Deep learning-based image reconstruction significantly improves image quality of MRI examinations of the orbit at 3 Tesla
Diagnostic and Interventional Imaging · 2026
DOI: 10.1016/j.diii.2025.11.003 ↗Kurz gefasst
Fachartikel in der Zeitschrift Diagnostic and Interventional Imaging, 2026. Verfasst von Aurore Sajust de Bergues de Escalup und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Neuroradiology, Hôpital Fondation Adolphe de Rothschild, 75019 Paris, France. Electronic address: asajustdebergues@for.paris. Hauptthemen laut MEDLINE: Deep Learning, Magnetic Resonance Imaging, Orbit, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 26 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: The purpose of this study was to assess the benefit of a deep learning-based image reconstruction (DLBIR) for improving image quality in orbital magnetic resonance imaging (MRI) at 3 Tesla (T). MATERIALS AND METHODS: Seventy-one patients (48 women and 23 men) with a mean age of 52 ± 19.5 (standard deviation [SD]) years (age range: 7-90 years) who underwent MRI examination of the orbit at 3 T between January and June of 2024, were included in the study. Coronal T2-weighted MR images obtained in 70 patients and post-contrast fat-saturated (FS) coronal T1-weighted MR images obtained in 25 patients, were reconstructed with and without DLBIR, resulting in four imaging sets. Two radiologists independently and blindly measured the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) of the optic nerves on the four imaging sets. Image quality and orbital abnormalities were assessed using a standardized 5-point Likert scale. Comparisons between MR images obtained with and without DLBIR were performed using Wilcoxon test for ordinal and quantitative variables and McNemar test for paired binary data. RESULTS: SNR and CNR of coronal T2-weighted MR images were significantly greater using DLBIR (26.67 ± 9.03 [SD], and 14.87 ± 10.31 [SD], respectively) than without DLBIR (18.91 ± 7.28 [SD], and 9.78 ± 8.47 [SD], respectively) (P < 0.001). There were no differences in SNR and CNR between post-contrast FS T1-weighted images obtained with DLBIR (85.56 ± 63.13 [SD], and 64 ± 41.38 [SD], respectively) and those obtained without DLBIR (91.36 ± 48.49 [SD], and 43.25 ± 20.4 [SD], respectively) (P = 0.35, and P = 0.14, respectively). Qualitatively, good-to-excellent image quality was obtained more frequently with DLBIR than without DLBIR for T2-weighted and post-contrast FS T1-weighted images with respect to optic nerve sharpness (67 % vs. 16 %, and 8 % vs. 0 %, respectively), brain sharpness (90 % vs. 6 %, and 68 % vs. 4 %, respectively), and overall image quality (73 % vs. 1 % and 36 % vs. 0 %, respectively) (all P ≤ 0.001). No significant differences in the detection rates of orbital abnormalities were found between MR images obtained with and without DLBIR, including optic nerve hyperintensity (34 % vs. 31 %, respectively; P = 0.16) and optic nerve atrophy (33 % for both) on T2-weighted images, and optic nerve enhancement on post-contrast FS T1-weighted images (16 % for both). CONCLUSION: DLBIR significantly improves image quality of MRI examinations of the orbit at 3 T, without losing clinically relevant information.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Neuroradiology, Hôpital Fondation Adolphe de Rothschild, 75019 Paris, France. Electronic address: asajustdebergues@for.paris.
- Department of Neuroradiology, Hôpital Fondation Adolphe de Rothschild, 75019 Paris, France; Université Paris Cité, Faculté de Médecine, 75006 Paris, France.
- Department of Neuroradiology, Hôpital Fondation Adolphe de Rothschild, 75019 Paris, France.
- Department of Neurology, Hôpital Fondation Adolphe de Rothschild, 75019 Paris, France.
- Department of Neuro-Ophthalmology, Hôpital Fondation Adolphe de Rothschild, 75019 Paris, France.
Zitieren
APA
Sajust de Bergues de Escalup, A., Lecler, A., Poirion, É., Papeix, C., Deschamps, R., Milea, D., Savatovsky, J., Duron, L., & O'Shaughnessy, E. (2026). Deep learning-based image reconstruction significantly improves image quality of MRI examinations of the orbit at 3 Tesla. Diagnostic and Interventional Imaging. https://doi.org/10.1016/j.diii.2025.11.003
Vancouver
Sajust de Bergues de Escalup A, Lecler A, Poirion É, Papeix C, Deschamps R, Milea D, et al. Deep learning-based image reconstruction significantly improves image quality of MRI examinations of the orbit at 3 Tesla. Diagnostic and Interventional Imaging. 2026. doi:10.1016/j.diii.2025.11.003
BibTeX
@article{SajustdeBerguesdeEscalup2026,
author = {Sajust de Bergues de Escalup, Aurore and Lecler, Augustin and Poirion, Émilie and Papeix, Caroline and Deschamps, Romain and Milea, Dan and Savatovsky, Julien and Duron, Loïc and O'Shaughnessy, Emma},
title = {Deep learning-based image reconstruction significantly improves image quality of MRI examinations of the orbit at 3 Tesla},
journal = {Diagnostic and Interventional Imaging},
year = {2026},
doi = {10.1016/j.diii.2025.11.003}
}
EUVIMED