Fachartikel in der Zeitschrift Journal of Chromatography A, 2023-07. Verfasst von Mark R Taylor und 2 weiteren Autor:innen. Beteiligte Einrichtung: Pfizer Global Research and Development, Discovery Park, Ramsgate Road, Sandwich, CT13 9NJ, UK. Hauptthemen laut MEDLINE: Water. Die Arbeit wurde bisher 25-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Pfizer, Agilent Technologies.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Sample injection can cause serious problems in hydrophilic interaction liquid chromatography (HILIC) when the injection solvent has higher elution strength than the mobile phase (mp). It can lead to asymmetric peak shapes and poor efficiency. The problem can occur when the mp contains a high proportion of organic e.g. 95% acetonitrile (a weak solvent) whereas the injection solvent contains a higher proportion of water (a strong solvent) that is necessary to dissolve polar samples. We investigated different strategies to overcome this problem. A simple method is pre-column dilution where the injector is programmed to deliver a plug of weak solvent (e.g. pure acetonitrile) along with the sample dissolved in a solvent with higher water content than the mp. Another option is to use alternative organic solvents to acetonitrile in the injection solvent, e.g. isopropanol, acetone or tetrahydrofuran, that may give enhanced sample solubility. The role of the volume of injection solvents was investigated as well as the possible effects of mass overload on the results. The use of small sample volumes is always recommended to reduce mismatch effects.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

WaterChromatography, LiquidSolventsHydrophobic and Hydrophilic InteractionsAcetonitrilesIndicators and Reagents

Stichwörter der Autor:innen

HILICHPLCSample injection

Substanzen

SolventsWateracetonitrileAcetonitrilesIndicators and Reagents

Autor:innen und Institutionen

  • Pfizer Global Research and Development, Discovery Park, Ramsgate Road, Sandwich, CT13 9NJ, UK.
  • Pfizer Global Research and Development, 280 Shennecossett Rd, Groton, CT 06340, USA.
  • Centre for Research in Biosciences, University of the West of England, Frenchay, Bristol, BS16 1QY, UK. Electronic address: David.Mccalley@uwe.ac.uk.

Förderung

  • Pfizer
  • Agilent Technologies

Zitieren

APA

Taylor, M. R., Kawakami, J., & McCalley, D. V. (2023). Managing sample introduction problems in hydrophilic interaction liquid chromatography. Journal of Chromatography A, 1700, 464006. https://doi.org/10.1016/j.chroma.2023.464006

Vancouver

Taylor MR, Kawakami J, McCalley DV. Managing sample introduction problems in hydrophilic interaction liquid chromatography. Journal of Chromatography A. 2023;1700:464006. doi:10.1016/j.chroma.2023.464006

BibTeX
@article{Taylor2023,
  author = {Taylor, Mark R and Kawakami, Jane and McCalley, David V},
  title = {Managing sample introduction problems in hydrophilic interaction liquid chromatography},
  journal = {Journal of Chromatography A},
  year = {2023},
  volume = {1700},
  pages = {464006},
  doi = {10.1016/j.chroma.2023.464006}
}
Kostenlos starten →