Advancing HIC method development: Retention-time modeling and tuning selectivity with ternary mobile-phase systems
Journal of Chromatography A · 2024
DOI: 10.1016/j.chroma.2024.465133 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Chromatography A, 2024. Verfasst von Raphael Ewonde Ewonde und 3 weiteren Autor:innen. Beteiligte Einrichtung: Vrije Universiteit Brussel (VUB), Department of Chemical Engineering, Brussels, Belgium. Hauptthemen laut MEDLINE: Muramidase, Hydrophobic and Hydrophilic Interactions, Ammonium Sulfate. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Gefördert von: Nederlandse Organisatie voor Wetenschappelijk Onderzoek, BASF, Ministry of Economic Affairs.
Abstract (Originalsprache)
The use of a ternary mobile-phase system comprising ammonium sulphate, sodium chloride, and phosphate buffer was explored to tune retention and enhance selectivity in hydrophobic interaction chromatography. The accuracy of the linear solvent-strength model to predict protein retention with the ternary mobile-phase system based on isocratic scouting runs is limited, as the extrapolated retention factor at aqueous buffer conditions (k0) cannot be reliably established. The Jandera retention model utilizing a salt concentration averaged retention factor (k¯0) in aqueous buffer for ternary systems overcomes this bottleneck. Gradient retention factors were derived based on isocratic scouting runs after numerical integration of the isocratic Jandera model, leading to retention-time prediction errors below 11 % for linear gradients. Furthermore, an analytical expression was formulated to predict HIC retention for both linear and segmented linear gradients, considering the linear solvent-strength (LSS) model within ternary salt systems, relying on a fixed k0. The approach involved conducting two gradient scouting runs for each of the two binary salt systems to determine model parameters. Retention-time prediction errors for linear gradients were below 12 % for lysozyme and 3 % for trypsinogen and α-chymotrypsinogen A. Finally, the analytical expression for a ternary mobile-phase system was used in combination with a genetic algorithm to tune the HIC selectivity. With an optimized segmented ternary gradient, a critical-pair separation for a mixture of 7 proteins was achieved within 15 min with retention-time prediction errors ranging between 0.7 and 15.7 %.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Vrije Universiteit Brussel (VUB), Department of Chemical Engineering, Brussels, Belgium.
- Vrije Universiteit Amsterdam, Amsterdam Institute of Molecular and Life Sciences, Division of Bioanalytical Chemistry, Amsterdam, the Netherlands; Centre for Analytical Sciences Amsterdam (CASA), Amsterdam, the Netherla…
- Vrije Universiteit Brussel (VUB), Department of Chemical Engineering, Brussels, Belgium. Electronic address: seeltink@vub.be.
Förderung
- Nederlandse Organisatie voor Wetenschappelijk Onderzoek
- BASF
- Ministry of Economic Affairs
- FWO
- NWO-I
Zitieren
APA
Ewonde Ewonde, R., Molenaar, S. R. A., Broeckhoven, K., & Eeltink, S. (2024). Advancing HIC method development: Retention-time modeling and tuning selectivity with ternary mobile-phase systems. Journal of Chromatography A. https://doi.org/10.1016/j.chroma.2024.465133
Vancouver
Ewonde Ewonde R, Molenaar SRA, Broeckhoven K, Eeltink S. Advancing HIC method development: Retention-time modeling and tuning selectivity with ternary mobile-phase systems. Journal of Chromatography A. 2024. doi:10.1016/j.chroma.2024.465133
BibTeX
@article{EwondeEwonde2024,
author = {Ewonde Ewonde, Raphael and Molenaar, Stef R A and Broeckhoven, Ken and Eeltink, Sebastiaan},
title = {Advancing HIC method development: Retention-time modeling and tuning selectivity with ternary mobile-phase systems},
journal = {Journal of Chromatography A},
year = {2024},
doi = {10.1016/j.chroma.2024.465133}
}
EUVIMED