Fachartikel in der Zeitschrift Biosensors and Bioelectronics, 2026. Verfasst von Lemin Yu und 6 weiteren Autor:innen. Beteiligte Einrichtung: College of Information and Electrical Engineering, Shenyang Agricultural University, Shenyang, 110866, China. Electronic address: 2023240103@stu.syau…. Hauptthemen laut MEDLINE: Salmonella typhimurium, Biosensing Techniques, Aptamers, Nucleotide, Microfluidic Analytical Techniques. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 41 Literaturangaben. Gefördert von: National Natural Science Foundation of China, Department of Science and Technology of Liaoning Province.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Salmonella typhimurium (S. typhimurium)-induced contamination is emerging as a severe threat to food safety. Therefore, rapid and sensitive detection of S. typhimurium is urgently required. Conventional detection methodologies exhibit persistent limitations. The combination of microfluidic chips with magnetic beads (MBs) for isolating bacteria such as S. typhimurium has emerged as a prominent research focus. However, flexible and tunable magnetic field control methods for optimizing MB chain formation and enhancing the capture efficiency of the target bacteria are lacking. We developed a microfluidic fluorescent biosensor to achieve an efficient on-site detection of S. typhimurium using magnetic bead (MB) chains for continuous-flow electromagnetic separation, a dual-aptamer recognition method for specifically labeling S. typhimurium, and a customized smartphone-based real-time fluorescence system for quantitative detection. Aptamer-conjugated MBs in the separation channel were distributed into high-density bead chains in a high-intensity gradient magnetic field generated by an electromagnetic actuation module. The samples continuously flew through the separation channel, in which the target bacteria were specifically captured by aptamer-conjugated MBs. Aptamer-functionalized fluorescent labels were passed through the separation channel to form bacteria-MB-fluorescent label complexes for quantitative detection. Under optimal conditions, the microfluidic biosensor completed sample-in-result-out detection within 30 min, with a detection limit of 2 colony-forming unit/mL for S. typhimurium. This biosensor can be potentially applied for detecting food pathogens, making it suitable for supervising food production and monitoring catering hygiene.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Salmonella typhimuriumBiosensing TechniquesAptamers, NucleotideMicrofluidic Analytical TechniquesLimit of DetectionEquipment DesignFood MicrobiologyFood Contamination

Stichwörter der Autor:innen

BiosensorDual aptamerElectromagnetic actuationMagnetic bead chainsMicrofluidic chipSalmonella typhimurium

Substanzen

Aptamers, Nucleotide

Autor:innen und Institutionen

  • College of Information and Electrical Engineering, Shenyang Agricultural University, Shenyang, 110866, China. Electronic address: 2023240103@stu.syau.edu.cn.
  • College of Information and Electrical Engineering, Shenyang Agricultural University, Shenyang, 110866, China. Electronic address: wangjingyi@syau.edu.cn.
  • College of Bioscience and Biotechnology, Shenyang Agricultural University, Shenyang, 110866, China.
  • College of Information and Electrical Engineering, Shenyang Agricultural University, Shenyang, 110866, China.
  • Department of Stomatology, General Hospital of Northern Theater Command, No. 83 Wenhua Road, Shenhe District, Shenyang, 110016, China. Electronic address: sphinirvanax@qq.com.

Förderung

  • National Natural Science Foundation of China
  • Department of Science and Technology of Liaoning Province

Zitieren

APA

Yu, L., Wang, J., Lin, S., Lu, W., Li, Q., Shi, Y., & Wang, J. (2026). Integrated sample-in-result-out microfluidic biosensor for rapid ultrasensitive detection of Salmonella typhimurium using electromagnetically actuated aptamer-conjugated magnetic bead chains. Biosensors and Bioelectronics. https://doi.org/10.1016/j.bios.2026.118422

Vancouver

Yu L, Wang J, Lin S, Lu W, Li Q, Shi Y, et al. Integrated sample-in-result-out microfluidic biosensor for rapid ultrasensitive detection of Salmonella typhimurium using electromagnetically actuated aptamer-conjugated magnetic bead chains. Biosensors and Bioelectronics. 2026. doi:10.1016/j.bios.2026.118422

BibTeX
@article{Yu2026,
  author = {Yu, Lemin and Wang, Jingyi and Lin, Shoukun and Lu, Wenzhe and Li, Quan and Shi, Yubo and Wang, Jingfu},
  title = {Integrated sample-in-result-out microfluidic biosensor for rapid ultrasensitive detection of Salmonella typhimurium using electromagnetically actuated aptamer-conjugated magnetic bead chains},
  journal = {Biosensors and Bioelectronics},
  year = {2026},
  doi = {10.1016/j.bios.2026.118422}
}