Accuracy of a point-of-care CoaguChek test versus standard laboratory coagulation monitoring in cardiac surgery involving cardiopulmonary bypass: randomized clinical trial
Brazilian Journal of Anesthesiology (English Edition) · 2026
DOI: 10.1016/j.bjane.2025.844714 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Brazilian Journal of Anesthesiology (English Edition), 2026. Verfasst von Fabrício Tavares Mendonça und 6 weiteren Autor:innen. Beteiligte Einrichtung: Hospital de Base do Distrito Federal, Departamento de Anestesiologia, Brasília, DF, Brasil. Hauptthemen laut MEDLINE: Cardiopulmonary Bypass, Cardiac Surgical Procedures, Point-of-Care Systems, Monitoring, Intraoperative. Die Arbeit nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Intraoperative coagulopathies are common in complex surgeries, and timely coagulation monitoring is crucial. We assessed the accuracy of the point-of-care CoaguChek XS test against standard laboratory measurements in patients undergoing cardiac surgery with Cardiopulmonary Bypass (CPB).
Methods
We conducted a single-center, diagnostic accuracy study to assess the coagulation profile of 50 participants before and after CPB. The index test was the CoaguChek XS device and the reference test was the standard laboratory assay. The primary outcome was the accuracy of the CoaguChek device in measuring the International Normalized Ratio (INR). We pre-specified a tolerance range of ± 0.5 INR units. Secondary outcomes included accuracy in measuring prothrombin time and prothrombin activity.
Results
We included 50 patients undergoing cardiac surgery with CPB between October 2023 and January 2024. The mean (standard deviation) age was 59.2 (12.3) years, and 32 participants (64%) were male. For INR values, Lin’s coefficient was 0.72 (95% CI: 0.60‒0.82) pre-CPB and 0.66 (95% CI: 0.50‒0.77) post-CPB, both indicating good agreement. In the pre-CPB period, on average, the index test readings exceeded reference readings by 0.045 INR units (95% CI: 0.030‒0.059, p < 0.001), while in post-CPB period, index readings were, on average, 0.064 INR units lower (95% CI: -0.09 to -0.04, p < 0.001). Although statistically significant differences were observed, they fell within predefined tolerance range and were considered clinically irrelevant. Analyses of secondary outcomes were consistent with the primary outcome findings.
Conclusion
In patients undergoing cardiac surgery with cardiopulmonary bypass, CoaguChek XS provided results comparable to standard laboratory coagulation monitoring, both pre-and post-cardiopulmonary bypass. Research Board Approval Plataforma Brasil CAAE 70266023.8.0000.5553. Clinical Trials Registration (https://clinicaltrials.gov/study/NCT06037720) NCT06037720, 14/09/2023.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Hospital de Base do Distrito Federal, Departamento de Anestesiologia, Brasília, DF, Brasil.
- Hospital de Base do Distrito Federal, Departamento de Anestesiologia, Brasília, DF, Brasil. Electronic address: larissagoveia@yahoo.com.
Zitieren
APA
Mendonça, F. T., Matos, S. H., Moreira, L. G., Pereira, I. L., Dino, L. L., Braga, M. B., & Dias, G. H. D. S. (2026). Accuracy of a point-of-care CoaguChek test versus standard laboratory coagulation monitoring in cardiac surgery involving cardiopulmonary bypass: randomized clinical trial. Brazilian Journal of Anesthesiology (English Edition). https://doi.org/10.1016/j.bjane.2025.844714
Vancouver
Mendonça FT, Matos SH, Moreira LG, Pereira IL, Dino LL, Braga MB, et al. Accuracy of a point-of-care CoaguChek test versus standard laboratory coagulation monitoring in cardiac surgery involving cardiopulmonary bypass: randomized clinical trial. Brazilian Journal of Anesthesiology (English Edition). 2026. doi:10.1016/j.bjane.2025.844714
BibTeX
@article{Mendona2026,
author = {Mendonça, Fabrício Tavares and Matos, Sérgio Honorato de and Moreira, Larissa Goveia and Pereira, Igor Louza and Dino, Lorenzo Leite and Braga, Matheus Beserra and Dias, Gustavo Henrique Dos Santos},
title = {Accuracy of a point-of-care CoaguChek test versus standard laboratory coagulation monitoring in cardiac surgery involving cardiopulmonary bypass: randomized clinical trial},
journal = {Brazilian Journal of Anesthesiology (English Edition)},
year = {2026},
doi = {10.1016/j.bjane.2025.844714}
}
EUVIMED