Carotid Artery Corrected Flow Time Measured by Wearable Doppler Ultrasound Detects Stroke Volume Change Measured by Transesophageal Echocardiography After Coronary Artery Bypass Grafting
CHEST Critical Care · 2025-06 · 3 (2) · 100138
DOI: 10.1016/j.chstcc.2025.100138 ↗Kurz gefasst
Fachartikel in der Zeitschrift CHEST Critical Care, 2025-06. Verfasst von Jon-Emile S. Kenny und 11 weiteren Autor:innen. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 50 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Northern Ontario Academic Medicine Association.
Themen: Schlaganfall
Abstract (Originalsprache)
Background
As a measure of preload responsiveness (PR), change in carotid artery corrected flow time (ccFTΔ) is a surrogate for change in stroke volume (SVΔ). However, the optimal threshold and accuracy of ccFTΔ to detect SVΔ are inconsistent in previous reports. Research Question: Does ccFTΔ from a wireless, wearable Doppler ultrasound accurately detect a 10% SVΔ measured by transesophageal echocardiography? Study Design and Methods: This was a prospective, single-center study of adult patients after elective coronary artery bypass grafting. PR was defined as ≥ 10% augmentation in transesophageal echocardiography left ventricular outflow tract velocity time integral (as a surrogate for SVΔ) during Trendelenburg positioning. Synchronous carotid Doppler imaging was captured by a wireless, wearable Doppler ultrasound. The optimal ccFTΔ threshold to detect PR, sensitivity, specificity, and area under the receiver operating characteristic curve (AUC) were calculated. Linear correlation between ccFTΔ and SVΔ was assessed by Pearson correlation coefficient. We also evaluated the effect of the number of consecutively averaged cardiac cycles on ccFTΔ accuracy.
Results
This analysis included 30 patients; 7 patients showed a ≥ 10% SVΔ during Trendelenburg positioning. The optimal ccFTΔ thresholds were +6.6 ms or 2.2% with sensitivities of 100%, specificities of 70%, and AUCs of 0.89 and 0.88, respectively. A strong, linear correlation between ccFTΔ and SVΔ was found (r = 0.70; P < .001). The mean AUC increased from 0.68 to 0.87 when using 1 vs 20 consecutively averaged cardiac cycles.
Interpretation
After cardiopulmonary bypass, ccFTΔ measured by wireless, wearable ultrasound detected SVΔ during Trendelenburg positioning with high accuracy. The AUC improved as a function of consecutively averaged cardiac cycles. As a surrogate for preload-induced SVΔ, ccFTΔ can direct fluid therapy in the postoperative period.
Förderung
- Northern Ontario Academic Medicine Association
Zitieren
APA
Kenny, J. S., Clarke, G., Atwi, S., Kerrebijn, I., Savery, T., Knott, M., Munding, C. E., Elfarnawany, M., Eibl, A. M., Eibl, J. K., Nalla, B., & Atoui, R. (2025). Carotid Artery Corrected Flow Time Measured by Wearable Doppler Ultrasound Detects Stroke Volume Change Measured by Transesophageal Echocardiography After Coronary Artery Bypass Grafting. CHEST Critical Care, 3(2), 100138. https://doi.org/10.1016/j.chstcc.2025.100138
Vancouver
Kenny JS, Clarke G, Atwi S, Kerrebijn I, Savery T, Knott M, et al. Carotid Artery Corrected Flow Time Measured by Wearable Doppler Ultrasound Detects Stroke Volume Change Measured by Transesophageal Echocardiography After Coronary Artery Bypass Grafting. CHEST Critical Care. 2025;3(2):100138. doi:10.1016/j.chstcc.2025.100138
BibTeX
@article{Kenny2025,
author = {Kenny, Jon-Emile S. and Clarke, Geoffrey and Atwi, Sarah and Kerrebijn, Isabel and Savery, Tracy and Knott, Meredith and Munding, Chelsea E. and Elfarnawany, Mai and Eibl, Andrew M. and Eibl, Joseph K. and Nalla, Bhanu and Atoui, Rony},
title = {Carotid Artery Corrected Flow Time Measured by Wearable Doppler Ultrasound Detects Stroke Volume Change Measured by Transesophageal Echocardiography After Coronary Artery Bypass Grafting},
journal = {CHEST Critical Care},
year = {2025},
volume = {3},
number = {2},
pages = {100138},
doi = {10.1016/j.chstcc.2025.100138}
}
EUVIMED