Concave Triple Branched Stent Graft System for Aortic Arch Pathologies: A First in Man Prospective Cohort Study
European Journal of Vascular and Endovascular Surgery · 2026
DOI: 10.1016/j.ejvs.2025.06.009 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift European Journal of Vascular and Endovascular Surgery, 2026. Verfasst von Chang Shu und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Vascular Surgery, The Second Xiangya Hospital of Central South University, Changsha, China; State Key Laboratory of Cardiovascular Dise…. Hauptthemen laut MEDLINE: Stents, Aorta, Thoracic, Endovascular Procedures, Blood Vessel Prosthesis. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 27 Literaturangaben.
Abstract (Originalsprache)
OBJECTIVE: Endovascular aortic arch repair with supra-aortic trunk (SAT) reconstruction is an alternative to open surgical replacement, but the ideal stent graft is still under investigation. This study describes the first experience and one year results of a novel, off the shelf, concave supra-arch triple branched stent graft system (CS system) for aortic arch diseases. METHODS: This was a prospective cohort study conducted in three hospitals, enrolling patients from March 2022 to June 2023, to evaluate use of the CS system for aortic arch pathologies. An interdisciplinary board assessed patients pre-operatively. All patients were treated using the CS system to cover the aortic arch pathologies and reconstruct all SATs. Follow up assessments were conducted with computed tomography scans at two weeks and six and 12 months. RESULTS: Ten patients (eight men) were treated using the CS system. The mean age was 65.3 years. Conditions included thoracic aortic aneurysm, non-A non-B aortic dissection, and penetrating aortic arch ulcer. Technical success was achieved in all patients, with one type III endoleak during the procedure. The mean procedure time was 160 ± 23 minutes and the mean fluoroscopy time was 63 ± 20 minutes. The in hospital mortality and stroke rates were both 0%. The mean follow up duration was 456.9 ± 93.6 days. All patients recovered well without complications. The CS system completely covered the aortic pathologies, and all reconstructed branches remained patent. The single endoleak sealed spontaneously within six months. Mild to moderate in stent stenosis was observed in three left common carotid arteries. No other morphological abnormalities were observed, and secondary interventions were not necessary. CONCLUSION: The CS system for endovascular treatment of aortic arch pathologies demonstrated promising safety and efficacy in this study, with no deaths, strokes, retrograde type A aortic dissections, re-interventions, or significant endoleaks during one year follow up. Stenosis of the reconstructed SATs should be monitored closely.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Vascular Surgery, The Second Xiangya Hospital of Central South University, Changsha, China; State Key Laboratory of Cardiovascular Diseases, Centre of Vascular Surgery, Fuwai Hospital, National Centre for…
- Department of Vascular Surgery, The Second Xiangya Hospital of Central South University, Changsha, China.
- State Key Laboratory of Cardiovascular Diseases, Centre of Vascular Surgery, Fuwai Hospital, National Centre for Cardiovascular Diseases, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, Ch…
- Department of Vascular Surgery, Fuwai Yunnan Cardiovascular Hospital, Affiliated Cardiovascular Hospital of Kunming Medical University, Kunming, China.
Zitieren
APA
Shu, C., Wang, T., Fang, K., Li, Q., Luo, M., He, H., Li, X., Guo, Y., & Li, M. (2026). Concave Triple Branched Stent Graft System for Aortic Arch Pathologies: A First in Man Prospective Cohort Study. European Journal of Vascular and Endovascular Surgery. https://doi.org/10.1016/j.ejvs.2025.06.009
Vancouver
Shu C, Wang T, Fang K, Li Q, Luo M, He H, et al. Concave Triple Branched Stent Graft System for Aortic Arch Pathologies: A First in Man Prospective Cohort Study. European Journal of Vascular and Endovascular Surgery. 2026. doi:10.1016/j.ejvs.2025.06.009
BibTeX
@article{Shu2026,
author = {Shu, Chang and Wang, Tun and Fang, Kun and Li, Quanming and Luo, Mingyao and He, Hao and Li, Xin and Guo, Yuanyuan and Li, Ming},
title = {Concave Triple Branched Stent Graft System for Aortic Arch Pathologies: A First in Man Prospective Cohort Study},
journal = {European Journal of Vascular and Endovascular Surgery},
year = {2026},
doi = {10.1016/j.ejvs.2025.06.009}
}
EUVIMED