EUVIMEDEUVIMED entdecken
Crossref

Increasing detection sensitivity in gas chromatography by cooling a nano-gravimetric detector

Ambroisine Michel, Guy Raffin, David Sanz, Régis Barattin, Éric Colinet, Jérôme Randon

Journal of Chromatography A · 2026-06 · 1777 · 466939

DOI: 10.1016/j.chroma.2026.466939 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Chromatography A, 2026-06. Verfasst von Ambroisine Michel und 5 weiteren Autor:innen. Beteiligte Einrichtung: Université Claude Bernard Lyon 1, CNRS, Institut des Sciences Analytiques, UMR 5280, 69100, Villeurbanne, France. Hauptthemen laut MEDLINE: Nanotechnology. Die Arbeit nennt 15 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Response of nanogravimetric detector is based on adsorption of molecules on the surface of a vibrating beam. For any compound, NGD temperature (originally from 40 to 200°C) is the main operating parameter to control its sensitivity. Introduction of a cooling unit to enlarge the operating temperature range (down to -100°C) led to a substantial improvement in detection sensitivity and a significant reduction in quantification limits. The response of tetrahydrothiophene (THT) over a wide temperature range was deeply investigated: peak height increased exponentially as temperature decreased, while noise was slightly lowered. Consequently, the limit of quantification (LOQ) of THT can be reduced from 22 ppm at 50°C to 0.16 ppm at -65°C. However, significant peak tailing was observed at very low temperatures, which may compromise the separation of complex mixtures. THT chromatographic peaks were successfully modeled using an Exponentially Modified Gaussian (EMG) function, and the temperature dependence of the EMG adjusted parameters has been discussed.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

NanotechnologyChromatography, GasLimit of DetectionThiophenesTemperature

Stichwörter der Autor:innen

Gas chromatographyLimit of quantificationNanogravimetric detectorTetrahydrothiophene

Substanzen

Thiophenes

Autor:innen und Institutionen

  • Université Claude Bernard Lyon 1, CNRS, Institut des Sciences Analytiques, UMR 5280, 69100, Villeurbanne, France.
  • APIX Analytics, 38000, Grenoble, France.
  • Université Claude Bernard Lyon 1, CNRS, Institut des Sciences Analytiques, UMR 5280, 69100, Villeurbanne, France. Electronic address: randon@univ-lyon1.fr.

Zitieren

APA

Michel, A., Raffin, G., Sanz, D., Barattin, R., Colinet, É., & Randon, J. (2026). Increasing detection sensitivity in gas chromatography by cooling a nano-gravimetric detector. Journal of Chromatography A, 1777, 466939. https://doi.org/10.1016/j.chroma.2026.466939

Vancouver

Michel A, Raffin G, Sanz D, Barattin R, Colinet É, Randon J. Increasing detection sensitivity in gas chromatography by cooling a nano-gravimetric detector. Journal of Chromatography A. 2026;1777:466939. doi:10.1016/j.chroma.2026.466939

BibTeX
@article{Michel2026,
  author = {Michel, Ambroisine and Raffin, Guy and Sanz, David and Barattin, Régis and Colinet, Éric and Randon, Jérôme},
  title = {Increasing detection sensitivity in gas chromatography by cooling a nano-gravimetric detector},
  journal = {Journal of Chromatography A},
  year = {2026},
  volume = {1777},
  pages = {466939},
  doi = {10.1016/j.chroma.2026.466939}
}
Kostenlos starten →