Retrospective analysis of treatment-positioning accuracy and dose error in boron neutron capture therapy using a sitting-position treatment system for head and neck cancer
Physica Medica · 2024
DOI: 10.1016/j.ejmp.2024.104818 ↗Kurz gefasst
Fachartikel in der Zeitschrift Physica Medica, 2024. Verfasst von Shinya Komori und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Physics and Technology, Southern Tohoku BNCT Research Center, 7-10 Yatsuyamada, Koriyama, Fukushima 963-8052, Japan. Electron…. Hauptthemen laut MEDLINE: Boron Neutron Capture Therapy, Head and Neck Neoplasms, Radiotherapy Dosage, Radiotherapy Planning, Computer-Assisted. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 46 Literaturangaben.
Abstract (Originalsprache)
The neutron beam in boron neutron capture therapy (BNCT) exhibits poor directionality and significantly decreasing neutron flux with increasing distance. Therefore, the treatment site must be close to the irradiation aperture. Some patients with head and neck cancer may benefit from a sitting-position setup. The study aim was to evaluate the treatment-positioning accuracy and dose error in sitting patients receiving BNCT. Thirty-two patients with head and neck cancer who underwent sitting-position BNCT at Southern Tohoku BNCT Research Center were included in the study. Horizontal (ΔX) and vertical (ΔY) errors were defined as the displacement between the treatment planning system (TPS) digital reconstructed radiograph and the pre-treatment X-ray image. Using in-house software, image matching was performed. The beam-axial directional (ΔZ) error was compared with the parameters entered into the TPS and the actual pre-treatment measured values. The translational-position error was reflected in the TPS's patient coordinate system with respect to the reference plan. Re-dose calculations were performed to evaluate the effect of positional error on tumor and normal-tissue doses. The [ΔX, ΔY, ΔZ] DRR-CR mean ± 1SD were - 0.40 ± 2.0, 0.30 ± 2.3, and - 1.4 ± 1.5 mm, respectively. The Dmean and D98% tumor-dose errors were 1.22 % ± 1.44 % and 0.99 % ± 1.63 %, respectively. The D2% pharyngeal and oral mucosal-dose errors were 0.98 % ± 1.91 % and 1.21 % ± 1.78 %, respectively. The tumor- and normal-tissue dose errors were typically < 5 %. High-precision treatment was feasible in sitting-positioned BNCT.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Physics and Technology, Southern Tohoku BNCT Research Center, 7-10 Yatsuyamada, Koriyama, Fukushima 963-8052, Japan. Electronic address: skomori.0922@gmail.com.
- Department of Radiation Oncology, Hirosaki University Graduate School of Medicine, 5 Zaifu-cho, Hirosaki, Aomori, 036-8562, Japan.
- Department of Radiation Physics and Technology, Southern Tohoku BNCT Research Center, 7-10 Yatsuyamada, Koriyama, Fukushima 963-8052, Japan.
- School of Health Sciences, Fukushima Medical University, 10-6 Sakaemachi, Fukushima, Fukushima, 960-8516, Japan.
- Department of Radiation Oncology, Southern Tohoku BNCT Research Center, 7-10 Yatsuyamada, Koriyama, Fukushima, 963-8052, Japan.
Zitieren
APA
Komori, S., Hirose, K., Sato, M., Yamazaki, Y., Takeuchi, A., Kato, R., Motoyanagi, T., Narita, Y., Kato, T., & Takai, Y. (2024). Retrospective analysis of treatment-positioning accuracy and dose error in boron neutron capture therapy using a sitting-position treatment system for head and neck cancer. Physica Medica. https://doi.org/10.1016/j.ejmp.2024.104818
Vancouver
Komori S, Hirose K, Sato M, Yamazaki Y, Takeuchi A, Kato R, et al. Retrospective analysis of treatment-positioning accuracy and dose error in boron neutron capture therapy using a sitting-position treatment system for head and neck cancer. Physica Medica. 2024. doi:10.1016/j.ejmp.2024.104818
BibTeX
@article{Komori2024,
author = {Komori, Shinya and Hirose, Katsumi and Sato, Mariko and Yamazaki, Yuhei and Takeuchi, Akihiko and Kato, Ryohei and Motoyanagi, Tomoaki and Narita, Yuki and Kato, Takahiro and Takai, Yoshihiro},
title = {Retrospective analysis of treatment-positioning accuracy and dose error in boron neutron capture therapy using a sitting-position treatment system for head and neck cancer},
journal = {Physica Medica},
year = {2024},
doi = {10.1016/j.ejmp.2024.104818}
}
EUVIMED