Influence of scanning parameters on image quality, and repeatability of renal blood oxygenation level-dependent imaging: A preliminary study on 5 T MRI
European Journal of Radiology · 2025
DOI: 10.1016/j.ejrad.2025.112460 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Radiology, 2025. Verfasst von Ying Xiong und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Chongqing University Three Gorges Hospital, Chongqing 404000, China. Electronic address: yingyyx99@163.com. Hauptthemen laut MEDLINE: Magnetic Resonance Imaging, Renal Insufficiency, Chronic, Kidney, Oxygen. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 32 Literaturangaben. Gefördert von: Science Health Joint Medical Scientific Research Project of Chongqing.
Abstract (Originalsprache)
OBJECTIVE: To investigate the influence of scanning parameters on image quality and the repeatability of renal Blood Oxygenation Level-Dependent (BOLD) imaging at 5 T. METHODS: 20 healthy subjects and 25 patients with chronic kidney disease (CKD) were prospectively recruited for BOLD-MRI scans at 5 T using multiple parameter combinations. The Fisher's exact test, ANOVA, the Friedman test, and paired t-tests were used to compare the qualitative and quantitative indices across groups. The generalized linear models (GLM) analysis was further used to evaluate the effect of acquisition parameters on the measured R2* values. Subsequently, Pearson correlation and the intraclass correlation coefficient (ICC) were applied to assess the repeatability of R2* values in the high-resolution parameters in patients. RESULTS: Renal cortical and medullary R2* values demonstrated excellent agreement (all ICCs > 0.75) across varied in-plane resolution, slice thickness, echo number, and maximum TE. GLM revealed a significant positive effect of slice thickness on R2* in both cortex and medulla, a region-specific positive effect of maximum TE in the cortex only, and no significant effect of in-plane resolution or echo number. High resolution (1.25 mm × 1.25 mm × 3 mm) provided superior corticomedullary demarcation and clinical image quality, and renal cortical and medullary R2* values exhibited high inter-observer agreement (ICC: 0.992 and 0.971) and test-retest repeatability (all r > 0.90, P < 0.05). Compared with healthy subjects, the renal cortical and medullary R2* values in patients with CKD were significantly higher (cortical R2*: 21.13 Hz vs. 22.88 Hz, P < 0.001; medullary R2*: 40.69 Hz vs. 49.14 Hz, P < 0.001). CONCLUSION: This study showed that image resolution may slightly affect the quantification of the R2* values on 5 T renal BOLD-MRI. High-resolution BOLD imaging (1.25 mm × 1.25 mm × 3 mm, 10 echoes, maximum TE 40 ms) can be achieved at 5 T MRI with excellent inter-observer agreement and test-retest repeatability, validating its clinical applicability.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Radiology, Chongqing University Three Gorges Hospital, Chongqing 404000, China. Electronic address: yingyyx99@163.com.
- Department of Radiology, Chongqing University Three Gorges Hospital, Chongqing 404000, China. Electronic address: 843081898@qq.com.
- MR Scientific Collaboration, United Imaging Healthcare, Shanghai 201815, China. Electronic address: shaoxin_xiang@163.com.
- Department of Radiology, Chongqing University Three Gorges Hospital, Chongqing 404000, China. Electronic address: zeng_wenbing@163.com.
- School of Medical Imaging, North Sichuan Medical College, Sichuan 637000,China. Electronic address: rt84640721@163.com.
- MR Scientific Collaboration, United Imaging Healthcare, Shanghai 201815, China. Electronic address: fengzc2016@163.com.
Förderung
- Science Health Joint Medical Scientific Research Project of Chongqing
Zitieren
APA
Xiong, Y., Guan, C., Xiang, S., Zeng, W., Rong, T., Feng, Z., Wen, Y., & Kong, J. (2025). Influence of scanning parameters on image quality, and repeatability of renal blood oxygenation level-dependent imaging: A preliminary study on 5 T MRI. European Journal of Radiology. https://doi.org/10.1016/j.ejrad.2025.112460
Vancouver
Xiong Y, Guan C, Xiang S, Zeng W, Rong T, Feng Z, et al. Influence of scanning parameters on image quality, and repeatability of renal blood oxygenation level-dependent imaging: A preliminary study on 5 T MRI. European Journal of Radiology. 2025. doi:10.1016/j.ejrad.2025.112460
BibTeX
@article{Xiong2025,
author = {Xiong, Ying and Guan, Chuanjiang and Xiang, Shaoxin and Zeng, Wenbing and Rong, Ting and Feng, Zhichao and Wen, Yun and Kong, Junfeng},
title = {Influence of scanning parameters on image quality, and repeatability of renal blood oxygenation level-dependent imaging: A preliminary study on 5 T MRI},
journal = {European Journal of Radiology},
year = {2025},
doi = {10.1016/j.ejrad.2025.112460}
}
EUVIMED