EUVIMEDEUVIMED entdecken
Multizentrische StudieBeobachtungsstudie

Nitrous Oxide Manifold and Other Reduction of Emissions (NoMoreGas): a multicentre observational study evaluating pre-utilisation loss of nitrous oxide

Megan A F Thomas, Christopher J Ward, Matthew E Sinnott, Thomas W Davies, Jan M Wong, Joanna K L Wong, Gudrun Kunst, Sibtain Anwar, Pan-London Perioperative Audit and Research Network Committee, Local investigators

British Journal of Anaesthesia · 2024-12 · 133 (6) · 1427-1434

DOI: 10.1016/j.bja.2024.08.027 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift British Journal of Anaesthesia, 2024-12. Verfasst von Megan A F Thomas und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesia, King's College Hospital NHS Foundation Trust, London, UK. Hauptthemen laut MEDLINE: Nitrous Oxide, Anesthetics, Inhalation. Die Arbeit wurde bisher 22-mal zitiert (Crossref) und nennt 30 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Nitrous oxide (N2O) is a potent greenhouse gas that contributes significantly to the healthcare sector's carbon footprint. Pre-utilisation losses of N2O are up to 95%. Decommissioning manifolds can reduce these losses. METHODS: Hospitals in our Greater London research network with at least one active N2O manifold were included in the Nitrous Oxide Manifold and Other Reduction of Emissions (NoMoreGas) study. N2O utilisation data were collected continuously over 5 days and extrapolated over a year, in addition to collecting procurement records from the preceding financial year. The primary outcome was the discrepancy between clinically utilised N2O and the quantity procured by hospitals, referred to as the 'N2O gap'. Secondary outcomes included anaesthetists' self-reported utilisation of N2O and their opinions on manifold decommissioning. RESULTS: Eighteen of 53 hospitals were included. In total, 6 487 200 L of N2O were procured with a median (IQR) of 304 200 (183 600-473 400) L per site. During the 5-day data collection period, sites utilised a median (IQR) of 501 (42-1409) L of N2O. Extrapolating over a year resulted in a median (IQR) annual utilisation of 36 573 (3066-102 857) L per site and a total of 1 175 348 L. This represented an estimated 18% of the N2O procured, suggesting pre-utilisation losses of 5 311 852 L. Among surveyed anaesthetists, 70% (n=309) reported using N2O within the previous year, with one-third (n=97) using it once a week or more. There was widespread support for decommissioning manifolds. CONCLUSIONS: Consistent with other reports, the data demonstrate a substantial discrepancy between the quantities of N2O procured and utilised clinically, indicative of significant pre-utilisation losses. Our findings support the decommissioning of N2O manifolds for environmental and economic benefits.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Nitrous OxideAnesthetics, InhalationHumansCarbon FootprintLondonGreenhouse Gases

Stichwörter der Autor:innen

carbon footprintdecommissiongreen anaesthesiamanifoldnitrous oxidepre-utilisation losssustainability

Substanzen

Nitrous OxideAnesthetics, InhalationGreenhouse Gases

Autor:innen und Institutionen

  • Department of Anaesthesia, King's College Hospital NHS Foundation Trust, London, UK.
  • Department of Anaesthesia, King's College Hospital NHS Foundation Trust, London, UK. Electronic address: c.ward8@nhs.net.
  • Department of Anaesthesia, Ashford and St Peter's Hospitals NHS Foundation Trust, Chertsey, UK.
  • Faculty of Medicine and Dentistry, Queen Mary University of London, London, UK.
  • Department of Anaesthesia, Great Ormond Street Hospital for Children NHS Foundation Trust, London, UK.
  • Department of Anaesthesia, Barts Health NHS Trust, London, UK.

Zitieren

APA

Thomas, M. A. F., Ward, C. J., Sinnott, M. E., Davies, T. W., Wong, J. M., Wong, J. K. L., Kunst, G., Anwar, S., Pan-London Perioperative Audit and Research Network Committee, & Local investigators (2024). Nitrous Oxide Manifold and Other Reduction of Emissions (NoMoreGas): a multicentre observational study evaluating pre-utilisation loss of nitrous oxide. British Journal of Anaesthesia, 133(6), 1427-1434. https://doi.org/10.1016/j.bja.2024.08.027

Vancouver

Thomas MAF, Ward CJ, Sinnott ME, Davies TW, Wong JM, Wong JKL, et al. Nitrous Oxide Manifold and Other Reduction of Emissions (NoMoreGas): a multicentre observational study evaluating pre-utilisation loss of nitrous oxide. British Journal of Anaesthesia. 2024;133(6):1427-1434. doi:10.1016/j.bja.2024.08.027

BibTeX
@article{Thomas2024,
  author = {Thomas, Megan A F and Ward, Christopher J and Sinnott, Matthew E and Davies, Thomas W and Wong, Jan M and Wong, Joanna K L and Kunst, Gudrun and Anwar, Sibtain and Pan-London Perioperative Audit and Research Network Committee, Pan-London Perioperative Audit and Research Network Committee and Local investigators, Local investigators},
  title = {Nitrous Oxide Manifold and Other Reduction of Emissions (NoMoreGas): a multicentre observational study evaluating pre-utilisation loss of nitrous oxide},
  journal = {British Journal of Anaesthesia},
  year = {2024},
  volume = {133},
  number = {6},
  pages = {1427--1434},
  doi = {10.1016/j.bja.2024.08.027}
}