Performance comparison of UniSprayTM and electrospray ionisation coupled to ion-mobility high-resolution mass spectrometer for screening of organic pollutants in water matrices

Victoria Eriksson, Kalpana Munnuru Singamshetty, Thomas Malte Molnár Karlsson, Selina Tisler, Jan H Christensen, Johan Dahlén, Foon Yin Lai, Thanh Wang

Journal of Chromatography A · 2026-12 · 1790 · 467569

DOI: 10.1016/j.chroma.2026.467569 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of Chromatography A, 2026-12. Verfasst von Victoria Eriksson und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, Chemistry, and Biology (IFM), Linköping University, SE-581 83, Linköping, Sweden. Electronic address: victoria.eriksson@liu.se. Hauptthemen laut MEDLINE: Water Pollutants, Chemical, Spectrometry, Mass, Electrospray Ionization, Liquid Chromatography-Mass Spectrometry, Organic Chemicals. Die Arbeit nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Lars Hierta Memorial Foundation, Horizon Europe, NoVo Foundation.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Suspect and non-target screening (SNTS) by high-resolution mass spectrometry is widely used to characterise complex environmental mixtures, but detection depends strongly on ionisation efficiency and matrix effects. UniSprayTM ionisation (USI) is a relatively new ion source that applies a voltage to a metal rod positioned after a grounded spray capillary in contrast to electrospray ionisation (ESI), where the voltage is applied to the spray capillary. USI can increase signal intensity in targeted LC-MS, though its performance for SNTS of complex water matrices remains unclear. USI and ESI were compared using standard mixtures, evaluating signal intensity, reproducibility, linearity, limits of detection, adduct formation, and in-source fragmentation, and further assessed in wastewater and stormwater extracts via feature detection, suspect hits, matrix effects, and correlations with physicochemical parameters such as water-solubility (LogS), retention time, and drift time. USI produced higher signal intensities than ESI for most compounds. In standards, 48/58 (positive) and 47/60 (negative) showed higher peak areas with USI (2.6-3.0-fold). In environmental samples, 80-85% of features had higher peak areas (2.3-2.4-fold). USI detected more features in five of six extracts, especially low-intensity signals, without proportional increases in suspect annotations. In total, 94 (positive) and 54 (negative) features were annotated (levels 1-3), mainly shared between sources. The clearest advantage of USI was observed for low-abundance suspects in diluted stormwater, where several compounds were uniquely detected at lower enrichment factors and matrix effects were shown to be compound-specific. Correlations indicated that USI preferentially enhances compounds that are more water-soluble, earlier-eluting and have shorter drift times. Overall, USI and ESI could provide complementary analytical response, as certain compounds are preferentially detected by one ion source over the other. Combined use could improve chemical-space coverage and enhance the detection of environmentally relevant compounds in wastewater and stormwater screening.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Water Pollutants, ChemicalSpectrometry, Mass, Electrospray IonizationLiquid Chromatography-Mass SpectrometryOrganic ChemicalsReproducibility of ResultsWastewaterLimit of DetectionLinear Models

Stichwörter der Autor:innen

Environmental monitoringIonisation efficiencyOrganic micropollutantsStormwaterWastewater

Substanzen

Water Pollutants, ChemicalWastewaterOrganic Chemicals

Autor:innen und Institutionen

  • Department of Physics, Chemistry, and Biology (IFM), Linköping University, SE-581 83, Linköping, Sweden. Electronic address: victoria.eriksson@liu.se.
  • Department of Thematic Studies (TEMA), Linköping University, SE-581 83, Linköping, Sweden. Electronic address: kalpana.munnuru.singamshetty@liu.se.
  • Department of Plant and Environmental Sciences, University of Copenhagen, DK-1871, Frederiksberg, Denmark. Electronic address: tmmk@plen.ku.dk.
  • Department of Plant and Environmental Sciences, University of Copenhagen, DK-1871, Frederiksberg, Denmark. Electronic address: seti@plen.ku.dk.
  • Department of Plant and Environmental Sciences, University of Copenhagen, DK-1871, Frederiksberg, Denmark. Electronic address: jch@plen.ku.dk.
  • Department of Physics, Chemistry, and Biology (IFM), Linköping University, SE-581 83, Linköping, Sweden. Electronic address: johan.dahlen@liu.se.

Förderung

  • Lars Hierta Memorial Foundation
  • Horizon Europe
  • NoVo Foundation
  • Linköping University
  • Längmanska Cultural Foundation
  • Magnus Bergvall Foundation
  • Swedish Research Council

Zitieren

APA

Eriksson, V., Singamshetty, K. M., Karlsson, T. M. M., Tisler, S., Christensen, J. H., Dahlén, J., Lai, F. Y., & Wang, T. (2026). Performance comparison of UniSprayTM and electrospray ionisation coupled to ion-mobility high-resolution mass spectrometer for screening of organic pollutants in water matrices. Journal of Chromatography A, 1790, 467569. https://doi.org/10.1016/j.chroma.2026.467569

Vancouver

Eriksson V, Singamshetty KM, Karlsson TMM, Tisler S, Christensen JH, Dahlén J, et al. Performance comparison of UniSprayTM and electrospray ionisation coupled to ion-mobility high-resolution mass spectrometer for screening of organic pollutants in water matrices. Journal of Chromatography A. 2026;1790:467569. doi:10.1016/j.chroma.2026.467569

BibTeX
@article{Eriksson2026,
  author = {Eriksson, Victoria and Singamshetty, Kalpana Munnuru and Karlsson, Thomas Malte Molnár and Tisler, Selina and Christensen, Jan H and Dahlén, Johan and Lai, Foon Yin and Wang, Thanh},
  title = {Performance comparison of UniSprayTM and electrospray ionisation coupled to ion-mobility high-resolution mass spectrometer for screening of organic pollutants in water matrices},
  journal = {Journal of Chromatography A},
  year = {2026},
  volume = {1790},
  pages = {467569},
  doi = {10.1016/j.chroma.2026.467569}
}
Kostenlos starten →