Shape-Sensing Robotic-Assisted Bronchoscopy vs Digital Tomosynthesis-Corrected Electromagnetic Navigation Bronchoscopy
Chest · 2023-04 · 163 (4) · 977-984
DOI: 10.1016/j.chest.2022.10.019 ↗Kurz gefasst
Fachartikel in der Zeitschrift Chest, 2023-04. Verfasst von See-Wei Low und 8 weiteren Autor:innen. Beteiligte Einrichtung: Division of Allergy, Pulmonary and Critical Care Medicine, Vanderbilt University Medical Center, Nashville, TN. Hauptthemen laut MEDLINE: Robotic Surgical Procedures, Lung Neoplasms. Die Arbeit wurde bisher 83-mal zitiert (Crossref) und nennt 23 Literaturangaben. Gefördert von: Medtronic.
Abstract (Originalsprache)
BACKGROUND: Electromagnetic navigational bronchoscopy has been the dominant bronchoscopic technology for targeting small peripheral lesions and now includes digital tomosynthesis-electromagnetic navigational bronchoscopy (DT-ENB), allowing near-real-time intraprocedural nodule visualization. Shape-sensing robotic-assisted bronchoscopy (ssRAB), with improved catheter stability and articulation recently became available. Although the diagnostic performance of these two methods seems higher than that of legacy systems, data remain limited. We sought to compare the diagnostic yield of these two novel platforms after their introduction at our institution. RESEARCH QUESTION: Does the diagnostic yield of ssRAB differ significantly from that of DT-ENB in patients undergoing biopsy of peripheral pulmonary lesions (PPLs)? STUDY DESIGN AND METHODS: This retrospective comparative cohort study analyzed prospectively collected data on consecutive procedures performed with DT-ENB and ssRAB in their first 6 months of use at our institution. Biopsies were considered diagnostic if histopathologic analysis revealed malignancy or specific benign features that readily explained the presence of a PPL. Nonspecific inflammation, normal lung or airway, and atypia not diagnostic of malignancy were considered nondiagnostic. RESULTS: SSRAB was used to biopsy 143 PPLs in 133 patients and DT-ENB was used to biopsy 197 PPLs in 170 patients. Diagnostic yield was 77% for ssRAB (110 of 143 PPLs) and 80% (158 of 197 PPLs) for DT-ENB (OR, 0.8; 95% CI, 0.5-1.4; P = .4). Median lesion diameters were 17 and 19 mm, respectively. No difference in diagnostic yield was found after adjustment for lesion size, bronchus sign, peripheral vs middle third location, and sex. Pneumothorax complicated 1.5% of ssRAB and 1.8% of DT-ENB procedures (P = .86). INTERPRETATION: SSRAB and DT-ENB showed comparable diagnostic yields and safety profiles in this comparative cohort study.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Division of Allergy, Pulmonary and Critical Care Medicine, Vanderbilt University Medical Center, Nashville, TN.
- Department of Biostatistics, Vanderbilt University Medical Center, Nashville, TN.
- Division of Pulmonary and Critical Care, Department of Internal Medicine, Rush University, Chicago, IL.
- Division of Pulmonary and Critical Care, St. Luke's Health System, University of Missouri-Kansas City, Kansas City, MO.
- Division of Allergy, Pulmonary and Critical Care Medicine, Vanderbilt University Medical Center, Nashville, TN. Electronic address: fabien.maldonado@vumc.org.
Förderung
- Medtronic
Zitieren
APA
Low, S. W., Lentz, R. J., Chen, H., Katsis, J., Aboudara, M. C., Whatley, S., Paez, R., Rickman, O. B., & Maldonado, F. (2023). Shape-Sensing Robotic-Assisted Bronchoscopy vs Digital Tomosynthesis-Corrected Electromagnetic Navigation Bronchoscopy. Chest, 163(4), 977-984. https://doi.org/10.1016/j.chest.2022.10.019
Vancouver
Low SW, Lentz RJ, Chen H, Katsis J, Aboudara MC, Whatley S, et al. Shape-Sensing Robotic-Assisted Bronchoscopy vs Digital Tomosynthesis-Corrected Electromagnetic Navigation Bronchoscopy. Chest. 2023;163(4):977-984. doi:10.1016/j.chest.2022.10.019
BibTeX
@article{Low2023,
author = {Low, See-Wei and Lentz, Robert J and Chen, Heidi and Katsis, James and Aboudara, Matthew C and Whatley, Samuel and Paez, Rafael and Rickman, Otis B and Maldonado, Fabien},
title = {Shape-Sensing Robotic-Assisted Bronchoscopy vs Digital Tomosynthesis-Corrected Electromagnetic Navigation Bronchoscopy},
journal = {Chest},
year = {2023},
volume = {163},
number = {4},
pages = {977--984},
doi = {10.1016/j.chest.2022.10.019}
}
EUVIMED