EUVIMEDEUVIMED entdecken
Crossref

Biomimetic self-sustained ion oscillation in miscibility gap and chemical inductance impact in potassium sensing

Marini L, Roop L Mahajan, Chithra Lekha P, T Devasena

Biosensors and Bioelectronics · 2026

DOI: 10.1016/j.bios.2026.118413 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Biosensors and Bioelectronics, 2026. Verfasst von Marini L und 3 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Nanoscience and Technology, AC Tech Campus, Anna University, Chennai, 600 025, Tamil Nadu, India. Hauptthemen laut MEDLINE: Potassium, Biosensing Techniques, Biomimetic Materials. Die Arbeit nennt 38 Literaturangaben. Gefördert von: All India Council for Technical Education.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A biomimetic potassium ion-selective electrode (ISE) was developed using a graphene oxide (GO)-coated nickel foam (NF) substrate functionalized with a polyethylene glycol (PEG)-based ion-selective layer (ISL). The sensing mechanism centers on a miscibility gap formed within the ISL due to PEG's amphiphilic nature when complexed with potassium iodide-driving phase separation into hydrophilic and hydrophobic domains. This phase heterogeneity introduces interfacial barriers that delay ionic relaxation and facilitate self-sustained ion oscillation, as evidenced by low-frequency inductive behavior in electrochemical impedance spectroscopy (EIS)-demonstrating a chemical inductor effect. The design eliminates the need for costly ionophores, such as valinomycin, presenting a scalable and cost-effective sensing platform. The sensor exhibited a linear response to K+ concentrations ranging from 0.5 to 10 mM, with a high sensitivity of 14.09 × 102 mA/mM·cm2 and a detection limit of 0.69 mM. Selectivity tests showed minimal interference from Na+, Ca2+, urea, NH4+, and ascorbic acid ions, and the sensor demonstrated excellent stability over 100 cycles. These findings provide a robust framework for developing low-cost, biomimetic electrochemical potassium sensing.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

PotassiumBiosensing TechniquesBiomimetic MaterialsGraphiteDielectric SpectroscopyPolyethylene GlycolsIon-Selective ElectrodesBiomimeticsLimit of DetectionNickel

Stichwörter der Autor:innen

BiomimicryBiosensorsChemical inductorElectrochemical impedance spectroscopyMiscibility gapSelf-sustained oscillations

Substanzen

PotassiumGraphitegraphene oxidePolyethylene GlycolsNickel

Autor:innen und Institutionen

  • Centre for Nanoscience and Technology, AC Tech Campus, Anna University, Chennai, 600 025, Tamil Nadu, India.
  • Department of Mechanical Engineering, Department of Materials Science and Engineering, Virginia Tech, Blacksburg, VA, 24061, USA.
  • Centre for Clinical and Translational Research, Healthcare Technologies Division, Virginia Tech India Research and Education Forum, Indian Institute of Technology Madras Research Park, Chennai, 600 113, Tamil Nadu, Indi…
  • Centre for Nanoscience and Technology, AC Tech Campus, Anna University, Chennai, 600 025, Tamil Nadu, India. Electronic address: tdevasenabio@annauniv.edu.

Förderung

  • All India Council for Technical Education

Zitieren

APA

L, M., Mahajan, R. L., Lekha P, C., & Devasena, T. (2026). Biomimetic self-sustained ion oscillation in miscibility gap and chemical inductance impact in potassium sensing. Biosensors and Bioelectronics. https://doi.org/10.1016/j.bios.2026.118413

Vancouver

L M, Mahajan RL, Lekha P C, Devasena T. Biomimetic self-sustained ion oscillation in miscibility gap and chemical inductance impact in potassium sensing. Biosensors and Bioelectronics. 2026. doi:10.1016/j.bios.2026.118413

BibTeX
@article{L2026,
  author = {L, Marini and Mahajan, Roop L and Lekha P, Chithra and Devasena, T},
  title = {Biomimetic self-sustained ion oscillation in miscibility gap and chemical inductance impact in potassium sensing},
  journal = {Biosensors and Bioelectronics},
  year = {2026},
  doi = {10.1016/j.bios.2026.118413}
}