EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Endoscopic Femoropopliteal Bypass Surgery Without Groin Incision: A Feasibility Study on Cadavers with the Da Vinci Single Port System

Rouven Berndt, E Sebastian Debus, Hubert Stein, Mark Preuss, Glenn Stante, Fabien Thaveau

EJVES Vascular Forum · 2026 · 65 · 222-226

DOI: 10.1016/j.ejvsvf.2026.04.005 ↗

Kurz gefasst

Fachartikel in der Zeitschrift EJVES Vascular Forum, 2026. Verfasst von Rouven Berndt und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Vascular Medicine, University Heart & Vascular Centre Hamburg-Eppendorf, Hamburg, Germany. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 16 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Femoropopliteal bypass surgery (FPBS) is a standard surgical procedure but is still associated with complications, especially surgical site infections and groin seromas. This feasibility study aimed to investigate whether endoscopic, robot assisted surgery enabled FPBS with minimal surgical trauma and without a groin incision.

Method

The da Vinci single port (SP) system (Intuitive Surgical, Sunnyvale, USA) was used to perform the novel technique of endoscopic FPBS on fresh frozen cadavers. For this novel technique, a 4 cm incision was made in the distal third of the medial thigh for supragenicular exposure of the proximal femoral artery. Thereafter, an end to side anastomosis between the common femoral artery and a polytetrafluoroethylene (PTFE) graft was performed endoscopically with the da Vinci SP system. Success of the procedure was defined by completion of the proximal bypass anastomosis and the avoidance of injuries to anatomic key structures. First evaluation of the technique was performed by assessment with a slightly modified Northwestern Objective Microanastomosis Assessment Tool (NOMAT) score.

Results

Two proximal bypass anastomoses were successfully sutured without injury to the five pre-defined key anatomic structures. Skin to skin complete duration of the intervention including removal of the robotic system was 105 minutes and 92 minutes, respectively. The modified NOMAT score representing the efficiency of surgical handling and quality of the anastomosis was 29 of 45 and 37 of 45, respectively.

Conclusion

The current study demonstrates that endoscopic, robot assisted FPBS was feasible within an acceptable timeframe and without injury to key anatomic structures. Further clinical development of endoscopic FPBS could enable minimally invasive bypass surgery including the advantages of conventional surgery but avoiding its limitations.

Schlagwörter

Stichwörter der Autor:innen

Endoscopic surgeryFresh frozen cadaverPeripheral bypass surgeryRobotic assisted surgeryVascular surgeryda Vinci single port system

Autor:innen und Institutionen

  • Department of Vascular Medicine, University Heart & Vascular Centre Hamburg-Eppendorf, Hamburg, Germany.
  • Intuitive Surgical Inc., Sunnyvale, CA, USA.
  • Department of Vascular Surgery, University Hospital Clermont-Ferrand, Clermont-Ferrand, France.

Zitieren

APA

Berndt, R., Debus, E. S., Stein, H., Preuss, M., Stante, G., & Thaveau, F. (2026). Endoscopic Femoropopliteal Bypass Surgery Without Groin Incision: A Feasibility Study on Cadavers with the Da Vinci Single Port System. EJVES Vascular Forum, 65, 222-226. https://doi.org/10.1016/j.ejvsvf.2026.04.005

Vancouver

Berndt R, Debus ES, Stein H, Preuss M, Stante G, Thaveau F. Endoscopic Femoropopliteal Bypass Surgery Without Groin Incision: A Feasibility Study on Cadavers with the Da Vinci Single Port System. EJVES Vascular Forum. 2026;65:222-226. doi:10.1016/j.ejvsvf.2026.04.005

BibTeX
@article{Berndt2026,
  author = {Berndt, Rouven and Debus, E Sebastian and Stein, Hubert and Preuss, Mark and Stante, Glenn and Thaveau, Fabien},
  title = {Endoscopic Femoropopliteal Bypass Surgery Without Groin Incision: A Feasibility Study on Cadavers with the Da Vinci Single Port System},
  journal = {EJVES Vascular Forum},
  year = {2026},
  volume = {65},
  pages = {222--226},
  doi = {10.1016/j.ejvsvf.2026.04.005}
}
Kostenlos starten →