Strength-duration time constant and rheobase measurements: Comparison of the threshold tracking method and a manual procedure
Clinical Neurophysiology · 2023
DOI: 10.1016/j.clinph.2023.06.026 ↗Kurz gefasst
Fachartikel in der Zeitschrift Clinical Neurophysiology, 2023. Verfasst von Maelle Tyberghein und 2 weiteren Autor:innen. Beteiligte Einrichtung: CHU Liège, Department of Neurophysiology, Sart Tilman B35, 4000 Liège, Belgium. Hauptthemen laut MEDLINE: Peripheral Nerves, Median Nerve. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 25 Literaturangaben.
Abstract (Originalsprache)
OBJECTIVE: To compare the strength-duration time constant (SDTC) and rheobase measurements obtained by the threshold tracking method (TT) and by a non-automated method (MM). METHODS: The MM procedure involved measuring, using a routine electrodiagnostic device, the intensity required to evoke a motor response whose amplitude corresponds to 40% of the maximum amplitude for four stimulus duration (1.0, 0.7, 0.5, 0.2 ms), and studying the linear relationship between stimulus charge and stimulus duration (slope = rheobase, intercept on the x-axis = SDTC). Using TT and MM, 30 successive healthy subjects (mean age = 38 years old) underwent a prospective evaluation of SDTC and rheobase of the median nerve motor axons at the wrist. Nerve stimulation and bipolar recording of evoked motor responses were performed with disposable self-adhesive surface electrodes. RESULTS: The Spearman correlations between the two methods were 0.78 (p < 0.0001) for SDTC and 0.96 (p < 0.0001) for the rheobase. The Bland-Altman analysis did not reveal any systematic bias of MM compared to TT. CONCLUSIONS: The MM procedure was reliable for strength-duration relationship analysis. SIGNIFICANCE: We encourage neurophysiologists, who do not have dedicated threshold tracking equipment, not to hesitate to use these simple tools to assess peripheral nerve excitability.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- CHU Liège, Department of Neurophysiology, Sart Tilman B35, 4000 Liège, Belgium.
- CHU Liège, Department of Neurophysiology, Sart Tilman B35, 4000 Liège, Belgium. Electronic address: fc.wang@chuliege.be.
Zitieren
APA
Tyberghein, M., Janssen, A., & Wang, F. C. (2023). Strength-duration time constant and rheobase measurements: Comparison of the threshold tracking method and a manual procedure. Clinical Neurophysiology. https://doi.org/10.1016/j.clinph.2023.06.026
Vancouver
Tyberghein M, Janssen A, Wang FC. Strength-duration time constant and rheobase measurements: Comparison of the threshold tracking method and a manual procedure. Clinical Neurophysiology. 2023. doi:10.1016/j.clinph.2023.06.026