Calibration of X-ray multimeters for radiography

Paula Toroi, Viivi Valkama, Touko Kaasalainen, Niko Kiiskinen, Mika Kortesniemi, Panu Pousi, Jaroslav Šolc, Aino Tietäväinen

Physica Medica · 2026-12 · 152 · 107239

DOI: 10.1016/j.ejmp.2026.107239 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physica Medica, 2026-12. Verfasst von Paula Toroi und 7 weiteren Autor:innen. Beteiligte Einrichtung: STUK - Radiation and Nuclear Safety Authority, Vantaa, Finland; HUS Diagnostic Center, Radiology, Helsinki University and Helsinki University Hospita…. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Partnership on Metrology.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

INTRODUCTION: Accurate determination of air kerma, tube‑voltage‑related quantities (TV), and half‑value layer (HVL) is essential for reliable dosimetry in X‑ray imaging. X‑ray multimeters (XMMs) are widely used for these measurements; however, their performance depends on spectral characteristics and measurement conditions. This study evaluates the consistency and clinical applicability of calibration coefficients derived using industrial and clinical X‑ray systems across a range of clinically relevant radiation qualities. METHODS: Eight XMMs from four manufacturers were calibrated for air kerma, TV, and HVL using one laboratory and three clinical systems. Reference values were established using traceable ionization chambers, high voltage dividers, and spectrometry. Calibration coefficients were assessed for standard aluminium filtered radiation qualities and for an extended range including 0-0.9 mm of copper filtration. Energy dependence, X‑ray system dependence, and the impact of applying calibration coefficients were evaluated. RESULTS: For air kerma, calibration coefficients derived using industrial and clinical systems were highly consistent. Energy dependence variations were within 5% and could be reduced to below 1% by applying radiation‑quality‑specific calibration coefficients. TV measurements showed the strongest dependence on the X‑ray system; energy dependence variations were generally below 4% and could be improved only using clinical‑system‑based calibrations. For HVL measurements, energy dependence variations were typically within 6% and were reduced below 1% by calibration, except for two XMMs exhibiting system‑dependent behaviour. CONCLUSIONS: XMMs can provide reliable measurements across clinically relevant radiation qualities, including copper‑filtered beams, when radiation‑quality‑specific calibration coefficients are applied. Interpolation based on a single radiation quality parameter is insufficient; at least two parameters are required.

Schlagwörter

Stichwörter der Autor:innen

AccuracyCalibrationDosimetryEnergy dependenceRadiation qualityTraceabilityUncertaintyX-ray multimeterXMM

Autor:innen und Institutionen

  • STUK - Radiation and Nuclear Safety Authority, Vantaa, Finland; HUS Diagnostic Center, Radiology, Helsinki University and Helsinki University Hospital, Helsinki, Finland. Electronic address: paula.toroi@stuk.fi.
  • STUK - Radiation and Nuclear Safety Authority, Vantaa, Finland.
  • HUS Diagnostic Center, Radiology, Helsinki University and Helsinki University Hospital, Helsinki, Finland.
  • CMI - Czech Metrology Institute, Okruzni 31, Brno, Czech Republic.

Förderung

  • European Partnership on Metrology

Zitieren

APA

Toroi, P., Valkama, V., Kaasalainen, T., Kiiskinen, N., Kortesniemi, M., Pousi, P., Šolc, J., & Tietäväinen, A. (2026). Calibration of X-ray multimeters for radiography. Physica Medica, 152, 107239. https://doi.org/10.1016/j.ejmp.2026.107239

Vancouver

Toroi P, Valkama V, Kaasalainen T, Kiiskinen N, Kortesniemi M, Pousi P, et al. Calibration of X-ray multimeters for radiography. Physica Medica. 2026;152:107239. doi:10.1016/j.ejmp.2026.107239

BibTeX
@article{Toroi2026,
  author = {Toroi, Paula and Valkama, Viivi and Kaasalainen, Touko and Kiiskinen, Niko and Kortesniemi, Mika and Pousi, Panu and Šolc, Jaroslav and Tietäväinen, Aino},
  title = {Calibration of X-ray multimeters for radiography},
  journal = {Physica Medica},
  year = {2026},
  volume = {152},
  pages = {107239},
  doi = {10.1016/j.ejmp.2026.107239}
}
Kostenlos starten →