Plaque and stent distinction using an experimental silicon-based photon-counting CT: An experimental and clinical comparison with conventional energy-integrated detector CT

Emma Verelst, Hakån Almqvist, Staffan Holmin, Gert Van Gompel, Hans Koenraad Nieboer, Frans Vandenbergh, Johan De Mey, Dominic Crotty, Paul Deak, Nico Buls

European Journal of Radiology · 2025

DOI: 10.1016/j.ejrad.2025.112207 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift European Journal of Radiology, 2025. Verfasst von Emma Verelst und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Vrije Universiteit Brussel, Universitair Ziekenhuis Brussel, 1090 Brussels, Belgium. Electronic address: emma.verelst@vub.be. Hauptthemen laut MEDLINE: Stents, Silicon, Tomography, X-Ray Computed, Plaque, Atherosclerotic. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Research Foundation Flanders.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To compare plaque-stent interface using an experimental Silicon-based photon-counting CT (Si-PCCT) with a conventional energy-integrating detector CT (EIDCT). METHODS: A human-resected and stented artery was embedded centrally into an agar-water phantom and scanned in a Si-PCCT and an EIDCT using a similar scan technique targeting a CT volumetric dose index (CTDIvol) of 9 mGy. Images were reconstructed using equivalent settings at 150 mm and 50 mm field-of-view (150FOV, 50FOV). Plaque-stent distinction, stent diameter measurement accuracy and blooming were quantitatively and qualitatively assessed using a 5-point Likert scale. Result comparisons between systems and FOVs were evaluated with a t-test and a Wilcoxon signed-rank test with inter-reader agreement assessed using intra-class correlation (ICC). Additionally, EIDCT and Si-PCCT images from a single clinical carotid angiography case were compared. RESULTS: Si-PCCT enabled improved plaque-stent contrast and reduced blooming (p < 0.001) at both FOVs, whereas diameter measurement accuracy was only improved at 50FOV (p = 0.003), compared to EIDCT. 150FOV Si-PCCT images were preferred over EIDCT images forstent appearance (p = 0.001), plaque appearance (p < 0.001), stent-plaque distinction (p = 0.002) and blooming (p = 0.001). 50FOV showed similar trends. 50FOV EIDCT images enabled improved plaque-stent contrast (p = 0.046), reduced blooming (p < 0.001), and improved stent appearance (p = 0.025). For Si-PCCT, no improvements were observed at 50FOV. Quantitative and qualitative ICC was 0.93 and 0.96, respectivelyThe experimental findings are reflected in the evaluation of the patient case. CONCLUSION: This study demonstrates the potential of Si-PCCT to improve image contrast and spatial resolution, enabling improved assessment of plaque and stent in an experimental phantom and a clinical patient case.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

StentsSiliconTomography, X-Ray ComputedPlaque, AtheroscleroticHumansPhantoms, ImagingPhotonsReproducibility of ResultsSensitivity and Specificity

Stichwörter der Autor:innen

Atherosclerotic plaquePhantom imagingPhoton-CountingStentsVascular imaging

Substanzen

Silicon

Autor:innen und Institutionen

  • Department of Radiology, Vrije Universiteit Brussel, Universitair Ziekenhuis Brussel, 1090 Brussels, Belgium. Electronic address: emma.verelst@vub.be.
  • Department of Clinical Neuroscience, Karolinska Institutet and Department of Neuroradiology, 171 74 Stockholm, Sweden; Department of Radiology, Capio St. Göran Hospital, 112 19 Stockholm, Sweden.
  • Department of Clinical Neuroscience, Karolinska Institutet and Department of Neuroradiology, 171 74 Stockholm, Sweden.
  • Department of Radiology, Vrije Universiteit Brussel, Universitair Ziekenhuis Brussel, 1090 Brussels, Belgium.
  • General Electric Healthcare, Waukesha, WI 53188 USA.

Förderung

  • Research Foundation Flanders

Zitieren

APA

Verelst, E., Almqvist, H., Holmin, S., Gompel, G. V., Nieboer, H. K., Vandenbergh, F., Mey, J., Crotty, D., Deak, P., & Buls, N. (2025). Plaque and stent distinction using an experimental silicon-based photon-counting CT: An experimental and clinical comparison with conventional energy-integrated detector CT. European Journal of Radiology. https://doi.org/10.1016/j.ejrad.2025.112207

Vancouver

Verelst E, Almqvist H, Holmin S, Gompel GV, Nieboer HK, Vandenbergh F, et al. Plaque and stent distinction using an experimental silicon-based photon-counting CT: An experimental and clinical comparison with conventional energy-integrated detector CT. European Journal of Radiology. 2025. doi:10.1016/j.ejrad.2025.112207

BibTeX
@article{Verelst2025,
  author = {Verelst, Emma and Almqvist, Hakån and Holmin, Staffan and Gompel, Gert Van and Nieboer, Hans Koenraad and Vandenbergh, Frans and Mey, Johan De and Crotty, Dominic and Deak, Paul and Buls, Nico},
  title = {Plaque and stent distinction using an experimental silicon-based photon-counting CT: An experimental and clinical comparison with conventional energy-integrated detector CT},
  journal = {European Journal of Radiology},
  year = {2025},
  doi = {10.1016/j.ejrad.2025.112207}
}
Kostenlos starten →