Analysis of semivolatile organics in liquid radioactive residue using sorptive stir bar and solvent back-extraction

Andrew J Boggess, Thomas L White

Journal of Chromatography A · 2025

DOI: 10.1016/j.chroma.2025.466167 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Chromatography A, 2025. Verfasst von Andrew J Boggess und 1 weiteren Autor:innen. Beteiligte Einrichtung: Analytical Characterization and Sample Management, Savannah River National Laboratory, SRNL, Aiken, SC 29808, USA. Electronic address: andrew.boggess…. Hauptthemen laut MEDLINE: Organic Chemicals, Radioactive Waste, Water Pollutants, Radioactive, Chemical Fractionation. Die Arbeit nennt 33 Literaturangaben. Gefördert von: Savannah River National Laboratory, Laboratory Directed Research and Development, Office of Environmental Management.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Current methods for semivolatiles analysis in radioactive samples can produce large volumes of radioactive solvent residue. A method utilizing stir-bar sorptive extraction has been explored in this work for its applicability to radioactive waste samples. This low solvent analytical method may accelerate remediation, minimize hazardous solvent waste, and reduce exposure risk to workers. Organic compounds (polyaromatic hydrocarbons, chlorinated aromatics, and phenolic compounds) were chosen as surrogates for common Liquid Waste System (LWS) contaminants at Savannah River Site (Aiken, SC). Stir-bar extraction parameters (extraction time, matrix modification, and effective pH range) and solvent back extraction parameters (solvent type, volume, and extraction time) were optimized experimentally for the chosen compounds. Affinity of the stir-bar extraction polymer to radionuclides Cs-137 and Am-241 was observed to determine radionuclide concentration effects. The stir-bar method achieved mean recovery of 100 ± 0.7 % (1σ), relative to 114 ± 7 % using solvent extraction, while reducing weekly method hands-on time by 93.4 % and solvent volume consumption by 99.3 %. Sensitivity was improved by 378 % in simulated tank waste and 278 % in real-world LWS matrix, relative to solvent extraction. This work has produced a safe and optimized method for the low solvent analysis of organics in legacy radioactive tank waste by stir-bar sorptive extraction.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Organic ChemicalsRadioactive WasteWater Pollutants, RadioactiveChemical FractionationSolventsReproducibility of ResultsLimit of DetectionLinear Models

Stichwörter der Autor:innen

GC-MSNuclear wastePolyaromatic hydrocarbonsStir-bar extraction

Substanzen

Organic ChemicalsSolventsRadioactive WasteWater Pollutants, Radioactive

Autor:innen und Institutionen

  • Analytical Characterization and Sample Management, Savannah River National Laboratory, SRNL, Aiken, SC 29808, USA. Electronic address: andrew.boggess@srnl.doe.gov.
  • Analytical Characterization and Sample Management, Savannah River National Laboratory, SRNL, Aiken, SC 29808, USA.

Förderung

  • Savannah River National Laboratory
  • Laboratory Directed Research and Development
  • Office of Environmental Management
  • U.S. Department of Energy

Zitieren

APA

Boggess, A. J., & White, T. L. (2025). Analysis of semivolatile organics in liquid radioactive residue using sorptive stir bar and solvent back-extraction. Journal of Chromatography A. https://doi.org/10.1016/j.chroma.2025.466167

Vancouver

Boggess AJ, White TL. Analysis of semivolatile organics in liquid radioactive residue using sorptive stir bar and solvent back-extraction. Journal of Chromatography A. 2025. doi:10.1016/j.chroma.2025.466167

BibTeX
@article{Boggess2025,
  author = {Boggess, Andrew J and White, Thomas L},
  title = {Analysis of semivolatile organics in liquid radioactive residue using sorptive stir bar and solvent back-extraction},
  journal = {Journal of Chromatography A},
  year = {2025},
  doi = {10.1016/j.chroma.2025.466167}
}
Kostenlos starten →