Development of a high-performance label-free electrochemical immunosensor for early cancer diagnosis using anti-CEA/Ag-MOF/GO/GCE nanocomposite

Peng Huang, Lingzhang Meng, Jun Pang, Haiting Huang, Jing Ma, Linlin He, Parnian Amani

Environmental Research · 2023

DOI: 10.1016/j.envres.2023.117178 ↗

Kurz gefasst

Zurückgezogene Publikation in der Zeitschrift Environmental Research, 2023. Verfasst von Peng Huang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nephrology, Affiliated Hospital of Youjiang Medical University for Nationalities, Guangxi, 533000, China. Hauptthemen laut MEDLINE: Metal-Organic Frameworks, Biosensing Techniques, Nanocomposites, Metal Nanoparticles. Die Arbeit wurde bisher 29-mal zitiert (Crossref) und nennt 64 Literaturangaben. Gefördert von: Natural Science Foundation of Guangxi Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In order to detect carcinoembryonic antigen (CEA) as a tumor marker in lung cancer for early cancer diagnosis, this study aimed to develop a label-free electrochemical immunosensor based on the immobilization of an Anti-CEA antibody on a metal-organic framework (MOF)-graphene oxide nanocomposite modified glassy carbon electrode (Anti-CEA/Ag-MOF/GO/GCE). Ag-MOF/GO nanocomposite was prepared on the GCE surface using the ultrasonic irradiation method, and Anti-CEA antibody was subsequently immobilized on the surface. Analysis of the crystal structure and morphology of the modified electrode using FE-SEM and XRD revealed that the correct combination of GO nanosheets and Ag-MOF nanoparticles produced a high surface area to trap the antibodies. Electrochemical tests utilizing the CV and DPV methods revealed that the immunosensor's sensitivity, stability, and selectivity were improved by Anti-CEA/Ag-MOF/GO/GCE. Results showed that, with a detection limit of 0.005 ng/mL, the change in the reduction peak current was inversely correlated with the logarithm concentration of CEA in the range of 10-3 to 5000 ng/mL. The suggested CEA immunosensor's applicability in a human serum sample was investigated, and findings of analytical studies via standard addition technique for both ELISA and DPV assays revealed that significant agreement existed between the outcomes of the two assays. Additionally, the recoveries ranged from 99.00% to 99.25%, and all relative standard deviations (RSDs) for the sample detections were below 5.01%, indicating satisfactory accuracy in results measured with the proposed CEA immunosensor, indicating that the prepared CEA immunosensor in this study can be used in clinical applications and human fluids.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Metal-Organic FrameworksBiosensing TechniquesNanocompositesMetal NanoparticlesNeoplasmsHumansCarcinoembryonic AntigenImmunoassayGoldLimit of Detection

Stichwörter der Autor:innen

Ag-MOFCarcinoembryonic antigenElectrochemical immunosensorGraphene oxideNanocomposite

Substanzen

Metal-Organic FrameworksCarcinoembryonic AntigenGold

Autor:innen und Institutionen

  • Department of Nephrology, Affiliated Hospital of Youjiang Medical University for Nationalities, Guangxi, 533000, China.
  • Center for Systemic Inflammation Research (CSIR), School of Preclinical Medicine, Youjiang Medical University for Nationalities, Guangxi, 533000, China.
  • Department of Molecular and Cell Biology, Faculty of Basic Sciences, University of Mazandaran, Babolsar, Iran. Electronic address: Parnianamanii1@gmail.com.

Förderung

  • Natural Science Foundation of Guangxi Province

Zitieren

APA

Huang, P., Meng, L., Pang, J., Huang, H., Ma, J., He, L., & Amani, P. (2023). Development of a high-performance label-free electrochemical immunosensor for early cancer diagnosis using anti-CEA/Ag-MOF/GO/GCE nanocomposite. Environmental Research. https://doi.org/10.1016/j.envres.2023.117178

Vancouver

Huang P, Meng L, Pang J, Huang H, Ma J, He L, et al. Development of a high-performance label-free electrochemical immunosensor for early cancer diagnosis using anti-CEA/Ag-MOF/GO/GCE nanocomposite. Environmental Research. 2023. doi:10.1016/j.envres.2023.117178

BibTeX
@article{Huang2023,
  author = {Huang, Peng and Meng, Lingzhang and Pang, Jun and Huang, Haiting and Ma, Jing and He, Linlin and Amani, Parnian},
  title = {Development of a high-performance label-free electrochemical immunosensor for early cancer diagnosis using anti-CEA/Ag-MOF/GO/GCE nanocomposite},
  journal = {Environmental Research},
  year = {2023},
  doi = {10.1016/j.envres.2023.117178}
}
Kostenlos starten →