A comparison of two cold atmospheric helium plasma devices which utilise the same RF power generator
Clinical Plasma Medicine · 2020-09 · 19-20 · 100108
DOI: 10.1016/j.cpme.2020.100108 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Plasma Medicine, 2020-09. Verfasst von Cian D. Madigan und 3 weiteren Autor:innen. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Irish Research Council.
Abstract (Originalsprache)
This research was conducted to analyse and compare a Renuvion hand piece and TheraDep's deposition device at various gas flow rates and power settings. Both devices were powered by an Apyx Ultimate Electrosurgical generator. Current, voltage, optical emission spectroscopy (OES) and thermal imaging were used to investigate the resulting plasma intensities and species being ionised. It was found that the Renuvion device could deliver 5.6 to 8.2 W of power whereas the Deposition Accessory was significantly less at between 1.9 and 4.2 W depending on the power setting selected. Differences in the OES results were also observed which are thought to have resulted from variations in the atmospheric entrainment between the two devices. Taken in combination, the electrical, optical emission and thermal data analysis has demonstrated that the addition of the Deposition Accessory transforms the Renuvion electrosurgical tool from a device that operates in the coagulation and cutting arena into a device with a dramatic reduction in the energy delivered to the target tissue. Consequently, this limits the localised heating of adjacent tissue and results in a device that has controlled coagulation capability and one which is not capable of cutting tissue. This offers significant potential to allow clinicians to evaluate applications in plasma medicine using an FDA cleared electrosurgical power source.
Förderung
- Irish Research Council
Zitieren
APA
Madigan, C. D., O'Sullivan, D., O'Neill, L., & Kavanagh, D. F. (2020). A comparison of two cold atmospheric helium plasma devices which utilise the same RF power generator. Clinical Plasma Medicine, 19-20, 100108. https://doi.org/10.1016/j.cpme.2020.100108
Vancouver
Madigan CD, O'Sullivan D, O'Neill L, Kavanagh DF. A comparison of two cold atmospheric helium plasma devices which utilise the same RF power generator. Clinical Plasma Medicine. 2020;19-20:100108. doi:10.1016/j.cpme.2020.100108
BibTeX
@article{Madigan2020,
author = {Madigan, Cian D. and O'Sullivan, Denis and O'Neill, Liam and Kavanagh, Darren F.},
title = {A comparison of two cold atmospheric helium plasma devices which utilise the same RF power generator},
journal = {Clinical Plasma Medicine},
year = {2020},
volume = {19-20},
pages = {100108},
doi = {10.1016/j.cpme.2020.100108}
}