EUVIMEDEUVIMED entdecken
Validierungsstudie

Rapid, robust LC–MS/MS quantification of cefazolin in whole blood microsamples, plasma, and plasma ultrafiltrate: Analytical method validation and preliminary clinical application in pediatric population

Arkadiusz Kocur, Agnieszka Czajkowska, Mateusz Moczulski, Anna Częczek, Małgorzata Mikaszewska-Sokolewicz

Clinica Chimica Acta · 2026

DOI: 10.1016/j.cca.2025.120808 ↗

Kurz gefasst

Validierungsstudie in der Zeitschrift Clinica Chimica Acta, 2026. Verfasst von Arkadiusz Kocur und 4 weiteren Autor:innen. Beteiligte Einrichtung: Therapeutic Drug Monitoring, Clinical Pharmacokinetics and Toxicology Laboratory Unit, Department of Clinical Biochemistry, The Children's Memorial H…. Hauptthemen laut MEDLINE: Tandem Mass Spectrometry, Cefazolin, Anti-Bacterial Agents, Blood Chemical Analysis. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 28 Literaturangaben. Gefördert von: Medical Research Agency.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Optimal dosing of cephalosporins in pediatric ICU patients is difficult due to high inter-individual variability in β-lactam pharmacokinetics, limited blood volume, and the lack of validated microsampling-compatible clinical assays. We developed and validated a rapid and reliable liquid chromatography-tandem mass spectrometry method to measure cefazolin concentrations in lithium-heparin plasma, plasma ultrafiltrate (unbound fraction), and capillary whole blood collected with volumetric absorptive microsampling (VAMS). This method was applied in a pilot cohort of pediatric patients. Samples were protein precipitated with cryo-cooled organic solvents, separated by reversed-phase liquid chromatography (reversed-phase HPLC) using a polar C18 column, and detected by multiple reaction monitoring (MRM) on a hybrid triple quadrupole-linear ion trap (QTRAP) mass spectrometer. Cefotaxime-D3 was used as the internal standard, and calibration curves (0.5-200 mg/L) were established for each matrix. The method was validated in accordance with ICH M10 bioanalytical guidelines for selectivity, linearity, accuracy, precision, carry-over, matrix effect, recovery, process efficiency, stability, and VAMS-specific factors (hematocrit and drying time). The total run time was 3 min, including the internal standard, with an LLOQ of 0.5 mg/L across all matrices. The coefficients of determination (R2) were > 0.995, IS-normalized matrix factors were close to 1.0, and accuracy and precision were within ±15 % at all quality control levels. Cefazolin remained stable under all preanalytical conditions studied. VAMS values were unaffected by changes in hematocrit within clinically relevant ranges. The feasibility of the multi-matrix analytical method in routine practice has been demonstrated. The multi-matrix, microsampling-compatible LC-MS/MS assay described is sensitive and specific for TDM and PK/PD studies, aimed at defining precise dosing regimens for cefazolin in neonatal and pediatric critical care settings.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tandem Mass SpectrometryCefazolinAnti-Bacterial AgentsBlood Chemical AnalysisHumansChildChild, PreschoolInfantChromatography, LiquidChromatography, High Pressure LiquidTime FactorsMaleLiquid Chromatography-Mass Spectrometry

Stichwörter der Autor:innen

CefazolinLC-MS/MSPerioperative antimicrobial prophylaxisPlasma ultrafiltrateTherapeutic drug monitoringUnbound drugsVolumetric absorptive microsampling

Substanzen

CefazolinAnti-Bacterial Agents

Autor:innen und Institutionen

  • Therapeutic Drug Monitoring, Clinical Pharmacokinetics and Toxicology Laboratory Unit, Department of Clinical Biochemistry, The Children's Memorial Health Institute, 04-730 Warsaw, Poland. Electronic address: a.kocur@ip…
  • Therapeutic Drug Monitoring, Clinical Pharmacokinetics and Toxicology Laboratory Unit, Department of Clinical Biochemistry, The Children's Memorial Health Institute, 04-730 Warsaw, Poland.
  • Department of Anesthesiology and Intensive Care, The Children's Memorial Health Institute, 04-730 Warsaw, Poland.

Förderung

  • Medical Research Agency

Zitieren

APA

Kocur, A., Czajkowska, A., Moczulski, M., Częczek, A., & Mikaszewska-Sokolewicz, M. (2026). Rapid, robust LC–MS/MS quantification of cefazolin in whole blood microsamples, plasma, and plasma ultrafiltrate: Analytical method validation and preliminary clinical application in pediatric population. Clinica Chimica Acta. https://doi.org/10.1016/j.cca.2025.120808

Vancouver

Kocur A, Czajkowska A, Moczulski M, Częczek A, Mikaszewska-Sokolewicz M. Rapid, robust LC–MS/MS quantification of cefazolin in whole blood microsamples, plasma, and plasma ultrafiltrate: Analytical method validation and preliminary clinical application in pediatric population. Clinica Chimica Acta. 2026. doi:10.1016/j.cca.2025.120808

BibTeX
@article{Kocur2026,
  author = {Kocur, Arkadiusz and Czajkowska, Agnieszka and Moczulski, Mateusz and Częczek, Anna and Mikaszewska-Sokolewicz, Małgorzata},
  title = {Rapid, robust LC–MS/MS quantification of cefazolin in whole blood microsamples, plasma, and plasma ultrafiltrate: Analytical method validation and preliminary clinical application in pediatric population},
  journal = {Clinica Chimica Acta},
  year = {2026},
  doi = {10.1016/j.cca.2025.120808}
}