Fachartikel in der Zeitschrift Clinical Mass Spectrometry, 2016. Verfasst von Adam P R Zabell und 2 weiteren Autor:innen. Beteiligte Einrichtung: Indigo BioAutomation, 7820 Innovation Blvd, Suite 250, Indianapolis, IN 46278, United States. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Instrument calibration, required for any accurate quantitative calculation, is a trivial process when performed correctly, but is also full of easily overlooked stumbling blocks. To minimize the risk of error associated with improper calibrations, national and international guidance dictates a minimum number of calibrators and the threshold at which a measurement becomes an outlier. Evidence from industry practice, which conflicts with regulatory guidance, suggests that most groups are focused on remapping their detector with each run. We present a post facto explanation for the calibrator minimum and provide recommendations for curve building, which include improved outlier detection for high-volume mass spectrometry laboratories.

Schlagwörter

Stichwörter der Autor:innen

CalibrationInstrument stabilityOutlier detectionStudentized deleted residual

Autor:innen und Institutionen

  • Indigo BioAutomation, 7820 Innovation Blvd, Suite 250, Indianapolis, IN 46278, United States.

Zitieren

APA

Zabell, A. P. R., Lytle, F. E., & Julian, R. K. (2016). A proposal to improve calibration and outlier detection in high-throughput mass spectrometry. Clinical Mass Spectrometry. https://doi.org/10.1016/j.clinms.2016.12.003

Vancouver

Zabell APR, Lytle FE, Julian RK. A proposal to improve calibration and outlier detection in high-throughput mass spectrometry. Clinical Mass Spectrometry. 2016. doi:10.1016/j.clinms.2016.12.003

BibTeX
@article{Zabell2016,
  author = {Zabell, Adam P R and Lytle, Fred E and Julian, Randall K},
  title = {A proposal to improve calibration and outlier detection in high-throughput mass spectrometry},
  journal = {Clinical Mass Spectrometry},
  year = {2016},
  doi = {10.1016/j.clinms.2016.12.003}
}
Kostenlos starten →