Comparative biomechanical study of different screw fixation methods for minimally invasive hallux valgus surgery: A finite element analysis
Foot and Ankle Surgery · 2025-02 · 31 (2) · 160-169
DOI: 10.1016/j.fas.2024.09.001 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Foot and Ankle Surgery, 2025-02. Verfasst von T L Lewis und 7 weiteren Autor:innen. Beteiligte Einrichtung: King's Foot and Ankle Unit, King's College Hospital NHS Foundation Trust, UK. Electronic address: t.lewis2@nhs.net. Hauptthemen laut MEDLINE: Hallux Valgus, Bone Screws, Minimally Invasive Surgical Procedures. Die Arbeit wurde bisher 20-mal zitiert (Crossref) und nennt 39 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND: There are different screw configurations utilised for minimally invasive hallux valgus (HV) deformity despite limited biomechanical data assessing the stability and strength of each construct. We aimed to compare the strength of various screw configurations for minimally invasive HV surgery using finite element analysis (FEA). METHODS: A FEA model was developed from a CT of a female with moderate HV deformity. Five screw configurations utilizing one or two bicortical or intramedullary screws were tested. Stress analysis considered osteotomy displacement, maximum and minimum principal stresses, and von Mises stress for both implants and bone for each screw configuration. RESULTS: Fixation with two screws (one bicortical and one intramedullary) demonstrated the lowest values for osteotomy displacement, minimum and maximum total stress, and equivalent von Mises stress on the bone and screws in both loading conditions. CONCLUSION: The optimal configuration when performing minimally invasive surgery for moderate HV is one bicortical and one intramedullary screw. LEVEL OF EVIDENCE: Level III.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- King's Foot and Ankle Unit, King's College Hospital NHS Foundation Trust, UK. Electronic address: t.lewis2@nhs.net.
- Department of Orthopedic Surgery,Hospital Santa Helena, Brasília, DF, Brazil.
- Foot and Ankle Surgery Group, Orthopaedics and Traumatology Unit, Prevent Senior, São Paulo, Brazil.
- Head of Foot and Ankle Surgery Group, Orthopaedics and Traumatology Unit, Prevent Senior, São Paulo, Brazil.
- São Paulo University, USP, Ribeirão Preto, SP, Brazil.
- Faculty of Medical Sciences, Belo Horizonte, Minas Gerais, Brazil; Mater Dei Healthcare Network, Belo Horizonte, Minas Gerais, Brazil.
Zitieren
APA
Lewis, T. L., Mansur, H., Ferreira, G. F., Filho, M. V. P., Battaglion, L. R., Zambelli, R., Ray, R., & Nunes, G. A. (2025). Comparative biomechanical study of different screw fixation methods for minimally invasive hallux valgus surgery: A finite element analysis. Foot and Ankle Surgery, 31(2), 160-169. https://doi.org/10.1016/j.fas.2024.09.001
Vancouver
Lewis TL, Mansur H, Ferreira GF, Filho MVP, Battaglion LR, Zambelli R, et al. Comparative biomechanical study of different screw fixation methods for minimally invasive hallux valgus surgery: A finite element analysis. Foot and Ankle Surgery. 2025;31(2):160-169. doi:10.1016/j.fas.2024.09.001
BibTeX
@article{Lewis2025,
author = {Lewis, T L and Mansur, H and Ferreira, G F and Filho, M V P and Battaglion, L R and Zambelli, R and Ray, R and Nunes, G A},
title = {Comparative biomechanical study of different screw fixation methods for minimally invasive hallux valgus surgery: A finite element analysis},
journal = {Foot and Ankle Surgery},
year = {2025},
volume = {31},
number = {2},
pages = {160--169},
doi = {10.1016/j.fas.2024.09.001}
}