Highly sensitive detection of adeno-associated virus serotype 9 enabled by a nanobody-functionalized multiphase TiO2 photoelectrochemical immunosensor
Biosensors and Bioelectronics · 2026
DOI: 10.1016/j.bios.2026.118734 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Biosensors and Bioelectronics, 2026. Verfasst von Ying Hou und 6 weiteren Autor:innen. Beteiligte Einrichtung: School of Chemistry and Chemical Engineering, University of Jinan, Jinan, 250022, China; State Key Laboratory of Analytical Chemistry for Life Scienc…. Hauptthemen laut MEDLINE: Titanium, Biosensing Techniques, Electrochemical Techniques, Dependovirus. Die Arbeit nennt 25 Literaturangaben. Gefördert von: Shangqiu Normal University, Science and Technology Commission of Shanghai Municipality, Shandong Province Natural Science Foundation.
Abstract (Originalsprache)
The reliable detection and quantification of adeno-associated virus serotype 9 (AAV9) is crucial for gene therapy development and quality control. This work presents a significant advancement by developing a novel, nanobody-based photoelectrochemical (PEC) immunosensor for the highly sensitive and specific detection of AAV9. The core is a laser-synthesized multiphase TiO2 heterojunction photoanode fabricated directly on a titanium substrate. This one-step, mask-free laser processing technique ensures excellent material consistency and superior photocurrent response, overcoming the batch-to-batch variability associated with traditional electrode modification methods. The sensing interface is further engineered by immobilizing a high-affinity anti-AAV9 nanobody, which offers minimal steric hindrance and enhanced epitope accessibility compared to conventional antibodies. The synergistic integration of a high-performance multiphase photoactive material with a biological recognition element, resulting in a satisfactory sensing platform. The fabricated immunosensor demonstrates a wide linear detection range from 5 × 107 to 1 × 1010 vg/mL (vg/mL: viral genomes per milliliter) for AAV9, with a remarkably low calculated detection limit (LOD) of 1 × 107.2 vg/mL and limit of quantitation (LOQ) of 3 × 107.2 vg/mL. To our knowledge, this represents the first successful application of an AAV9-specific nanobody within a PEC biosensing platform. This study not only provides a powerful tool for virological analysis but also establishes a versatile framework that can be adapted for detecting other viral targets by exchanging the nanobody probe, showcasing broad potential in biomedical diagnostics.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Chemistry and Chemical Engineering, University of Jinan, Jinan, 250022, China; State Key Laboratory of Analytical Chemistry for Life Science, School of Chemistry and Chemical Engineering, Nanjing University, N…
- Shanghai Novamab Biopharmaceuticals Co., Ltd., Shanghai, 201318, China.
- School of Chemistry and Chemical Engineering, University of Jinan, Jinan, 250022, China.
- Shanghai Novamab Biopharmaceuticals Co., Ltd., Shanghai, 201318, China. Electronic address: ykwan@novamab.com.
- School of Chemistry and Chemical Engineering, University of Jinan, Jinan, 250022, China. Electronic address: lijuan910219@126.com.
- Henan Key Laboratory of Biomarker Detection and Diagnosis for Neurodegenerative Diseases, School of Chemistry and Chemical Engineering, Shangqiu Normal University, Shangqiu, 476000, China; School of Chemistry and Chemic…
Förderung
- Shangqiu Normal University
- Science and Technology Commission of Shanghai Municipality
- Shandong Province Natural Science Foundation
Zitieren
APA
Hou, Y., Zhu, M., Tian, X., Xu, R., Wan, Y., Li, J., & Jia, Y. (2026). Highly sensitive detection of adeno-associated virus serotype 9 enabled by a nanobody-functionalized multiphase TiO2 photoelectrochemical immunosensor. Biosensors and Bioelectronics. https://doi.org/10.1016/j.bios.2026.118734
Vancouver
Hou Y, Zhu M, Tian X, Xu R, Wan Y, Li J, et al. Highly sensitive detection of adeno-associated virus serotype 9 enabled by a nanobody-functionalized multiphase TiO2 photoelectrochemical immunosensor. Biosensors and Bioelectronics. 2026. doi:10.1016/j.bios.2026.118734
BibTeX
@article{Hou2026,
author = {Hou, Ying and Zhu, Min and Tian, Xiaoxue and Xu, Ruixian and Wan, Yakun and Li, Juan and Jia, Yue},
title = {Highly sensitive detection of adeno-associated virus serotype 9 enabled by a nanobody-functionalized multiphase TiO2 photoelectrochemical immunosensor},
journal = {Biosensors and Bioelectronics},
year = {2026},
doi = {10.1016/j.bios.2026.118734}
}