EUVIMEDEUVIMED entdecken
Crossref

Light-addressable sandwich photoelectrochemical immunosensor array and lateral flow immunoassays with self-calibration using quantum dots-sensitized and porphyrin-engineered MOFs for accurate detection of amyloid β-proteins

Louqun Wang, Jingge Niu, Xiaoli Chen, Yun Zhu, Xiaohong Hou

Biosensors and Bioelectronics · 2026

DOI: 10.1016/j.bios.2026.118424 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors and Bioelectronics, 2026. Verfasst von Louqun Wang und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Pharmacy, Shenyang Pharmaceutical University, Shenyang, Liaoning Province, 110016, PR China. Hauptthemen laut MEDLINE: Amyloid beta-Peptides, Quantum Dots, Biosensing Techniques, Alzheimer Disease. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 41 Literaturangaben. Gefördert von: National Natural Science Foundation of China.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Alzheimer's disease (AD) is a neurodegenerative disorder characterized by the accumulation of amyloid-beta (Aβ) plaques, making Aβ a crucial biomarker for early diagnosis. In this study, we developed a novel light-addressable sandwich photoelectrochemical (PEC) immunosensor array for the detection of Aβ42, aiming to enhance diagnostic methods for AD. The array utilized cadmium telluride quantum dot-sensitized UiO-66 (U@C) as the sensing matrix and PCN-224@ZnIn2S4 nanoflower heterojunctions (P@Z) as immunoprobes, enabling multiplexed detection through laser pen activation on a single FTO electrode with a self-calibration method to minimize background noise and baseline drift. The assay required 60 min for incubation steps before detection; however, with 11 simultaneous detection points, it significantly reduced batch processing time and enhanced practicality. For point-of-care applications, we also developed a PEC-lateral flow immunoassay (PEC-LFIA) test strip that allowed for real-time, quantitative detection of Aβ42 with only 15 min of incubation required for testing. Both platforms effectively detected Aβ42 in PBS, artificial cerebrospinal fluid (CSF), and human plasma at concentrations ranging from fg/mL to ng/mL. Notably, sensitivity in human plasma was five times greater than that in artificial CSF. The PEC array achieved detection limits of 19.5 fg/mL in CSF and 3.1 fg/mL in plasma, while the PEC-LFIA strip demonstrated limits of 17.8 fg/mL in CSF and 3.1 fg/mL in human plasma. These advancements significantly reduced patient burden and brought us closer to utilizing a single drop of blood for monitoring brain aging.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Amyloid beta-PeptidesQuantum DotsBiosensing TechniquesAlzheimer DiseaseMetal-Organic FrameworksPeptide FragmentsHumansImmunoassayElectrochemical TechniquesCadmium CompoundsLimit of DetectionPorphyrinsTelluriumAntibodies, ImmobilizedPhthalic Acids

Stichwörter der Autor:innen

Alzheimer's diseaseAβ42Human plasmaLateral flow immunoassayMetal-organic frameworksPhotoelectrochemical detection

Substanzen

Amyloid beta-PeptidesCadmium CompoundsPorphyrinsMetal-Organic Frameworkscadmium tellurideTelluriumamyloid beta-protein (1-42)UiO-66Peptide FragmentsAntibodies, Immobilized

Autor:innen und Institutionen

  • School of Pharmacy, Shenyang Pharmaceutical University, Shenyang, Liaoning Province, 110016, PR China.
  • School of Pharmaceutical Engineering, Shenyang Pharmaceutical University, Shenyang, Liaoning Province, 110016, PR China.
  • School of Pharmaceutical Engineering, Shenyang Pharmaceutical University, Shenyang, Liaoning Province, 110016, PR China. Electronic address: syphu_houxiaohong@163.com.

Förderung

  • National Natural Science Foundation of China

Zitieren

APA

Wang, L., Niu, J., Chen, X., Zhu, Y., & Hou, X. (2026). Light-addressable sandwich photoelectrochemical immunosensor array and lateral flow immunoassays with self-calibration using quantum dots-sensitized and porphyrin-engineered MOFs for accurate detection of amyloid β-proteins. Biosensors and Bioelectronics. https://doi.org/10.1016/j.bios.2026.118424

Vancouver

Wang L, Niu J, Chen X, Zhu Y, Hou X. Light-addressable sandwich photoelectrochemical immunosensor array and lateral flow immunoassays with self-calibration using quantum dots-sensitized and porphyrin-engineered MOFs for accurate detection of amyloid β-proteins. Biosensors and Bioelectronics. 2026. doi:10.1016/j.bios.2026.118424

BibTeX
@article{Wang2026,
  author = {Wang, Louqun and Niu, Jingge and Chen, Xiaoli and Zhu, Yun and Hou, Xiaohong},
  title = {Light-addressable sandwich photoelectrochemical immunosensor array and lateral flow immunoassays with self-calibration using quantum dots-sensitized and porphyrin-engineered MOFs for accurate detection of amyloid β-proteins},
  journal = {Biosensors and Bioelectronics},
  year = {2026},
  doi = {10.1016/j.bios.2026.118424}
}
Kostenlos starten →