Fachartikel in der Zeitschrift Biosensors and Bioelectronics, 2026. Verfasst von Sadhna Singh und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Physics, Indian Institute of Technology Roorkee, Roorkee, 247667, India. Hauptthemen laut MEDLINE: Biosensing Techniques, Spectrum Analysis, Raman, Glial Fibrillary Acidic Protein, Brain Injuries, Traumatic. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 46 Literaturangaben. Gefördert von: DRDO Institute of Nuclear Medicine and Allied Sciences, Council of Scientific and Industrial Research, India, Ministry of Education, India.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

To prevent neurological damage and enable effective treatment, especially for military personnel, there is a clinical requirement for rapid, accurate, and early diagnosis of blast-exposed traumatic brain injury (TBI). A silver nano-sculptured thin film (nSTF) based surface-enhanced Raman spectroscopy (SERS) biosensor is developed to detect TBI biomarker GFAP. The biosensor chip is prepared by immobilizing an anti-GFAP antibody onto the 4-aminothiophenol (4ATP) adsorbed over Ag-nSTF deposited over Si substrate. The GFAP biosensor is based on the enhancement in the intensity of SERS signal of 4ATP. SERS signals are recorded for varying concentrations of GFAP over the sensor chip. The limit of detection (LOD) of the sensor for GFAP in PBS, and in spiked control serum is achieved to be 1 pg/ml. The concentrations of GFAP in three blast-exposed moderately injured rat sera, validated with ELISA, are diluted 1000 times and found to be 15, 24, and 37 pg/ml. The sensor's performance is further validated in blind-fold tests of 1000 times diluted rat sera to predict blast or no-blast cases, in accordance with ELISA. The fabricated sensor offers specificity, repeatability, and reusability, along with a compact size, minimal analyte requirement (3 μl), and a rapid Raman readout (500 ms) with a total response time

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Biosensing TechniquesSpectrum Analysis, RamanGlial Fibrillary Acidic ProteinBrain Injuries, TraumaticBlast InjuriesAnimalsRatsSilverLimit of DetectionAniline CompoundsAntibodies, ImmobilizedBiomarkersHumansNanostructuresSulfhydryl Compounds

Stichwörter der Autor:innen

Glial fibrillary acidic proteinNano-sculptured thin filmSurface enhanced Raman spectroscopyTraumatic brain injury

Substanzen

Glial Fibrillary Acidic ProteinSilverAniline Compounds4-aminothiophenolAntibodies, ImmobilizedGfap protein, ratBiomarkersSulfhydryl Compounds

Autor:innen und Institutionen

  • Department of Physics, Indian Institute of Technology Roorkee, Roorkee, 247667, India.
  • Department of Neurobehavioral Research, INMAS, DRDO, India.
  • Traumatic Brain Injury and Metabolomics (TBIM) Department, INMAS, DRDO, India. Electronic address: priyanka.inmas@gov.in.
  • Department of Physics, Indian Institute of Technology Roorkee, Roorkee, 247667, India; Centre for Photonics and Quantum Communication Technology, Indian Institute of Technology Roorkee, Roorkee, 247667, India. Electroni…

Förderung

  • DRDO Institute of Nuclear Medicine and Allied Sciences
  • Council of Scientific and Industrial Research, India
  • Ministry of Education, India
  • Defence Research and Development Organisation
  • Science and Engineering Research Board
  • India Ministry of Science & Technology Department of Science and Technology

Zitieren

APA

Singh, S., Yadav, A., Arora, P., Trivedi, R., Sharma, P., & Srivastava, S. K. (2026). Surface enhanced Raman spectroscopy-based biosensor using nano-sculptured thin film to detect glial fibrillary acidic protein (GFAP) in blast-exposed moderately injured rat serum. Biosensors and Bioelectronics. https://doi.org/10.1016/j.bios.2026.118677

Vancouver

Singh S, Yadav A, Arora P, Trivedi R, Sharma P, Srivastava SK. Surface enhanced Raman spectroscopy-based biosensor using nano-sculptured thin film to detect glial fibrillary acidic protein (GFAP) in blast-exposed moderately injured rat serum. Biosensors and Bioelectronics. 2026. doi:10.1016/j.bios.2026.118677

BibTeX
@article{Singh2026,
  author = {Singh, Sadhna and Yadav, Arti and Arora, Palkin and Trivedi, Richa and Sharma, Priyanka and Srivastava, Sachin Kumar},
  title = {Surface enhanced Raman spectroscopy-based biosensor using nano-sculptured thin film to detect glial fibrillary acidic protein (GFAP) in blast-exposed moderately injured rat serum},
  journal = {Biosensors and Bioelectronics},
  year = {2026},
  doi = {10.1016/j.bios.2026.118677}
}
Kostenlos starten →