Round Robin Study for Evaluation an in vitro skin irritation test for medical device extracts using KeraSkinTM in Korea
Food and Chemical Toxicology · 2024
DOI: 10.1016/j.fct.2024.114942 ↗Kurz gefasst
Fachartikel in der Zeitschrift Food and Chemical Toxicology, 2024. Verfasst von Joohwan Kim und 4 weiteren Autor:innen. Beteiligte Einrichtung: Korean Center for the Validation of Alternative Methods (KoCVAM), National Institute of Food and Drug Safety Evaluation (NIFDS), Ministry of Food and…. Hauptthemen laut MEDLINE: Skin Irritancy Tests, Equipment and Supplies, Irritants. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 16 Literaturangaben. Gefördert von: Ministry of Food and Drug Safety.
Abstract (Originalsprache)
With the growing importance of alternative test methods that implement the 3Rs principles (Reduction, Refinement and Replacement) and the global importance of biological safety assessment data for medical devices is increasing. We have developed and optimized the 'KeraSkin™ Skin Irritation Test (KeraSkin™ SIT) for medical device' for regulatory application in biological evaluation according to ISO 10993-23. We conducted a round robin study to optimize and evaluate the performance of KeraSkin™ SIT for medical devices using KeraSkin™ Reconstructed Human Epidermis (RhE), which is developed and manufactured in Korea. This round robin study was performed to assess the transferability, reproducibility (within and between laboratories) and predictive capacity in 1 lead laboratory and 3 participating laboratories based on OECD Guidance Document 34. The predictive capacity, the results showed 83.3 % of sensitivity, 100 % of specificity and 91.6 % of accuracy. In conclusion, the results demonstrate that 'KeraSkin™ SIT for medical device' provides a robust test method for detecting irritant activity of medical device extracts and can be utilized for identifying low levels of potent irritants in medical device extracts. Therefore, it fulfills the requirements to be included as a 'me-too' test method to EpiDerm™ and SkinEthic™ skin irritation test in ISO 10993-23.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Korean Center for the Validation of Alternative Methods (KoCVAM), National Institute of Food and Drug Safety Evaluation (NIFDS), Ministry of Food and Drug Safety (MFDS), 187, Osongsaengmyong 2-ro, Osong-eup, Heungdeok-g…
- Korea Conformity Labotratories (KCL), 20, Bodeum 3-ro, Seo-gu, Inchen, Republic of Korea.
- Korea Testing & Research Institute (KTR), 12-63, Sandan-gil, Hwasun-eup, Hwasun-gun, Jeollanam-do, Republic of Korea.
- Korea Institute of Toxicology, 141, Gajeong-ro, Yuseong-gu, Daejeon, Republic of Korea.
Förderung
- Ministry of Food and Drug Safety
Zitieren
APA
Kim, J., Kang, N. H., Kim, J. S., Kim, I. S., & Kim, Y. (2024). Round Robin Study for Evaluation an in vitro skin irritation test for medical device extracts using KeraSkinTM in Korea. Food and Chemical Toxicology. https://doi.org/10.1016/j.fct.2024.114942
Vancouver
Kim J, Kang NH, Kim JS, Kim IS, Kim Y. Round Robin Study for Evaluation an in vitro skin irritation test for medical device extracts using KeraSkinTM in Korea. Food and Chemical Toxicology. 2024. doi:10.1016/j.fct.2024.114942
BibTeX
@article{Kim2024,
author = {Kim, Joohwan and Kang, Nam-Hee and Kim, Jin Sik and Kim, Il Soo and Kim, Younhee},
title = {Round Robin Study for Evaluation an in vitro skin irritation test for medical device extracts using KeraSkinTM in Korea},
journal = {Food and Chemical Toxicology},
year = {2024},
doi = {10.1016/j.fct.2024.114942}
}
EUVIMED