Long-term effects of tear film component deposition on the surface and optical properties of two different orthokeratology lenses

Chen-Ying Su, Lung-Kun Yeh, Pin-Hsuan Huang, Wen-Pin Lin, Hsueh-Fang Huang, Chi-Chun Lai, Hsu-Wei Fang

Contact Lens and Anterior Eye · 2023

DOI: 10.1016/j.clae.2023.101852 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Contact Lens and Anterior Eye, 2023. Verfasst von Chen-Ying Su und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Chemical Engineering and Biotechnology, National Taipei University of Technology. 1, Sec. 3, Zhongxiao E. Rd., Taipei 10608, Taiwan. Hauptthemen laut MEDLINE: Contact Lenses, Hydrophilic. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 36 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To understand the effects of long-term deposition of tear film components on the surface and optical properties of orthokeratology (ortho-k) lenses, two different lenses, Brighten 22 and Optimum Extra, were tested here. METHODS: Ortho-k lenses were immersed in artificial tears and cleaned with a commercial care solution repeatedly for up to 90 days. Both the daily and accumulated lysozyme deposition amounts using an Enzyme-Linked ImmunoSorbent Assay were then analyzed. The base curve, central thickness, power, and transmission of visible light, ultraviolet A, and ultraviolet B were analyzed before and after repeated tear film component deposition procedures. The surface roughness using atomic force microscopy was observed and an energy dispersive spectrometer was used to analyze the composition of the deposits. RESULTS: The highest levels of lysozyme were adsorbed on both lens materials during the first four days of the procedure and became saturated by day 6. For both lens materials, contamination on the lenses was easily observed by day 30, and the degree of surface roughness was higher. The transmission levels of different light spectrums were reduced showing that the optical characteristics of both lenses were also affected. CONCLUSIONS: The results provide in vitro evidence that lysozyme could not be completely removed from orthokeratology lenses. Both surface and optical properties were affected by the deposition of tear film components. However, only one commercial multipurpose care solution was used to clean the lens in this study when the main ingredient was a surfactant, and the results might be different when other care regimens with other key ingredients are used. In addition, whether tear film component deposition might result in increased risks of infection or corneal abrasion will require further investigation.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Contact Lenses, HydrophilicHumansTearsLubricant Eye Drops

Stichwörter der Autor:innen

LysozymeOptical propertyOrthokeratology lensSurface roughnessTear film components

Substanzen

Lubricant Eye Drops

Autor:innen und Institutionen

  • Department of Chemical Engineering and Biotechnology, National Taipei University of Technology. 1, Sec. 3, Zhongxiao E. Rd., Taipei 10608, Taiwan.
  • Department of Ophthalmology, Chang Gung Memorial Hospital, Linkou, No. 5, Fuxing St., Taoyuan 333, Taiwan; College of Medicine, Chang Gung University, No. 259, Wenhua 1st Rd., Taoyuan 333, Taiwan.
  • Research and Development Center, Brighten Optix Co., 6F-1, No. 150, Sec.4, Chengde Rd., Shilin Dist., Taipei 111, Taiwan; Department of Optometry, University of Kang Ning, No. 137, Alley 75, Sec. 3, Kang Ning Road, Neih…
  • Research and Development Center, Brighten Optix Co., 6F-1, No. 150, Sec.4, Chengde Rd., Shilin Dist., Taipei 111, Taiwan.
  • Department of Chemical Engineering and Biotechnology, National Taipei University of Technology. 1, Sec. 3, Zhongxiao E. Rd., Taipei 10608, Taiwan; Institute of Biomedical Engineering and Nanomedicine, National Health Re…

Zitieren

APA

Su, C. Y., Yeh, L. K., Huang, P. H., Lin, W. P., Huang, H. F., Lai, C. C., & Fang, H. W. (2023). Long-term effects of tear film component deposition on the surface and optical properties of two different orthokeratology lenses. Contact Lens and Anterior Eye. https://doi.org/10.1016/j.clae.2023.101852

Vancouver

Su CY, Yeh LK, Huang PH, Lin WP, Huang HF, Lai CC, et al. Long-term effects of tear film component deposition on the surface and optical properties of two different orthokeratology lenses. Contact Lens and Anterior Eye. 2023. doi:10.1016/j.clae.2023.101852

BibTeX
@article{Su2023,
  author = {Su, Chen-Ying and Yeh, Lung-Kun and Huang, Pin-Hsuan and Lin, Wen-Pin and Huang, Hsueh-Fang and Lai, Chi-Chun and Fang, Hsu-Wei},
  title = {Long-term effects of tear film component deposition on the surface and optical properties of two different orthokeratology lenses},
  journal = {Contact Lens and Anterior Eye},
  year = {2023},
  doi = {10.1016/j.clae.2023.101852}
}
Kostenlos starten →