Reduction of radiation dose and contrast medium volume in computed tomography pulmonary angiography: adaptation of dual-energy computed tomography (CT) protocols to the body mass index
Clinical Radiology · 2025
DOI: 10.1016/j.crad.2025.106944 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Clinical Radiology, 2025. Verfasst von X Zhang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Southern University of Science and Technology Hospital, Shenzhen, Guangdong, China. Hauptthemen laut MEDLINE: Contrast Media, Pulmonary Embolism, Radiation Dosage, Computed Tomography Angiography. Die Arbeit nennt 40 Literaturangaben.
Abstract (Originalsprache)
AIM: To evaluate the image quality and diagnostic value of a body mass index (BMI)-based dual-energy computed tomography (CT) protocol in computed tomography pulmonary angiography (CTPA) for reduction of radiation dose and contrast medium (CM) volume. MATERIALS AND METHODS: Patients suspected of having pulmonary embolism were prospectively included and randomly assigned to one of three protocols: Protocol A (a standard protocol, 100 kV/140 mAs/50 mL iodinated CM), B (a conventional dual-energy protocol, 80-140 kV switching/automatic tube current/weight-dependent CM volume) and C (a BMI-based dual-energy protocol, 80-140 kV switching/BMI-based tube current/weight-dependent CM volume, BMI-based CM injection rate). Subjective and objective image analysis were performed by two radiologists independently. Diagnostic accuracy of pulmonary embolism were evaluated. Inter-group comparison was performed. RESULTS: Ninety patients (mean age 54.1 ± 13.3 years, 57 men) were included, with 30 patients in each group. The CT values, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) of the pulmonary arteries in group C were significantly higher than those in group B (358.2 ± 50.5 vs 324.9 ± 57.2, p=0.047, 17.8 ± 3.2 vs 15.3 ± 2.9, p=0.010, 19.2 ± 4.1 vs 15.7 ± 3.8, p=0.014). The dose length product (DLP) and CM volume were significantly reduced in groups B and C (A vs B vs C, DLP, 287.9 ± 34.0 vs 177.2 ± 39.2 vs 183.8 ± 23.1 mGy·cm), p
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Radiology, Southern University of Science and Technology Hospital, Shenzhen, Guangdong, China.
- Department of Radiology, Southern University of Science and Technology Hospital, Shenzhen, Guangdong, China. Electronic address: chenmin@bjmu.edu.cn.
- Department of Radiology, Affiliated Hospital of Guangdong Medical University, Zhanjiang, Guangdong, China. Electronic address: junqingxu@126.com.
Zitieren
APA
Zhang, X., Huang, H., Huang, Y., Sun, Y., Chen, Z., Chen, M., & Xu, J. (2025). Reduction of radiation dose and contrast medium volume in computed tomography pulmonary angiography: adaptation of dual-energy computed tomography (CT) protocols to the body mass index. Clinical Radiology. https://doi.org/10.1016/j.crad.2025.106944
Vancouver
Zhang X, Huang H, Huang Y, Sun Y, Chen Z, Chen M, et al. Reduction of radiation dose and contrast medium volume in computed tomography pulmonary angiography: adaptation of dual-energy computed tomography (CT) protocols to the body mass index. Clinical Radiology. 2025. doi:10.1016/j.crad.2025.106944
BibTeX
@article{Zhang2025,
author = {Zhang, X and Huang, H and Huang, Y and Sun, Y and Chen, Z and Chen, M and Xu, J},
title = {Reduction of radiation dose and contrast medium volume in computed tomography pulmonary angiography: adaptation of dual-energy computed tomography (CT) protocols to the body mass index},
journal = {Clinical Radiology},
year = {2025},
doi = {10.1016/j.crad.2025.106944}
}
EUVIMED