EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

On-line aptamer affinity solid-phase extraction capillary electrophoresis-mass spectrometry for procalcitonin analysis as a sepsis protein biomarker

Laura Pont, Belén Cebrián, Estela Giménez, Fernando Benavente

Journal of Chromatography A · 2025

DOI: 10.1016/j.chroma.2025.466183 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Chromatography A, 2025. Verfasst von Laura Pont und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Chemical Engineering and Analytical Chemistry, Institute for Research on Nutrition and Food Safety (INSA·UB), University of Barcelona, …. Hauptthemen laut MEDLINE: Procalcitonin, Sepsis, Solid Phase Extraction, Mass Spectrometry. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 36 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Agencia Estatal de Investigación, European Regional Development Fund.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Sepsis, a life-threatening condition characterized by organ and tissue damage, primarily caused by bacterial infection, remains a leading cause of global mortality. This study introduces an integrated analytical method for the purification, enrichment, separation, identification, and quantification of procalcitonin (PCT), a key biomarker for sepsis, using on-line aptamer affinity solid-phase extraction capillary electrophoresis-mass spectrometry (AA-SPE-CE-MS). The method employed SPE microcartridges packed with magnetic bead particles functionalized with a PCT-selective aptamer. Optimizing the analytical parameters was challenging due to the specific properties of PCT, which adopts a rigid structure, partly attributed to the presence of a disulfide bond. The final optimized conditions utilized 1 M acetic acid (pH 2.3) as the background electrolyte (BGE) for the separation, and an aqueous-organic solvent mixture of methanol:water (60:40, v/v) with 0.5 M ammonia (apparent pH 11.3) for the elution. The developed method demonstrated excellent repeatability and reproducibility, with linearity within the 2.5 and 10 µg·mL⁻¹ range and a limit of detection (LOD) of 1 µg·mL⁻¹, yielding a 100-fold enhancement in sensitivity over CE-MS. Additionally, it was successfully applied to analyze PCT in serum samples after a precipitation-based clean-up pretreatment, showcasing its potential as a reliable MS-based analytical methodology for the analysis of intact PCT as a sepsis biomarker.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ProcalcitoninSepsisSolid Phase ExtractionMass SpectrometryAptamers, NucleotideElectrophoresis, CapillaryBiomarkersHumansReproducibility of ResultsLimit of DetectionLinear Models

Stichwörter der Autor:innen

AptamerCapillary electrophoresisIn-line solid-phase extractionMass spectrometryProtein

Substanzen

ProcalcitoninBiomarkersAptamers, Nucleotide

Autor:innen und Institutionen

  • Department of Chemical Engineering and Analytical Chemistry, Institute for Research on Nutrition and Food Safety (INSA·UB), University of Barcelona, Spain; Serra Húnter Program, Generalitat de Catalunya, Spain.
  • Department of Chemical Engineering and Analytical Chemistry, Institute for Research on Nutrition and Food Safety (INSA·UB), University of Barcelona, Spain.
  • Department of Chemical Engineering and Analytical Chemistry, Institute for Research on Nutrition and Food Safety (INSA·UB), University of Barcelona, Spain. Electronic address: fbenavente@ub.edu.

Förderung

  • Agencia Estatal de Investigación
  • European Regional Development Fund

Zitieren

APA

Pont, L., Cebrián, B., Giménez, E., & Benavente, F. (2025). On-line aptamer affinity solid-phase extraction capillary electrophoresis-mass spectrometry for procalcitonin analysis as a sepsis protein biomarker. Journal of Chromatography A. https://doi.org/10.1016/j.chroma.2025.466183

Vancouver

Pont L, Cebrián B, Giménez E, Benavente F. On-line aptamer affinity solid-phase extraction capillary electrophoresis-mass spectrometry for procalcitonin analysis as a sepsis protein biomarker. Journal of Chromatography A. 2025. doi:10.1016/j.chroma.2025.466183

BibTeX
@article{Pont2025,
  author = {Pont, Laura and Cebrián, Belén and Giménez, Estela and Benavente, Fernando},
  title = {On-line aptamer affinity solid-phase extraction capillary electrophoresis-mass spectrometry for procalcitonin analysis as a sepsis protein biomarker},
  journal = {Journal of Chromatography A},
  year = {2025},
  doi = {10.1016/j.chroma.2025.466183}
}
Kostenlos starten →