Meropenem interference on phosphate measurement: utility of reaction curve analysis in automated biochemistry analyzers

Chun-Fung So, Rachel Sze-Wan Yiu, Tsz-Ki Ling, Man-Ting Wong, Melody Yee-Man Wong, Felix Chi-Kin Wong, Timothy Ming-Him Yeung, Ching-Wan Lam

Clinica Chimica Acta · 2026-06 · 588 · 120976

DOI: 10.1016/j.cca.2026.120976 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Clinica Chimica Acta, 2026-06. Verfasst von Chun-Fung So und 7 weiteren Autor:innen. Beteiligte Einrichtung: Division of Chemical Pathology, Department of Pathology, Queen Mary Hospital, Hong Kong, China. Hauptthemen laut MEDLINE: Phosphates, Thienamycins, Blood Chemical Analysis. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Analytical errors in plasma phosphate measurement can result in inappropriate treatment. Meropenem, not previously known to affect plasma phosphate assays, was suspected to be the interferent after a case of negative phosphate measurement with abnormal reaction curve for a meropenem-containing specimen on the Roche Cobas 8000 Analytic System. We studied the effect of meropenem on Roche's phosphate assay and the utility of reaction curve analysis in interference detection. METHODS: Meropenem-spiked plasma specimens with a range of estimated meropenem concentrations were prepared by mixing pooled human plasma from 46 patients and 20 mg/mL meropenem in normal saline in predetermined volume ratios. The relationship between meropenem concentration and percentage difference on apparent phosphate concentration on Roche was recorded as an interferogram. Reaction curves generated by the analyzer were also examined. RESULTS: Percentage difference on apparent phosphate concentration on Roche increases with meropenem concentration, with 50× therapeutic levels). Reaction curves of meropenem-containing samples were markedly different from the control. CONCLUSION: Meropenem leads to negative, spectral, concentration-dependent, platform-specific interference on Roche phosphate assay. Reaction curve analysis improves interference detection in automated biochemistry analyzers and should be implemented in the form of laboratory protocols to prevent analytical errors and patient harm.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PhosphatesThienamycinsBlood Chemical AnalysisMeropenemHumansAutomation

Stichwörter der Autor:innen

Automated biochemistry analyzerInterferenceMeropenemPhosphateReaction curve

Substanzen

MeropenemPhosphatesThienamycins

Autor:innen und Institutionen

  • Division of Chemical Pathology, Department of Pathology, Queen Mary Hospital, Hong Kong, China.
  • Division of Chemical Pathology, Department of Pathology, Queen Mary Hospital, Hong Kong, China; Department of Pathology, The University of Hong Kong, Hong Kong, China. Electronic address: ching-wanlam@pathology.hku.hk.

Zitieren

APA

So, C. F., Yiu, R. S., Ling, T. K., Wong, M. T., Wong, M. Y., Wong, F. C., Yeung, T. M., & Lam, C. W. (2026). Meropenem interference on phosphate measurement: utility of reaction curve analysis in automated biochemistry analyzers. Clinica Chimica Acta, 588, 120976. https://doi.org/10.1016/j.cca.2026.120976

Vancouver

So CF, Yiu RS, Ling TK, Wong MT, Wong MY, Wong FC, et al. Meropenem interference on phosphate measurement: utility of reaction curve analysis in automated biochemistry analyzers. Clinica Chimica Acta. 2026;588:120976. doi:10.1016/j.cca.2026.120976

BibTeX
@article{So2026,
  author = {So, Chun-Fung and Yiu, Rachel Sze-Wan and Ling, Tsz-Ki and Wong, Man-Ting and Wong, Melody Yee-Man and Wong, Felix Chi-Kin and Yeung, Timothy Ming-Him and Lam, Ching-Wan},
  title = {Meropenem interference on phosphate measurement: utility of reaction curve analysis in automated biochemistry analyzers},
  journal = {Clinica Chimica Acta},
  year = {2026},
  volume = {588},
  pages = {120976},
  doi = {10.1016/j.cca.2026.120976}
}
Kostenlos starten →