Yaqub, Muhammad Shaharyar and Jo, Wooyeon and Nadeem, Tamer and Topsakal, Erdem and Ahmed, Irfan},
title = {Digital forensic readiness of regulated medical devices: An empirical study of the GE venue ultrasound system},
journal = {Forensic Science International: Digital Investigation},
year = {2026},
volume = {57},
pages = {302121},
doi = {10.1016/j.fsidi.2026.302121}
}