The effect of CT scanning and metal artefact reduction strategies on measuring knee implant displacement: iMAR – helpful or harmful?
European Journal of Radiology · 2025-10 · 191 · 112281
DOI: 10.1016/j.ejrad.2025.112281 ↗Kurz gefasst
Fachartikel in der Zeitschrift European Journal of Radiology, 2025-10. Verfasst von Maaike A Ter Wee und 6 weiteren Autor:innen. Beteiligte Einrichtung: Amsterdam UMC Location University of Amsterdam, Biomedical Engineering and Physics, Meibergdreef 9, Amsterdam, the Netherlands; Amsterdam Movement Sc…. Hauptthemen laut MEDLINE: Artifacts, Knee Prosthesis, Tomography, X-Ray Computed, Prosthesis Failure. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 47 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Amsterdam University Medical Centers.
Abstract (Originalsprache)
BACKGROUND: A recent development in diagnosing total knee arthroplasty (TKA) tibial component loosening uses Computed Tomography (CT) images of loaded knees to assess implant micro-motion, but metal artefacts can affect reliability. This study investigates the impact of increased exposure and tube voltage, iterative metal artefact reduction reconstruction (iMAR), Sn-filtering and an extended CT scale (ECTS), compared to a standard knee CT protocol on load-induced implant displacement measurement. METHOD: CT scans of a fresh frozen cadaver leg with a TKA were acquired using a standard protocol and four metal artefact reduction strategies: Increased exposure/tube voltage, iMAR, Sn-filtering, and ECTS reconstruction. Each scan was repeated 10 times with random leg repositioning ensuring no implant-bone motion. Image analysis included implant and proximal tibial cortex segmentation, object registration across scans, and quantification of apparent relative displacement. Grey-level similarity was assessed using the correlation coefficient (CC). Implant displacement error was measured using translation and rotation norms, mean target registration error (mTRE), and maximum total point motion (MTPM). RESULTS: CT scans with metal artefact reduction showed less noise and streaking. Grey-level similarity of the implant was similar across protocols (CC = 0.88). iMAR had the highest bone similarity (CC = 0.99). Implant displacement errors were comparable for increased exposure/tube voltage, Sn-filtering and ECTS (median: 0.08 mm translation, 0.13° rotation, 0.09 mm mTRE, 0.14 mm MTPM). However, iMAR statistically significantly increased median rotation norm to 0.19˚. CONCLUSION: Increased exposure/tube voltage, Sn-filtering, and ECTS do not impact the measurement error of 3D CT-based knee implant loosening detection. iMAR showed increased rotational error while other displacement parameters remained unchanged, suggesting it may be less suitable for quantifying implant displacement.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Amsterdam UMC Location University of Amsterdam, Biomedical Engineering and Physics, Meibergdreef 9, Amsterdam, the Netherlands; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, the Netherlands. Electronic…
- Amsterdam UMC Location University of Amsterdam, Biomedical Engineering and Physics, Meibergdreef 9, Amsterdam, the Netherlands.
- Amsterdam UMC Location University of Amsterdam, Biomedical Engineering and Physics, Meibergdreef 9, Amsterdam, the Netherlands; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, the Netherlands.
- Amsterdam UMC Location University of Amsterdam, Department of Orthopedic Surgery and Sports Medicine, Meibergdreef 9, Amsterdam, the Netherlands; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, the Nethe…
- Amsterdam UMC location University of Amsterdam, Radiology, Meibergdreef 9, Amsterdam, the Netherlands; Amsterdam Movement Sciences, Musculoskeletal Health, Amsterdam, the Netherlands.
Förderung
- Amsterdam University Medical Centers
Zitieren
APA
Ter Wee, M. A., Speelman, R. P. J., Dobbe, J. G. G., Kievit, A. J., Maas, M., Blankevoort, L., & Streekstra, G. J. (2025). The effect of CT scanning and metal artefact reduction strategies on measuring knee implant displacement: iMAR – helpful or harmful?. European Journal of Radiology, 191, 112281. https://doi.org/10.1016/j.ejrad.2025.112281
Vancouver
Ter Wee MA, Speelman RPJ, Dobbe JGG, Kievit AJ, Maas M, Blankevoort L, et al. The effect of CT scanning and metal artefact reduction strategies on measuring knee implant displacement: iMAR – helpful or harmful?. European Journal of Radiology. 2025;191:112281. doi:10.1016/j.ejrad.2025.112281
BibTeX
@article{TerWee2025,
author = {Ter Wee, Maaike A and Speelman, Rogier P J and Dobbe, Johannes G G and Kievit, Arthur J and Maas, Mario and Blankevoort, Leendert and Streekstra, Geert J},
title = {The effect of CT scanning and metal artefact reduction strategies on measuring knee implant displacement: iMAR – helpful or harmful?},
journal = {European Journal of Radiology},
year = {2025},
volume = {191},
pages = {112281},
doi = {10.1016/j.ejrad.2025.112281}
}
EUVIMED