Image quality comparison between low-dose thin-slice deep-learning reconstruction and standard-dose thick-slice hybrid iterative reconstruction in pediatric abdominal CT

Ryota Harai, Yasunori Nagayama, Soichiro Ishiuchi, Ryuya Yoshida, Taihei Inoue, Takumi Osaki, Kaori Shiraishi, Masafumi Kidoh, Seitaro Oda, Takeshi Nakaura, Toshinori Hirai

European Journal of Radiology · 2026

DOI: 10.1016/j.ejrad.2026.112703 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift European Journal of Radiology, 2026. Verfasst von Ryota Harai und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic Radiology, Graduate School of Medical Sciences, Kumamoto University, 1-1-1, Honjo, Chuo-ku, Kumamoto 860-8556, Japan. Hauptthemen laut MEDLINE: Radiography, Abdominal, Radiation Dosage, Deep Learning, Tomography, X-Ray Computed. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Kumamoto University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVES: To compare image quality between low-dose thin-slice deep-learning reconstruction (DLR) and standard-dose thick-slice hybrid iterative reconstruction (HIR) in pediatric abdominal CT. METHODS: 82 children (≤6 years) who underwent contrast-enhanced abdominal CT with standard-dose (STD, n = 41) or low-dose (LD, n = 41) protocol matched by age and weight were retrospectively identified. STD and LD images were reconstructed at 3.0-mm using HIR (STD-HIR/3.0 and LD-HIR/3.0, respectively). LD images were also reconstructed at 0.5-mm using HIR (LD-HIR/0.5) and DLR (LD-DLR/0.5). Size-specific dose estimate (SSDE) was compared between groups. Image noise and contrast-to-noise ratio (CNR) were quantified. Noise power spectrum (NPS) and edge-rise slope (ERS) were employed for noise texture and edge sharpness measures, respectively. For subjective evaluation, noise magnitude, noise texture, edge sharpness, delineation of small structures, and diagnostic confidence were rated on a 5-point scale (1 = undiagnostic, 5 = best). RESULTS: SSDE was on average 59.5% lower in LD than in STD group (1.7 ± 0.4 vs. 4.2 ± 0.8 mGy, p < 0.001). Image noise was lower in LD-DLR/0.5 compared to STD-HIR/3.0, LD-HIR/3.0, and LD-HIR/0.5 (8.7 ± 1.5, 9.6 ± 1.1, 11.7 ± 1.9, and 17.9 ± 2.3 HU, respectively, all p ≤ 0.017). LD-DLR/0.5 showed equivalent CNR (e.g., liver CNR: 7.1 ± 2.1 vs. 7.0 ± 2.8, p = 0.827) and higher ERS (64.3 ± 23.9 vs. 53.3 ± 10.6 HU/mm, p = 0.011) with similar average NPS frequency (0.281 ± 0.042 vs. 0.291 ± 0.027 mm-1, p = 0.177) compared to STD-HIR/3.0. Subjective scores for all criteria were higher in LD-DLR/0.5 than in STD-HIR/3.0 (e.g., diagnostic confidence score: 4.5 ± 0.5 vs. 3.4 ± 0.4, p < 0.001). CONCLUSION: Low-dose thin-slice DLR improved edge sharpness, small structure visualization, and diagnostic confidence in pediatric abdominal CT without increasing noise compared to standard-dose thick-slice HIR.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiography, AbdominalRadiation DosageDeep LearningTomography, X-Ray ComputedRadiographic Image Interpretation, Computer-AssistedRadiation ProtectionHumansFemaleChild, PreschoolInfantMaleReproducibility of ResultsContrast MediaChildSensitivity and SpecificityRetrospective StudiesRadiographic Image Enhancement

Stichwörter der Autor:innen

Deep learningImage enhancementMultidetector computed tomographyPediatricsRadiation exposure

Substanzen

Contrast Media

Autor:innen und Institutionen

  • Department of Diagnostic Radiology, Graduate School of Medical Sciences, Kumamoto University, 1-1-1, Honjo, Chuo-ku, Kumamoto 860-8556, Japan.
  • Department of Diagnostic Radiology, Graduate School of Medical Sciences, Kumamoto University, 1-1-1, Honjo, Chuo-ku, Kumamoto 860-8556, Japan. Electronic address: nagayama.yasunori@kuh.kumamoto-u.ac.jp.

Förderung

  • Kumamoto University

Zitieren

APA

Harai, R., Nagayama, Y., Ishiuchi, S., Yoshida, R., Inoue, T., Osaki, T., Shiraishi, K., Kidoh, M., Oda, S., Nakaura, T., & Hirai, T. (2026). Image quality comparison between low-dose thin-slice deep-learning reconstruction and standard-dose thick-slice hybrid iterative reconstruction in pediatric abdominal CT. European Journal of Radiology. https://doi.org/10.1016/j.ejrad.2026.112703

Vancouver

Harai R, Nagayama Y, Ishiuchi S, Yoshida R, Inoue T, Osaki T, et al. Image quality comparison between low-dose thin-slice deep-learning reconstruction and standard-dose thick-slice hybrid iterative reconstruction in pediatric abdominal CT. European Journal of Radiology. 2026. doi:10.1016/j.ejrad.2026.112703

BibTeX
@article{Harai2026,
  author = {Harai, Ryota and Nagayama, Yasunori and Ishiuchi, Soichiro and Yoshida, Ryuya and Inoue, Taihei and Osaki, Takumi and Shiraishi, Kaori and Kidoh, Masafumi and Oda, Seitaro and Nakaura, Takeshi and Hirai, Toshinori},
  title = {Image quality comparison between low-dose thin-slice deep-learning reconstruction and standard-dose thick-slice hybrid iterative reconstruction in pediatric abdominal CT},
  journal = {European Journal of Radiology},
  year = {2026},
  doi = {10.1016/j.ejrad.2026.112703}
}
Kostenlos starten →