A Simplex algorithm approach to determine the optimal keV and window setting for appendicitis visualization
European Journal of Radiology · 2025-10 · 191 · 112282
DOI: 10.1016/j.ejrad.2025.112282 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Radiology, 2025-10. Verfasst von Nevo Adam und 6 weiteren Autor:innen. Beteiligte Einrichtung: Hadassah Medical Center, Radiology Department, Israel. Electronic address: nevoad@post.bgu.ac.il. Hauptthemen laut MEDLINE: Appendicitis, Algorithms, Tomography, X-Ray Computed, Radiographic Image Interpretation, Computer-Assisted. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 29 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: Contrast-enhanced dual energy CT, with low-mono-energetic CT scanning (conventional 70 keV and low-mono energetic 40 keV) can improve visualization of acute inflammatory processes, but the high iodine attenuation often requires adjustment of window display settings to optimize interpretation. Here, we aim to determine optimal keV and window settings for the visualization of acute appendicitis from images derived from low-mono-energetic-CT. METHODS: 34 patients diagnosed with appendicitis, using dual energy CT, were retrospectively identified. An iterative four-dimensional Simplex optimization algorithm was used to determine optimal keV and windowing for assessment. 14 different keVs (40-70 range) and window parameters (center 5-290 HU; width 205-570 HU) were evaluated by 6 radiologists. Each reader individually graded the conspicuity of appendix wall enhancement using a 5-point scale and the overall image quality on a 3-point scale. Three-dimensional contour maps of the relationship between window settings and overall averages of reader grade at 40 keV were constructed. RESULTS: Convergence to optimal windowing was achieved over ten parameter iterations following the initial evaluation of 4 different parameter sets. This amounted to a total of 2,856 reads (14 keVs and window settings x 6 readers x 34 cases). The highest average conspicuity grade (4.4 ± 0.0) was achieved at 40 keV with windows of 57-79 HU center and 357-418 HU width, with best background in this range at window 57 HU and center 418 HU. CONCLUSIONS: Algorithmic optimization of dual energy contrast-enhanced CT scans may allow better visualization of acute inflammatory pathological processes such as mucosal enhancement in appendicitis by tailoring keV levels and window settings.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Hadassah Medical Center, Radiology Department, Israel. Electronic address: nevoad@post.bgu.ac.il.
- Hadassah Medical Center, Radiology Department, Israel. Electronic address: sgoldber@bidmc.harvard.edu.
- Hadassah Medical Center, Radiology Department, Israel. Electronic address: marron.daud01@ateneopv.it.
- Hadassah Medical Center, Radiology Department, Israel. Electronic address: Romman@hadassah.org.il.
- Hadassah Medical Center, Radiology Department, Israel. Electronic address: zeinehwissam@gmail.com.
- Hadassah Medical Center, Radiology Department, Israel. Electronic address: Jacobs@hadassah.org.il.
Zitieren
APA
Adam, N., Nahum, G. S., Marron, D., Romman, Z., Wissam, Z., Jacob, S., & Naama, L. C. (2025). A Simplex algorithm approach to determine the optimal keV and window setting for appendicitis visualization. European Journal of Radiology, 191, 112282. https://doi.org/10.1016/j.ejrad.2025.112282
Vancouver
Adam N, Nahum GS, Marron D, Romman Z, Wissam Z, Jacob S, et al. A Simplex algorithm approach to determine the optimal keV and window setting for appendicitis visualization. European Journal of Radiology. 2025;191:112282. doi:10.1016/j.ejrad.2025.112282
BibTeX
@article{Adam2025,