EUVIMEDEUVIMED entdecken
Open Access · CC BY-NC-ND 4.0

A portable, low-cost and high-throughput electrochemical impedance spectroscopy device for point-of-care biomarker detection

Xuanjie Ye, Tianxiang Jiang, Yuhao Ma, Daniel To, Shuren Wang, Jie Chen

Biosensors and Bioelectronics: X · 2023-05 · 13 · 100301

DOI: 10.1016/j.biosx.2022.100301 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors and Bioelectronics: X, 2023-05. Verfasst von Xuanjie Ye und 5 weiteren Autor:innen. Die Arbeit wurde bisher 14-mal zitiert (Crossref) und nennt 49 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: China Scholarship Council.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This paper presents a portable, fast and accurate electrochemical impedance spectroscopy (EIS) device with 8-well interdigitated electrode chips for biomarker detection. The design adopts low crest factor multisine signal synthesis at low frequencies ( 1 kHz), which significantly increases measurement speed without sacrificing accuracy. In addition, the low excitation amplitude of 10 mV preserves impedance linearity and protects the biosamples. The system achieved an average magnitude accuracy error of 0.30% in the frequency range of interest and it requires only 0.46 s to scan 28 frequency points from 10 Hz to 1 MHz. Experiments were conducted to test the capability to detect antibodies against SARS-CoV-2. Gold nanoparticles bound with protein G (GNP-G) were employed as the conjugated secondary antibody probe to detect anti-SARS-CoV-2 IgG in serum. A highly statistical significance (p = 7×10−6) could be found in the impedance data at 10 kHz. The impedance magnitude alteration caused by the GNP-G of the positive and negative groups were 27.2%±13.6% and 4.1%±1.7%, respectively. The results imply that the proposed system enables rapid COVID-19 antibody biomarker detection. Moreover, the EIS system and GNPs have the potential to be modified to detect other biomarkers.

Förderung

  • China Scholarship Council

Zitieren

APA

Ye, X., Jiang, T., Ma, Y., To, D., Wang, S., & Chen, J. (2023). A portable, low-cost and high-throughput electrochemical impedance spectroscopy device for point-of-care biomarker detection. Biosensors and Bioelectronics: X, 13, 100301. https://doi.org/10.1016/j.biosx.2022.100301

Vancouver

Ye X, Jiang T, Ma Y, To D, Wang S, Chen J. A portable, low-cost and high-throughput electrochemical impedance spectroscopy device for point-of-care biomarker detection. Biosensors and Bioelectronics: X. 2023;13:100301. doi:10.1016/j.biosx.2022.100301

BibTeX
@article{Ye2023,
  author = {Ye, Xuanjie and Jiang, Tianxiang and Ma, Yuhao and To, Daniel and Wang, Shuren and Chen, Jie},
  title = {A portable, low-cost and high-throughput electrochemical impedance spectroscopy device for point-of-care biomarker detection},
  journal = {Biosensors and Bioelectronics: X},
  year = {2023},
  volume = {13},
  pages = {100301},
  doi = {10.1016/j.biosx.2022.100301}
}
Kostenlos starten →