EUVIMEDEUVIMED entdecken
EvaluationsstudieOpen Access · CC BY 4.0

Accuracy evaluation of an automatic landmarking method based on 3D segmented CT scans of full lower limbs with knee osteoarthritis

Clément Horteur, Nolwenn Fougeron, Benoît Gaulin, Ysé Roch, Marek Bucki, Estelle Palluel, Yohan Payan, Antoine Perrier

European Journal of Radiology · 2025-10 · 191 · 112292

DOI: 10.1016/j.ejrad.2025.112292 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift European Journal of Radiology, 2025-10. Verfasst von Clément Horteur und 7 weiteren Autor:innen. Beteiligte Einrichtung: Univ. Grenoble Alpes, CNRS UMR 5525, VetAgro Sup, Grenoble INP, TIMC, 38000 Grenoble, France; Department of Orthopedics and Sport Surgery, Grenoble-A…. Hauptthemen laut MEDLINE: Osteoarthritis, Knee, Imaging, Three-Dimensional, Tomography, X-Ray Computed, Anatomic Landmarks. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Région Auvergne-Rhône-Alpes, Foundation for Medical Research, Centre Hospitalier, Université Grenoble Alpes.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

In preoperative planning, a number of techniques for registration of anatomical landmarks on bones, or "bones landmarking", have been developed for automatic measurement of lower limbs. However, the evaluation of these techniques is often neglected, reducing confidence in results and limiting clinical applications. We propose a protocol to evaluate landmarking techniques applied to thirty 3D images of osteoarthritic lower limbs. The protocol includes the evaluation of the ground truth based on manual landmarking performed by two experienced clinicians. Intra- and inter-operator variability were quantified. Automatic landmarking was evaluated using the same procedure and compared to manual landmarking by computing distances with bounding boxes including manual landmarks. The method was applied to 87 landmarks on the femur, tibia, fibula, and patella. For intra-operator variability, the mean median error was 2.0 mm, with a maximum of 42.1 mm. For inter-operator variability, errors ranged from 0.0 to 43.5 mm, with a mean median error of 2.3 mm across all landmarks. The mean median error between manual and automatic landmarking methods was 2.4 mm, with a range from 0.0 mm to 22.8 mm. Confidence intervals of automatic landmarking errors were consistent with intra- and inter-operator variability confidence intervals. Approximately 42 % of the automatic landmarks were inside their bounding boxes, with a mean distance of 1.4 mm from the box for landmarks on the outside. Results showed different errors depending on landmark location and coordinates. Intra- and inter-operator variability shed light on the difficulty of defining ground truth based on manual segmentation. This finding emphasizes the need for a rigorous evaluation protocol for landmarking validation, including the verification of statistical test hypotheses and population size influence.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Osteoarthritis, KneeImaging, Three-DimensionalTomography, X-Ray ComputedAnatomic LandmarksRadiographic Image Interpretation, Computer-AssistedPattern Recognition, AutomatedHumansReproducibility of ResultsSensitivity and SpecificityFemaleMaleAlgorithmsAgedMiddle AgedRadiographic Image Enhancement

Stichwörter der Autor:innen

Automatic registrationKneeLandmarkingLower limbStatistical evaluation

Autor:innen und Institutionen

  • Univ. Grenoble Alpes, CNRS UMR 5525, VetAgro Sup, Grenoble INP, TIMC, 38000 Grenoble, France; Department of Orthopedics and Sport Surgery, Grenoble-Alpes CHU South Teaching Hospital, Kimberley Avenue, BP 338, 38434 Échi…
  • Univ. Grenoble Alpes, CNRS UMR 5525, VetAgro Sup, Grenoble INP, TIMC, 38000 Grenoble, France.
  • Univ. Grenoble Alpes, CNRS UMR 5525, VetAgro Sup, Grenoble INP, TIMC, 38000 Grenoble, France; Twinsight, Grenoble, France.
  • Twinsight, Grenoble, France.
  • Univ. Grenoble Alpes, CNRS UMR 5525, VetAgro Sup, Grenoble INP, TIMC, 38000 Grenoble, France; Twinsight, Grenoble, France; Department of Orthopedics and Traumatology, Groupe Hospitalier Diaconesses Croix St-Simon, Paris…

Förderung

  • Région Auvergne-Rhône-Alpes
  • Foundation for Medical Research
  • Centre Hospitalier, Université Grenoble Alpes
  • Grand Équipement National De Calcul Intensif

Zitieren

APA

Horteur, C., Fougeron, N., Gaulin, B., Roch, Y., Bucki, M., Palluel, E., Payan, Y., & Perrier, A. (2025). Accuracy evaluation of an automatic landmarking method based on 3D segmented CT scans of full lower limbs with knee osteoarthritis. European Journal of Radiology, 191, 112292. https://doi.org/10.1016/j.ejrad.2025.112292

Vancouver

Horteur C, Fougeron N, Gaulin B, Roch Y, Bucki M, Palluel E, et al. Accuracy evaluation of an automatic landmarking method based on 3D segmented CT scans of full lower limbs with knee osteoarthritis. European Journal of Radiology. 2025;191:112292. doi:10.1016/j.ejrad.2025.112292

BibTeX
@article{Horteur2025,
  author = {Horteur, Clément and Fougeron, Nolwenn and Gaulin, Benoît and Roch, Ysé and Bucki, Marek and Palluel, Estelle and Payan, Yohan and Perrier, Antoine},
  title = {Accuracy evaluation of an automatic landmarking method based on 3D segmented CT scans of full lower limbs with knee osteoarthritis},
  journal = {European Journal of Radiology},
  year = {2025},
  volume = {191},
  pages = {112292},
  doi = {10.1016/j.ejrad.2025.112292}
}