EUVIMEDEUVIMED entdecken
Validierungsstudie

Validation of retinal oximetry vessel selection using fluorescein angiography in patients with optic disc drusen

Mia Uhre Guldfeldt, Freja Pind Pilegaard, Lasse Malmqvist, Oliver Niels Klefter, Steffen Hamann

Experimental Eye Research · 2024

DOI: 10.1016/j.exer.2024.109882 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Experimental Eye Research, 2024. Verfasst von Mia Uhre Guldfeldt und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Rigshospitalet, Copenhagen, Denmark; Department of Clinical Medicine, Faculty of Health and Medical Sciences, University…. Hauptthemen laut MEDLINE: Oximetry, Fluorescein Angiography, Retinal Vessels, Optic Disk Drusen. Die Arbeit nennt 34 Literaturangaben. Gefördert von: Velux Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Retinal oximetry could provide insights into the pathophysiology of optic nerve disease, including optic disc drusen (ODD). Vessel selection for oximetry analysis is based on morphological characteristics of arterioles and venules and supported by an overlay of estimated blood oxygen saturations. The purpose of this cross-sectional study was to determine the validity of this vessel selection procedure by comparing it with vessel selection supported by video fluorescein angiography (FA). The study included 36 eyes of 36 patients with ODD who underwent retinal oximetry (Oxymap retinal oximeter T1) followed by FA (Heidelberg Spectralis). Two trained graders selected vessel segments in a pre-defined measurement area around the optic disc. One of these graders additionally performed the vessel segment selection with the support of FA images. When performed by the same grader, FA-supported and non-FA-supported vessel selection did not lead to significant differences in total vessel segment length, estimated oxygen saturations or vessel diameters (all p > 0.05). Inter-grader differences were found for arterial and venous segment lengths and arterial saturation (p < 0.05). A similar tendency was found for the arteriovenous saturation difference (p = 0.10). In conclusion, identifying vessel segments for retinal oximetry analysis based on vessel morphology and supported by a color-coded saturation overlay appears to be a valid method without the need for invasive angiography. A numerically small inter-grader variation may influence oximetry results. Further studies of retinal oximetry in ODD are warranted.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

OximetryFluorescein AngiographyRetinal VesselsOptic Disk DrusenHumansFemaleMaleCross-Sectional StudiesMiddle AgedAdultOxygenReproducibility of ResultsAgedOxygen SaturationOptic Disk

Stichwörter der Autor:innen

Fluorescein angiographyIn vivo imagingOptic disc drusenOxymapRetinal blood flowRetinal diseasesRetinal oximetryRetinal vessel saturation

Substanzen

Oxygen

Autor:innen und Institutionen

  • Department of Ophthalmology, Rigshospitalet, Copenhagen, Denmark; Department of Clinical Medicine, Faculty of Health and Medical Sciences, University of Copenhagen, Denmark.
  • Department of Ophthalmology, Rigshospitalet, Copenhagen, Denmark.
  • Department of Ophthalmology, Rigshospitalet, Copenhagen, Denmark; Department of Clinical Medicine, Faculty of Health and Medical Sciences, University of Copenhagen, Denmark. Electronic address: mia.guldfeldt@gmail.com.

Förderung

  • Velux Foundation

Zitieren

APA

Guldfeldt, M. U., Pilegaard, F. P., Malmqvist, L., Klefter, O. N., & Hamann, S. (2024). Validation of retinal oximetry vessel selection using fluorescein angiography in patients with optic disc drusen. Experimental Eye Research. https://doi.org/10.1016/j.exer.2024.109882

Vancouver

Guldfeldt MU, Pilegaard FP, Malmqvist L, Klefter ON, Hamann S. Validation of retinal oximetry vessel selection using fluorescein angiography in patients with optic disc drusen. Experimental Eye Research. 2024. doi:10.1016/j.exer.2024.109882

BibTeX
@article{Guldfeldt2024,
  author = {Guldfeldt, Mia Uhre and Pilegaard, Freja Pind and Malmqvist, Lasse and Klefter, Oliver Niels and Hamann, Steffen},
  title = {Validation of retinal oximetry vessel selection using fluorescein angiography in patients with optic disc drusen},
  journal = {Experimental Eye Research},
  year = {2024},
  doi = {10.1016/j.exer.2024.109882}
}
Kostenlos starten →