EMVision: An electromagnetic tracking guidance system for accurate placement of interstitial brachytherapy applicators
Brachytherapy · 2024
DOI: 10.1016/j.brachy.2024.06.006 ↗Kurz gefasst
Fachartikel in der Zeitschrift Brachytherapy, 2024. Verfasst von Christopher L Deufel und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Mayo Clinic, Rochester, MN 55905. Electronic address: deufel.christopher@mayo.edu. Hauptthemen laut MEDLINE: Brachytherapy, Electromagnetic Phenomena. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 20 Literaturangaben.
Abstract (Originalsprache)
PURPOSE: To design, construct, and evaluate a system for image-guided placement of brachytherapy applicators using electromagnetic tracking (EMT) technology for improved procedure quality and efficiency. The system, named EMVision, provides a three-dimensional display of planned needle sites, visibility of the anatomy and needle position during placement, and reference tracking to account for generator or target anatomy shifts. METHODS: The EMVision EMT guidance system registers CT and EMT reference frames using an automated point-based algorithm according to tandem/ovoid locations. Subsequent needle placement and digitization is guided with an intraluminal EMT sensor and user interface that displays the needle's location in axial, coronal, sagittal, and 3D-volumetric CT views. The interface can overlay contours, preplanned needles, and provide 6DOF compensation for tandem movement during needle placement/manipulation. EMVision accuracy was evaluated in phantom and human cadaver by comparing EMVision's DICOM needle positions with a ground-truth, postimplant CT. RESULTS: Proof of concept was demonstrated for EMT-assisted placement of brachytherapy needles. EMVision accuracy in phantom (mean ± standard deviation) on a brachysuite CT table was 0.76 ± 0.13 mm for needle tips placed up to 75 mm from the tandem/ovoids and 0.52 ± 0.27 mm for needle shafts at distances up to 100 mm from the tandem/ovoids. Performance in human cadaver was similar, with tip and shaft accuracies of 0.77 ± 0.14 mm and 0.40 ± 0.21 mm, respectively. CONCLUSION: EMVision provides sub-millimeter accuracy for the placement of brachytherapy needles without repeated or continuous imaging. The technology can be used to reduce brachytherapy procedure times, improve the correspondence between intended and actual needle positions, or decrease the trainee learning curve.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiation Oncology, Mayo Clinic, Rochester, MN 55905. Electronic address: deufel.christopher@mayo.edu.
- Department of Radiation Oncology, Mayo Clinic, Rochester, MN 55905.
Zitieren
APA
Deufel, C. L., Brost, E. E., Dupere, J. M., Petersen, I. A., Haddock, M. G., & Garda, A. E. (2024). EMVision: An electromagnetic tracking guidance system for accurate placement of interstitial brachytherapy applicators. Brachytherapy. https://doi.org/10.1016/j.brachy.2024.06.006
Vancouver
Deufel CL, Brost EE, Dupere JM, Petersen IA, Haddock MG, Garda AE. EMVision: An electromagnetic tracking guidance system for accurate placement of interstitial brachytherapy applicators. Brachytherapy. 2024. doi:10.1016/j.brachy.2024.06.006
BibTeX
@article{Deufel2024,
author = {Deufel, Christopher L and Brost, Eric E and Dupere, Justine M and Petersen, Ivy A and Haddock, Michael G and Garda, Allison E},
title = {EMVision: An electromagnetic tracking guidance system for accurate placement of interstitial brachytherapy applicators},
journal = {Brachytherapy},
year = {2024},
doi = {10.1016/j.brachy.2024.06.006}
}
EUVIMED