Statistical comparison for assessing agreement between two mass spectrometric methods for the analysis of polychlorinated dibenzo-p-dioxins and furans (PCDDs/Fs) in contaminated soils
Chemosphere · 2024-09 · 363 · 142806
DOI: 10.1016/j.chemosphere.2024.142806 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Chemosphere, 2024-09. Verfasst von Jungmin Jo und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Environmental Science & Engineering, Ewha Womans University, Seoul, 03760, South Korea. Hauptthemen laut MEDLINE: Polychlorinated Dibenzodioxins, Soil Pollutants, Gas Chromatography-Mass Spectrometry, Environmental Monitoring. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Korea Basic Science Institute, Korea Ministry of Environment.
Abstract (Originalsprache)
A gas chromatography coupled to high-resolution mass spectrometry (GC-HR/MS) has been used as the standard method for the quantification of polychlorinated dibenzo-p-dioxins and furans (PCDDs/Fs), which are regulated at screening and action levels in the environment. However, several alternative methods have been attempted due to the disadvantage of its high cost. Although a gas chromatography with triple quadrupole mass spectrometry (GC-QqQ-MS/MS) has been used in a wide variety of sample matrices, showing that they are interchangeable, there has been a lack of comprehensive studies on statistical agreement with GC-HR/MS. In this study, a pairwise comparison of the total concentrations of PCDDs/Fs in 90 soil field samples obtained by two mass spectrometric methods was performed using the Passing-Bablok (P&B) regression and Bland-Altman (B&A) analysis for the method comparison. According to the result of the B&A analysis, the concentration range of PCDDs/Fs was between 98.2 and 1760 pg/g showed good agreement between two methods at the 95 % confidence level (CL). Although there was a large discrepancy between the two methods in the low concentrations (
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Environmental Science & Engineering, Ewha Womans University, Seoul, 03760, South Korea.
- Department of Environmental Engineering, Kumoh National Institute of Technology, Gumi, 39177, South Korea; Department of Energy Engineering Convergence, Kumoh National Institute of Technology, 39177, South Korea.
- Department of Environmental Engineering, Jeju National University, 63243, South Korea.
- Metropolitan Seoul Center, Korea Basic Science Institute, University-Industry Cooperation Building, 150, Bugahyeon-ro, Seodaemun-gu, Seoul, 03759, South Korea.
- Metropolitan Seoul Center, Korea Basic Science Institute, University-Industry Cooperation Building, 150, Bugahyeon-ro, Seodaemun-gu, Seoul, 03759, South Korea. Electronic address: ygahn@kbsi.re.kr.
Förderung
- Korea Basic Science Institute
- Korea Ministry of Environment
Zitieren
APA
Jo, J., Son, Y., Park, M. K., Lee, J. Y., Chu, H., & Ahn, Y. G. (2024). Statistical comparison for assessing agreement between two mass spectrometric methods for the analysis of polychlorinated dibenzo-p-dioxins and furans (PCDDs/Fs) in contaminated soils. Chemosphere, 363, 142806. https://doi.org/10.1016/j.chemosphere.2024.142806
Vancouver
Jo J, Son Y, Park MK, Lee JY, Chu H, Ahn YG. Statistical comparison for assessing agreement between two mass spectrometric methods for the analysis of polychlorinated dibenzo-p-dioxins and furans (PCDDs/Fs) in contaminated soils. Chemosphere. 2024;363:142806. doi:10.1016/j.chemosphere.2024.142806
BibTeX
@article{Jo2024,
author = {Jo, Jungmin and Son, Younggyu and Park, Min-Kyu and Lee, Ji Yi and Chu, Haena and Ahn, Yun Gyong},
title = {Statistical comparison for assessing agreement between two mass spectrometric methods for the analysis of polychlorinated dibenzo-p-dioxins and furans (PCDDs/Fs) in contaminated soils},
journal = {Chemosphere},
year = {2024},
volume = {363},
pages = {142806},
doi = {10.1016/j.chemosphere.2024.142806}
}
EUVIMED