Prospective evaluation of hydrogel use in gynecologic brachytherapy

Darien Colson-Fearon, Rohini Bhatia, Michael Roumeliotis, Marc Morcos, Akila N Viswanathan

Brachytherapy · 2026-07 · 25 (4) · 578-583

DOI: 10.1016/j.brachy.2025.12.006 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Brachytherapy, 2026-07. Verfasst von Darien Colson-Fearon und 4 weiteren Autor:innen. Beteiligte Einrichtung: Johns Hopkins University School of Medicine, Baltimore, MD. Hauptthemen laut MEDLINE: Brachytherapy, Genital Neoplasms, Female, Hydrogel, Polyethylene Glycol Dimethacrylate, Hydrogels. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 21 Literaturangaben. Gefördert von: Augmenix Inc.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To evaluate hydrogel visualization as a marker on CT and MR imaging and consider any potential dosimetric effects incurred by placement of the marker on brachytherapy for the treatment of gynecologic malignancies. METHODS: This prospective cohort study enrolled 25 patients with gynecologic cancers who underwent brachytherapy at Johns Hopkins Hospital. MR and CT imaging were used to compare TraceITTM hydrogel visualization immediately after placement with the brachytherapy applicator in place. Rectal dosimetric data of a 2 cm3 volume (D2cc) and equi-effective doses of 2 Gy fractions (EQD2) were obtained. For patients with data collected before and after spacer placement, a comparative analysis was conducted to assess the impact of the marker on rectal dose. RESULTS: Hydrogel was safely injected into the space between the rectum and vagina in 25 patients, with no side effects related to gel placement. When comparing the agreement in gel visualization by 2 interpreters, inter-rater reliability (kappa statistic) was calculated and showed 58.3% (moderate) agreement (p = 0.0007). In patients who received a brachytherapy fraction both before and after TraceITTM placement, the average rectal displacement was 1.33 cm. The change in EQD2 D2cc rectum was significantly lower, with an average pre-TraceITTM dose of 5.02 Gy and a post-TraceITTM average of 4.60 Gy (p = 0.002). CONCLUSIONS: Hydrogel placement is feasible during brachytherapy procedures for the treatment of gynecologic malignancies. Using TraceITTM as a marker resulted in better visualization on MR than on CT. Displacement significantly reduced the D2cc to the rectum. Further research is needed to determine the optimal amount of hydrogel for spacing in gynecologic cancer patients.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrachytherapyGenital Neoplasms, FemaleHydrogel, Polyethylene Glycol DimethacrylateHydrogelsHumansFemaleProspective StudiesMiddle AgedAgedRadiotherapy DosageTomography, X-Ray ComputedMagnetic Resonance ImagingFiducial MarkersRectumAdult

Stichwörter der Autor:innen

BrachytherapyGynecologic malignancyHydrogel spacing

Substanzen

Hydrogel, Polyethylene Glycol DimethacrylateHydrogels

Autor:innen und Institutionen

  • Johns Hopkins University School of Medicine, Baltimore, MD.
  • Johns Hopkins University School of Medicine, Department of Radiation Oncology and Molecular Radiation Sciences, Baltimore, MD.
  • Johns Hopkins University School of Medicine, Department of Radiation Oncology and Molecular Radiation Sciences, Baltimore, MD. Electronic address: anv@jhu.edu.

Förderung

  • Augmenix Inc

Zitieren

APA

Colson-Fearon, D., Bhatia, R., Roumeliotis, M., Morcos, M., & Viswanathan, A. N. (2026). Prospective evaluation of hydrogel use in gynecologic brachytherapy. Brachytherapy, 25(4), 578-583. https://doi.org/10.1016/j.brachy.2025.12.006

Vancouver

Colson-Fearon D, Bhatia R, Roumeliotis M, Morcos M, Viswanathan AN. Prospective evaluation of hydrogel use in gynecologic brachytherapy. Brachytherapy. 2026;25(4):578-583. doi:10.1016/j.brachy.2025.12.006

BibTeX
@article{ColsonFearon2026,
  author = {Colson-Fearon, Darien and Bhatia, Rohini and Roumeliotis, Michael and Morcos, Marc and Viswanathan, Akila N},
  title = {Prospective evaluation of hydrogel use in gynecologic brachytherapy},
  journal = {Brachytherapy},
  year = {2026},
  volume = {25},
  number = {4},
  pages = {578--583},
  doi = {10.1016/j.brachy.2025.12.006}
}
Kostenlos starten →